Videolaryngoscopes for Double Lumen Tube Intubations in Simulated Easy and Difficult Airway

24. december 2015 opdateret af: Imam Abdulrahman Bin Faisal University

A Comparison of Three Videolaryngoscopes for Double-lumen Tubes Intubation in Simulated Easy and Difficult Airways

The investigators hypothesized that the use of the King Vision™ and the Airtraq® VL would reduce the time to DLT intubation compared with the GlideScope® and Macintosh in simulated easy and difficult airways. The investigators have considered to assess the efficacy of each device in manikins before considering to evaluate them in patients undergoing thoracic procedures.

Twenty-one staff anaesthesiologists who had limited prior experience in using the VLs for DLT intubation participated in this randomised crossover study. Following a brief demonstration and two practice attempts, participants were volunteered to insert a DLT using the Macintosh, GlideScope®, Airtraq®, and King Vision™ on two high-fidelity easy and difficult airway simulators in a computer-generated randomized sequence. The primary endpoint, time to DLT intubation, as well as, the views obtained at laryngoscopy, ease of intubation, numbers of laryngoscopy attempts and optimisation manoeuvers, and failure to intubation; defined as an attempt took longer than 150 seconds, were recorded.

Studieoversigt

Detaljeret beskrivelse

Several regional surveys demonstrated that most thoracic anesthesiologists are using the double-lumen endobronchial tubes (DLT) as the first-choice lung separation technique. DLT, when compared with single lumen tracheal tube, can be more difficult to insert in patients with difficult airways. The videolaryngoscopes (VL) have the potential to facilitate the placement of the DLTs for lung separation in patients with predicted or unanticipated difficult airway.

The use of GlideScope® (Verathon Inc., Bothell, WA, USA), a VL with an angulated blade, has been associated with variable times to DLT intubation according to the prior experience of the operators, despite superior visualization of the glottis The channeled Airtraq® (Prodol Meditec S.A., Vizcaya, Spain) and standard non-channeled blade of the King Vision™ (Ambu, Ballerup, Copenhagen, Denmark), may offer additional benefits for DLT intubation in patients with limited mouth opening or restricted neck movement, in whom the use of traditional VL like as the Glidescope® could be difficult. This is because of the larger outer diameter, the distal curvature and the increased rigidity of the DLT. Of note, the longer times to DLT intubation with the use of different VL could be shortened with building up the operator's experience.

To the best of the investigators knowledge, the comparison of the effects of the Macintosh, GlideScope®, Airtraq®, and King Vision™ VL on the time to DLT intubation in simulated easy and difficult airways has not yet been studied.

Two high-fidelity simulators (Airway Management Trainer, model AA-3100, Laerdal, Kent, UK) were prepared to simulate easy and difficult airway situations, as described by Marshall et al. and Wang et al. The "easy" airway was established with the manikin in a neutral position. The "difficult" airway setting was obtained by placing an Oasis Elite™ Prone Head Rest, Adult (140 mm in height) (Covidien, Mansfield, MA) under the occiput and securing the head position with adhesive tape, object to replicate cervical-collar use. Positioning was confirmed after each attempt to ensure consistency.

Undersøgelsestype

Interventionel

Tilmelding (Faktiske)

21

Fase

  • Ikke anvendelig

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiesteder

    • EP
      • Al-Khobar, EP, Saudi Arabien, 31952
        • Dammam University KFHU

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

23 år til 65 år (Voksen, Ældre voksen)

Tager imod sunde frivillige

Ja

Køn, der er berettiget til at studere

Alle

Beskrivelse

Inclusion Criteria:

  • Staff anesthesiologists
  • Signed consent
  • Limited experience in using videolaryngoscopes for inserting double lumen tubes

Exclusion Criteria:

  • Decline to sign the consent
  • Non receiving the prior training demonstration
  • Non practicing once with each device tested on each simulator scenario

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Tildeling: Randomiseret
  • Interventionel model: Crossover opgave
  • Maskning: Ingen (Åben etiket)

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Placebo komparator: Macintosh
The participants intubated the easy airway simulator with a double lumen tube using the Macintosh laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the Macintosh laryngoscope followed with intubating the difficult airway simulator
Placebo komparator: GlideScope
The participants intubated the easy airway simulator with a double lumen tube using the GlideScope laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the GlideScope followed with intubating the difficult airway simulator
Aktiv komparator: Airtraq
The participants intubated the easy airway simulator with a double lumen tube using the Airtraq laryngoscope followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the Airtraq followed with intubating the difficult airway simulator
Aktiv komparator: King Vision
The participants intubated the easy airway simulator with a double lumen tube using the King Vision followed with intubating the difficult airway simulator
The participants intubated the easy airway simulator with a double lumen tube using the King Vision followed with intubating the difficult airway simulator

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Time to achieve successful double lumen tube intubation
Tidsramme: for 6 min after using each device
defined as the time when the investigated laryngoscope passed the central incisors to when the tip of the bronchial lumen passed through the glottis, as confirmed visually by the operator (in the Macintosh group) or by the investigator, thanks to remote screens (in the VLs groups).
for 6 min after using each device

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Best view during laryngoscopy using the classification described by Cormack and Lehane
Tidsramme: for 1 min after laryngoscopy
using the classification described by Cormack and Lehane
for 1 min after laryngoscopy
Difficulty of intubation using a Visual Analog Scale (VAS)
Tidsramme: for 6 min after using each device
The difficulty of intubation was evaluated using a visual analog scale (VAS) (ranging from 0: extremely easy, to 100: extremely difficult) expressed by the anesthesiologists after the DLT intubation
for 6 min after using each device
Number of first-pass success
Tidsramme: for 3 min after using each device
first-pass success calculated as number of first-attempt successes/number of patients with an intubation attempt
for 3 min after using each device
Number of the required optimisation manoeuvers
Tidsramme: for 6 min after using each device
When the anaesthesiologists encountered difficulty in visualising vocal cords or placing the tube, they were allowed to use any manoeuver they would normally use to navigate the tube into the trachea including readjustment of the blade or tube or to ask the supervising investigator to help solve the problem, and gave suggestions or instructions
for 6 min after using each device

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Efterforskere

  • Ledende efterforsker: Mohamed R El Tahan, MD, Anesthesiology Department, Dammam University, Dammam, Saudi Arabia
  • Studieleder: Abdulmohsen A Al Ghamdi, MD, Anesthesiology Department, Dammam University, Dammam, Saudi Arabia

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart

1. marts 2015

Primær færdiggørelse (Faktiske)

1. maj 2015

Studieafslutning (Faktiske)

1. juni 2015

Datoer for studieregistrering

Først indsendt

22. december 2015

Først indsendt, der opfyldte QC-kriterier

24. december 2015

Først opslået (Skøn)

28. december 2015

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Skøn)

28. december 2015

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

24. december 2015

Sidst verificeret

1. december 2015

Mere information

Begreber relateret til denne undersøgelse

Andre undersøgelses-id-numre

  • ANESTH-03-15

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

INGEN

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .

Abonner