Esta página se tradujo automáticamente y no se garantiza la precisión de la traducción. por favor refiérase a versión inglesa para un texto fuente.

Clinical Validation of a Blood Loss Monitoring System

31 de mayo de 2026 actualizado por: Zhiping Wang, Lanzhou University Second Hospital

Prospective Multicenter Clinical Validation of a Novel Multifunctional Intraoperative Blood Loss Monitoring System

This observational study aims to evaluate the practicality of a novel multifunctional intraoperative blood loss monitoring system in real-world surgical settings. The primary questions it seeks to address are:

Using the direct measurement method as the reference standard, what is the magnitude of measurement error of the monitoring system, and does it tend to overestimate or underestimate blood loss? Does the monitoring system demonstrate superior accuracy compared with visual estimation, the gravimetric method, and the hemoglobin concentration-based method? During surgery, intraoperative blood loss will be measured simultaneously using five approaches: the novel multifunctional monitoring system, the direct measurement method, visual estimation, the gravimetric method, and the hemoglobin concentration method. These measurements will be conducted without interfering with the surgical procedure.

Descripción general del estudio

Estado

Inscripción por invitación

Tipo de estudio

De observación

Inscripción (Estimado)

120

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Gansu
      • Lanzhou, Gansu, Porcelana, 730030
        • Lanzhou University Second Hospital
      • Lanzhou, Gansu, Porcelana, 730030
        • Lanzhou University Frist Hospital
      • Lanzhou, Gansu, Porcelana, 730030
        • Gansu Provincial People's Hospital

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Método de muestreo

Muestra no probabilística

Población de estudio

The study population will be recruited from adult patients undergoing elective urological and general surgical procedures at two tertiary hospitals in China, including Lanzhou University Second Hospital and Lanzhou University First Hospital. Participants will be selected from routine clinical practice settings where intraoperative blood loss monitoring is required. This population represents adult surgical patients managed under standard perioperative care, providing a real-world clinical context for evaluating intraoperative blood loss measurement methods.

Descripción

Inclusion Criteria:

  • Patients undergoing elective urological or general surgical procedures
  • Age 18 years or older
  • Preoperative hemoglobin concentration available
  • Intraoperative blood loss assessment required during surgery

Exclusion Criteria:

  • Patients with incomplete intraoperative blood loss measurement data,
  • Severe hemolysis or other conditions that may affect hemoglobin measurement
  • Emergency surgery
  • Intraoperative blood loss assessment required during surgery 24 hours after surgery
  • Refusal to participate, when applicable

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

Cohortes e Intervenciones

Grupo / Cohorte
Self-controlled

This study includes a single self-controlled cohort of patients undergoing urological and general surgical procedures. All participants will undergo simultaneous intraoperative blood loss assessment using five different methods within the same surgical procedure: a novel multifunctional intraoperative blood loss monitoring system, the direct measurement method (reference standard), visual estimation, the gravimetric method, and the hemoglobin concentration-based method.

No intervention is applied in this observational study. All measurements are performed according to a predefined protocol and do not interfere with the standard surgical workflow. The purpose of this design is to enable within-subject comparison of measurement accuracy across different methods under identical clinical conditions.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Absolute error of the automatic intraoperative blood loss monitor compared with the direct measurement method
Periodo de tiempo: At the end of surgery

Absolute error is defined as the numerical difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. A positive value indicates that the monitor overestimates blood loss compared with the direct measurement method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters (mL)

At the end of surgery

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Percentage error of the automatic intraoperative blood loss monitor compared with the direct measurement method
Periodo de tiempo: At the end of surgery

Percentage error is defined as the relative difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method, expressed as a percentage of the average value obtained by the two methods. A positive value indicates overestimation by the monitor, whereas a negative value indicates underestimation.

Unit of Measure: Percentage

At the end of surgery
Absolute error of visual estimation compared with the direct measurement method
Periodo de tiempo: At the end of surgery

Absolute error is defined as the numerical difference between intraoperative blood loss estimated by visual estimation and intraoperative blood loss measured by the direct measurement method. A positive value indicates overestimation by visual estimation, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

At the end of surgery
Absolute error of the gravimetric method compared with the direct measurement method
Periodo de tiempo: At the end of surgery

The accuracy of five methods (monitoring system, direct measurement, visual Absolute error is defined as the numerical difference between intraoperative blood loss estimated by the gravimetric method and intraoperative blood loss measured by the direct measurement method. A positive value indicates overestimation by the gravimetric method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

At the end of surgery
Absolute error of the hemoglobin concentration-based method compared with the direct measurement method
Periodo de tiempo: 24 hours after surgery

Absolute error is defined as the numerical difference between blood loss estimated by the hemoglobin concentration-based method and intraoperative blood loss measured by the direct measurement method. The hemoglobin concentration-based estimate is calculated using preoperative hemoglobin concentration, hemoglobin concentration measured 24 hours after surgery, and estimated blood volume. A positive value indicates overestimation by the hemoglobin concentration-based method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

24 hours after surgery
Intraclass correlation coefficient between the automatic intraoperative blood loss monitor and the direct measurement method
Periodo de tiempo: At the end of surgery

The intraclass correlation coefficient is used to quantify agreement between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. Higher values indicate stronger agreement between the two methods.

Unit of Measure: Unitless coefficient

At the end of surgery
Bland-Altman mean bias between the automatic intraoperative blood loss monitor and the direct measurement method
Periodo de tiempo: Unit of Measure: Milliliters
Bland-Altman mean bias is defined as the mean difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. A positive mean bias indicates systematic overestimation by the monitor, whereas a negative mean bias indicates systematic underestimation.
Unit of Measure: Milliliters
Bland-Altman limits of agreement between the automatic intraoperative blood loss monitor and the direct measurement method
Periodo de tiempo: At the end of surgery

Bland-Altman limits of agreement describe the range within which most differences between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method are expected to fall.

Unit of Measure: Milliliters

At the end of surgery

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

10 de septiembre de 2025

Finalización primaria (Estimado)

30 de mayo de 2026

Finalización del estudio (Estimado)

30 de mayo de 2026

Fechas de registro del estudio

Enviado por primera vez

5 de mayo de 2026

Primero enviado que cumplió con los criterios de control de calidad

23 de mayo de 2026

Publicado por primera vez (Actual)

1 de junio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

3 de junio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

31 de mayo de 2026

Última verificación

1 de abril de 2026

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • 2025A-987

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

INDECISO

Descripción del plan IPD

The clinical data involves sensitive patient privacy. We are currently undecided on sharing the IPD, as it requires specific informed consent from the patients and authorization from the hospital's ethics committee.

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir