Videolaryngoscopes for Double Lumen Tube Intubations in Simulated Easy and Difficult Airway

24. prosince 2015 aktualizováno: Imam Abdulrahman Bin Faisal University

A Comparison of Three Videolaryngoscopes for Double-lumen Tubes Intubation in Simulated Easy and Difficult Airways

The investigators hypothesized that the use of the King Vision™ and the Airtraq® VL would reduce the time to DLT intubation compared with the GlideScope® and Macintosh in simulated easy and difficult airways. The investigators have considered to assess the efficacy of each device in manikins before considering to evaluate them in patients undergoing thoracic procedures.

Twenty-one staff anaesthesiologists who had limited prior experience in using the VLs for DLT intubation participated in this randomised crossover study. Following a brief demonstration and two practice attempts, participants were volunteered to insert a DLT using the Macintosh, GlideScope®, Airtraq®, and King Vision™ on two high-fidelity easy and difficult airway simulators in a computer-generated randomized sequence. The primary endpoint, time to DLT intubation, as well as, the views obtained at laryngoscopy, ease of intubation, numbers of laryngoscopy attempts and optimisation manoeuvers, and failure to intubation; defined as an attempt took longer than 150 seconds, were recorded.

Přehled studie

Detailní popis

Several regional surveys demonstrated that most thoracic anesthesiologists are using the double-lumen endobronchial tubes (DLT) as the first-choice lung separation technique. DLT, when compared with single lumen tracheal tube, can be more difficult to insert in patients with difficult airways. The videolaryngoscopes (VL) have the potential to facilitate the placement of the DLTs for lung separation in patients with predicted or unanticipated difficult airway.

The use of GlideScope® (Verathon Inc., Bothell, WA, USA), a VL with an angulated blade, has been associated with variable times to DLT intubation according to the prior experience of the operators, despite superior visualization of the glottis The channeled Airtraq® (Prodol Meditec S.A., Vizcaya, Spain) and standard non-channeled blade of the King Vision™ (Ambu, Ballerup, Copenhagen, Denmark), may offer additional benefits for DLT intubation in patients with limited mouth opening or restricted neck movement, in whom the use of traditional VL like as the Glidescope® could be difficult. This is because of the larger outer diameter, the distal curvature and the increased rigidity of the DLT. Of note, the longer times to DLT intubation with the use of different VL could be shortened with building up the operator's experience.

To the best of the investigators knowledge, the comparison of the effects of the Macintosh, GlideScope®, Airtraq®, and King Vision™ VL on the time to DLT intubation in simulated easy and difficult airways has not yet been studied.

Two high-fidelity simulators (Airway Management Trainer, model AA-3100, Laerdal, Kent, UK) were prepared to simulate easy and difficult airway situations, as described by Marshall et al. and Wang et al. The "easy" airway was established with the manikin in a neutral position. The "difficult" airway setting was obtained by placing an Oasis Elite™ Prone Head Rest, Adult (140 mm in height) (Covidien, Mansfield, MA) under the occiput and securing the head position with adhesive tape, object to replicate cervical-collar use. Positioning was confirmed after each attempt to ensure consistency.

Typ studie

Intervenční

Zápis (Aktuální)

21

Fáze

  • Nelze použít

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní místa

    • EP
      • Al-Khobar, EP, Saudská arábie, 31952
        • Dammam University KFHU

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

23 let až 65 let (Dospělý, Starší dospělý)

Přijímá zdravé dobrovolníky

Ano

Pohlaví způsobilá ke studiu

Všechno

Popis

Inclusion Criteria:

  • Staff anesthesiologists
  • Signed consent
  • Limited experience in using videolaryngoscopes for inserting double lumen tubes

Exclusion Criteria:

  • Decline to sign the consent
  • Non receiving the prior training demonstration
  • Non practicing once with each device tested on each simulator scenario

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Přidělení: Randomizované
  • Intervenční model: Crossover Assignment
  • Maskování: Žádné (otevřený štítek)

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Komparátor placeba: Macintosh
The participants intubated the easy airway simulator with a double lumen tube using the Macintosh laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the Macintosh laryngoscope followed with intubating the difficult airway simulator
Komparátor placeba: GlideScope
The participants intubated the easy airway simulator with a double lumen tube using the GlideScope laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the GlideScope followed with intubating the difficult airway simulator
Aktivní komparátor: Airtraq
The participants intubated the easy airway simulator with a double lumen tube using the Airtraq laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the Airtraq followed with intubating the difficult airway simulator
Aktivní komparátor: King Vision
The participants intubated the easy airway simulator with a double lumen tube using the King Vision followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the King Vision followed with intubating the difficult airway simulator

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Time to achieve successful double lumen tube intubation
Časové okno: for 6 min after using each device
defined as the time when the investigated laryngoscope passed the central incisors to when the tip of the bronchial lumen passed through the glottis, as confirmed visually by the operator (in the Macintosh group) or by the investigator, thanks to remote screens (in the VLs groups).
for 6 min after using each device

Sekundární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Best view during laryngoscopy using the classification described by Cormack and Lehane
Časové okno: for 1 min after laryngoscopy
using the classification described by Cormack and Lehane
for 1 min after laryngoscopy
Difficulty of intubation using a Visual Analog Scale (VAS)
Časové okno: for 6 min after using each device
The difficulty of intubation was evaluated using a visual analog scale (VAS) (ranging from 0: extremely easy, to 100: extremely difficult) expressed by the anesthesiologists after the DLT intubation
for 6 min after using each device
Number of first-pass success
Časové okno: for 3 min after using each device
first-pass success calculated as number of first-attempt successes/number of patients with an intubation attempt
for 3 min after using each device
Number of the required optimisation manoeuvers
Časové okno: for 6 min after using each device
When the anaesthesiologists encountered difficulty in visualising vocal cords or placing the tube, they were allowed to use any manoeuver they would normally use to navigate the tube into the trachea including readjustment of the blade or tube or to ask the supervising investigator to help solve the problem, and gave suggestions or instructions
for 6 min after using each device

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Vyšetřovatelé

  • Vrchní vyšetřovatel: Mohamed R El Tahan, MD, Anesthesiology Department, Dammam University, Dammam, Saudi Arabia
  • Ředitel studie: Abdulmohsen A Al Ghamdi, MD, Anesthesiology Department, Dammam University, Dammam, Saudi Arabia

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia

1. března 2015

Primární dokončení (Aktuální)

1. května 2015

Dokončení studie (Aktuální)

1. června 2015

Termíny zápisu do studia

První předloženo

22. prosince 2015

První předloženo, které splnilo kritéria kontroly kvality

24. prosince 2015

První zveřejněno (Odhad)

28. prosince 2015

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Odhad)

28. prosince 2015

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

24. prosince 2015

Naposledy ověřeno

1. prosince 2015

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • ANESTH-03-15

Plán pro data jednotlivých účastníků (IPD)

Plánujete sdílet data jednotlivých účastníků (IPD)?

NE

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .

Předplatit