Agreement and Precision Study of the Nidek Mirante
Agreement and Precision Study of the Nidek Mirante OCT Compared to the Optovue RTVue XR Avanti OCT and SLO Image Comparison of the Nidek Mirante and the OPTOS P200DTx
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
This is a prospective comparative clinical study to be conducted at one clinical site located in the United States. Three Nidek Mirante devices, three Optovue RTVue XR Avanti devices and one OPTOS P200DTx will be used. For agreement and precision analysis of OCT scans and agreement analysis of image quality of OCT ACA scans, each Nidek Mirante device will be paired with one of the Optovue RTVue XR Avanti devices. One device operator will be assigned to each device pair to create three distinct operator/device configurations.
For agreement analysis of SLO image quality, one OPTOS P200DTx will be included in one of three operator/device configurations.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Massachusetts
-
Andover, Massachusetts, United States, 01840
- Andover Eye Associates
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria - Normal Group:
- Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
- Subjects who agree to participate in the study;
- Subjects with normal eye examinations (without pathology other than cataract) in both eyes on the date of the study visit as observed with a +90 Diopter Fundus Lens;
- Subjects with current best-spectacle-corrected visual acuity (BSCVA) of 20/40 or better in both eyes on the date the study visit.
Exclusion Criteria - Normal Group:
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in either eye;
- Subjects with any current ocular pathology other than cataract in either eye, as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
- Subjects with (or history of) leukemia, dementia or multiple sclerosis;
- Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
Inclusion Criteria - Glaucoma Group:
- Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent was signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
- Subjects who agree to participate in the study;
Subjects who have been diagnosed with glaucoma in the glaucoma study eye(s), with optic nerve damage as evidenced by either of the following optic disc or RNFL structural abnormalities observed via fundus exam during the study visit:
- Diffuse thinning, focal narrowing, or notching of the optic disc rim, especially at the inferior or superior poles with or without disc hemorrhage; or
- Optic disc neural rim asymmetry of the two eyes consistent with loss of neural tissue;
- Subjects with a current BSCVA of 20/40 or better in the glaucoma study eye(s) on the date of the visit;
History of a reliable Humphrey Field Analyzer (HFA) visual field (24-2 SITA Standard, white on white) measured on the day of study visit 1, or within the previous six (6) months from study visit 1 with reliable results, defined as fixation losses < 20%, or false positives < 33%, or false negatives < 33% in the glaucoma study eye(s) that is consistent with glaucomatous optic nerve damage in glaucoma study eye(s) based on at least one of the following findings:
- On pattern deviation (PD), there exists a cluster of 3 or more points in an expected location of the visual field depressed below the 5% level, at least 1 of which is depressed below the 1% level; or
- Glaucoma hemi-field test "outside normal limits."
Exclusion Criteria - Glaucoma Group:
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the glaucoma study eye(s);
- Subjects with retinal disease in the glaucoma study eye(s), as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
- Subjects with (or history of) leukemia, dementia or multiple sclerosis;
- Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
Inclusion Criteria - Retinal Disease Group:
- Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
- Subjects who agree to participate in the study;
- Subjects with a current BSCVA of 20/400 or better in the retinal disease study eye(s) at the study visit;
- Subjects diagnosed with retinal pathology including but not limited to: Non Exudative Macular Degeneration (dry AMD), Diabetic Macular Edema, Macular Hole, Epiretinal Membrane, Cystoid Macular Edema, or other retinal disease in the study eye(s) as confirmed within the past six (6) months, who exhibit structural lesions in the study eye.
Exclusion Criteria - Retinal Disease Group
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the study eye;
- Subjects with glaucoma in the retinal disease study eye(s), as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
- Subjects with (or history of) leukemia, dementia or multiple sclerosis;
- Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
Inclusion Criteria - Corneal Group
- Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
- Subjects who agree to participate in the study;
- Subjects with a current BSCVA of 20/400 or better in the study eye(s) at the study visit;
- Subjects with corneal pathologies including but not limited to: Post status LASIK surgery, Keratoconus, or other corneal dystrophies or degenerations in the study eye(s) as confirmed within the past six (6) months.
