Ta strona została przetłumaczona automatycznie i dokładność tłumaczenia nie jest gwarantowana. Proszę odnieść się do angielska wersja za tekst źródłowy.

Clinical Validation of a Blood Loss Monitoring System

31 maja 2026 zaktualizowane przez: Zhiping Wang, Lanzhou University Second Hospital

Prospective Multicenter Clinical Validation of a Novel Multifunctional Intraoperative Blood Loss Monitoring System

This observational study aims to evaluate the practicality of a novel multifunctional intraoperative blood loss monitoring system in real-world surgical settings. The primary questions it seeks to address are:

Using the direct measurement method as the reference standard, what is the magnitude of measurement error of the monitoring system, and does it tend to overestimate or underestimate blood loss? Does the monitoring system demonstrate superior accuracy compared with visual estimation, the gravimetric method, and the hemoglobin concentration-based method? During surgery, intraoperative blood loss will be measured simultaneously using five approaches: the novel multifunctional monitoring system, the direct measurement method, visual estimation, the gravimetric method, and the hemoglobin concentration method. These measurements will be conducted without interfering with the surgical procedure.

Przegląd badań

Status

Rejestracja na zaproszenie

Typ studiów

Obserwacyjny

Zapisy (Szacowany)

120

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Lokalizacje studiów

    • Gansu
      • Lanzhou, Gansu, Chiny, 730030
        • Lanzhou University Second Hospital
      • Lanzhou, Gansu, Chiny, 730030
        • Lanzhou University Frist Hospital
      • Lanzhou, Gansu, Chiny, 730030
        • Gansu Provincial People's Hospital

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

  • Dorosły
  • Starszy dorosły

Akceptuje zdrowych ochotników

Nie

Metoda próbkowania

Próbka bez prawdopodobieństwa

Badana populacja

The study population will be recruited from adult patients undergoing elective urological and general surgical procedures at two tertiary hospitals in China, including Lanzhou University Second Hospital and Lanzhou University First Hospital. Participants will be selected from routine clinical practice settings where intraoperative blood loss monitoring is required. This population represents adult surgical patients managed under standard perioperative care, providing a real-world clinical context for evaluating intraoperative blood loss measurement methods.

Opis

Inclusion Criteria:

  • Patients undergoing elective urological or general surgical procedures
  • Age 18 years or older
  • Preoperative hemoglobin concentration available
  • Intraoperative blood loss assessment required during surgery

Exclusion Criteria:

  • Patients with incomplete intraoperative blood loss measurement data,
  • Severe hemolysis or other conditions that may affect hemoglobin measurement
  • Emergency surgery
  • Intraoperative blood loss assessment required during surgery 24 hours after surgery
  • Refusal to participate, when applicable

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

Kohorty i interwencje

Grupa / Kohorta
Self-controlled

This study includes a single self-controlled cohort of patients undergoing urological and general surgical procedures. All participants will undergo simultaneous intraoperative blood loss assessment using five different methods within the same surgical procedure: a novel multifunctional intraoperative blood loss monitoring system, the direct measurement method (reference standard), visual estimation, the gravimetric method, and the hemoglobin concentration-based method.

No intervention is applied in this observational study. All measurements are performed according to a predefined protocol and do not interfere with the standard surgical workflow. The purpose of this design is to enable within-subject comparison of measurement accuracy across different methods under identical clinical conditions.

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Absolute error of the automatic intraoperative blood loss monitor compared with the direct measurement method
Ramy czasowe: At the end of surgery

Absolute error is defined as the numerical difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. A positive value indicates that the monitor overestimates blood loss compared with the direct measurement method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters (mL)

At the end of surgery

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Percentage error of the automatic intraoperative blood loss monitor compared with the direct measurement method
Ramy czasowe: At the end of surgery

Percentage error is defined as the relative difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method, expressed as a percentage of the average value obtained by the two methods. A positive value indicates overestimation by the monitor, whereas a negative value indicates underestimation.

Unit of Measure: Percentage

At the end of surgery
Absolute error of visual estimation compared with the direct measurement method
Ramy czasowe: At the end of surgery

Absolute error is defined as the numerical difference between intraoperative blood loss estimated by visual estimation and intraoperative blood loss measured by the direct measurement method. A positive value indicates overestimation by visual estimation, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

At the end of surgery
Absolute error of the gravimetric method compared with the direct measurement method
Ramy czasowe: At the end of surgery

The accuracy of five methods (monitoring system, direct measurement, visual Absolute error is defined as the numerical difference between intraoperative blood loss estimated by the gravimetric method and intraoperative blood loss measured by the direct measurement method. A positive value indicates overestimation by the gravimetric method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

At the end of surgery
Absolute error of the hemoglobin concentration-based method compared with the direct measurement method
Ramy czasowe: 24 hours after surgery

Absolute error is defined as the numerical difference between blood loss estimated by the hemoglobin concentration-based method and intraoperative blood loss measured by the direct measurement method. The hemoglobin concentration-based estimate is calculated using preoperative hemoglobin concentration, hemoglobin concentration measured 24 hours after surgery, and estimated blood volume. A positive value indicates overestimation by the hemoglobin concentration-based method, whereas a negative value indicates underestimation.

Unit of Measure: Milliliters

24 hours after surgery
Intraclass correlation coefficient between the automatic intraoperative blood loss monitor and the direct measurement method
Ramy czasowe: At the end of surgery

The intraclass correlation coefficient is used to quantify agreement between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. Higher values indicate stronger agreement between the two methods.

Unit of Measure: Unitless coefficient

At the end of surgery
Bland-Altman mean bias between the automatic intraoperative blood loss monitor and the direct measurement method
Ramy czasowe: Unit of Measure: Milliliters
Bland-Altman mean bias is defined as the mean difference between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method. A positive mean bias indicates systematic overestimation by the monitor, whereas a negative mean bias indicates systematic underestimation.
Unit of Measure: Milliliters
Bland-Altman limits of agreement between the automatic intraoperative blood loss monitor and the direct measurement method
Ramy czasowe: At the end of surgery

Bland-Altman limits of agreement describe the range within which most differences between intraoperative blood loss measured by the automatic intraoperative blood loss monitor and intraoperative blood loss measured by the direct measurement method are expected to fall.

Unit of Measure: Milliliters

At the end of surgery

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Rzeczywisty)

10 września 2025

Zakończenie podstawowe (Szacowany)

30 maja 2026

Ukończenie studiów (Szacowany)

30 maja 2026

Daty rejestracji na studia

Pierwszy przesłany

5 maja 2026

Pierwszy przesłany, który spełnia kryteria kontroli jakości

23 maja 2026

Pierwszy wysłany (Rzeczywisty)

1 czerwca 2026

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

3 czerwca 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

31 maja 2026

Ostatnia weryfikacja

1 kwietnia 2026

Więcej informacji

Terminy związane z tym badaniem

Inne numery identyfikacyjne badania

  • 2025A-987

Plan dla danych uczestnika indywidualnego (IPD)

Planujesz udostępniać dane poszczególnych uczestników (IPD)?

NIEZDECYDOWANY

Opis planu IPD

The clinical data involves sensitive patient privacy. We are currently undecided on sharing the IPD, as it requires specific informed consent from the patients and authorization from the hospital's ethics committee.

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Nie

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .

Badania kliniczne na Śródoperacyjna utrata krwi

Subskrybuj