Exclusion Criteria - Corneal Group
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the study eye;
- Subjects with (or history of) leukemia, dementia or multiple sclerosis;
- Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Randomized
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: NIDEK Mirante Comparison - Normal
Subjects without any current ocular pathology other than cataract in either eye
|
The Nidek Mirante Optical Coherence Tomography (OCT) including scanning laser ophthalmoscope (SLO) function is an non-contact system for imaging the fundus and for axial cross sectional imaging of anterior and posterior ocular structures.
The Avanti is an optical coherence tomography system indicated for the in vivo imaging,axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures.
The P200DTx scanning laser ophthalmoscope (SLO) is indicated for use as a widefield and retinal fluorescence and autofluorescence imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases and disorders that manifest in the retina.
|
|
Experimental: NIDEK Mirante Comparison - Retina
Subjects diagnosed with retinal pathology
|
The Nidek Mirante Optical Coherence Tomography (OCT) including scanning laser ophthalmoscope (SLO) function is an non-contact system for imaging the fundus and for axial cross sectional imaging of anterior and posterior ocular structures.
The Avanti is an optical coherence tomography system indicated for the in vivo imaging,axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures.
The P200DTx scanning laser ophthalmoscope (SLO) is indicated for use as a widefield and retinal fluorescence and autofluorescence imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases and disorders that manifest in the retina.
|
|
Experimental: NIDEK Mirante Comparison - Glaucoma
Subjects who have been diagnosed with glaucoma
|
The Nidek Mirante Optical Coherence Tomography (OCT) including scanning laser ophthalmoscope (SLO) function is an non-contact system for imaging the fundus and for axial cross sectional imaging of anterior and posterior ocular structures.
The Avanti is an optical coherence tomography system indicated for the in vivo imaging,axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures.
The P200DTx scanning laser ophthalmoscope (SLO) is indicated for use as a widefield and retinal fluorescence and autofluorescence imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases and disorders that manifest in the retina.
|
|
Experimental: NIDEK Mirante Comparison - Corneal disease
Subjects with corneal pathologies
|
The Nidek Mirante Optical Coherence Tomography (OCT) including scanning laser ophthalmoscope (SLO) function is an non-contact system for imaging the fundus and for axial cross sectional imaging of anterior and posterior ocular structures.
The Avanti is an optical coherence tomography system indicated for the in vivo imaging,axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Agreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Agreement of the measured Macular thickness (μm) between the test device and the predicate device for the OCT function of NIDEK Mirante.
|
1 day
|
|
Agreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Agreement of the measured Retinal nerve fiber layer thickness around Optic nerve head between the test device and the predicate device for the OCT function of NIDEK Mirante.
|
1 day
|
|
Agreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Agreement of the measured Corneal thickness between the test device and the predicate device for the OCT function of NIDEK Mirante.
|
1 day
|
|
Agreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTx
Time Frame: 1 day
|
Agreement of the SLO Image quality between the test device and the predicate device for the SLO function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility for each image patterns (SLO Color Fundus, SLO B-FAF and SLO G-FAF) on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality. |
1 day
|
|
Agreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Agreement of the ACA Image quality between the test device and the predicate device for the OCT function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality. |
1 day
|
|
Precision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Precision (repeatability) of the measured Full retinal thickness at Macula for the OCT function of the NIDEK TONOREF III.
|
1 day
|
|
Precision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Precision (repeatability) of the measured Retinal nerve fiber layer thickness around Optic nerve head at Macula for the OCT function of the NIDEK TONOREF III.
|
1 day
|
|
Precision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT
Time Frame: 1 day
|
Precision (repeatability) of the measured Thickness of the Cornea for the OCT function of the NIDEK TONOREF III.
|
1 day
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Adverse Events
Time Frame: 1 day
|
An adverse event (AE) is defined as any unfavorable and unintended sign (including an abnormal laboratory finding), symptom, or disease temporally associated with the use of an investigational device but not necessarily related to the investigational device.
|
1 day
|
Collaborators and Investigators
Sponsor
Investigators
- Study Director: Tadakazu Ichimura, Nidek Co. LTD.
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- Nidek Mirante-001
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.