Volumetric Visual Fields at Intermediate and Near Distances in Presbyopic Patients Using Occupational and General-Purpose Progressive Addition Lenses (CIVIUS2025PAL)

lauantai 23. toukokuuta 2026 päivittänyt: José Manuel Serrano Morales, University of Seville

Ensayo clínico, unicéntrico, Transversal, Randomizado, Cruzado, Doble Ciego Para Evaluar Las Diferencias de Los Campos volumétricos de visión en Distancia Intermedia (4m) y Cerca (25cm) de Tres Tipos de Lentes oftálmicas de adición Progresiva (PAL), Dos de visión Intermedia y Cerca (PC-PAL) y Otra Generalista de Vision Lejos, Intermedia y Cerca (GPPAL), Neutros en Vision de Lejos y de 2.00 D en Vision de Cerca en Pacientes présbitas.

This single-center, randomized, double-blind, crossover clinical trial aims to compare the volumetric visual fields provided by two occupational progressive addition lenses (OPLs) optimized for intermediate and near vision in presbyopic patients. The study evaluates Vimax Zoom 1.5 FWD 4 m and Vimax PC 1.5 FWD 4 m lenses and Vimax Fit FRM, all of them with plano distance power and a +2.00 D near addition, in order to determine differences in the extent of clear vision at multiple working distances.

Participants will undergo a comprehensive optometric examination including medical and ocular history, visual acuity assessment, accommodative amplitude measurement, autorefraction, interpupillary distance measurement, ocular motility evaluation, confrontation visual fields, Amsler grid testing, and binocular vision assessment. Visual field measurements will then be performed using both study lenses under standardized conditions. Subjects will be instructed to maintain a fixed head position while identifying optotypes located at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. Both lenses will be tested in random order under masked conditions so that neither participants nor investigators know which lens design is being evaluated during each phase.

At least 23 presbyopic participants are expected to be enrolled. The study procedures are non-invasive and are anticipated to be completed in a single one-hour visit, although additional sessions may be scheduled if necessary. No significant risks are expected beyond possible transient visual fatigue associated with the duration of testing.

The primary objective is to determine which occupational progressive lens design provides a greater spatial volume of clear vision and wider horizontal and vertical visual fields at different intermediate and near distances. The findings may contribute to improving personalized recommendations of occupational lenses according to patients' visual demands and working distances.

Tutkimuksen yleiskatsaus

Opintotyyppi

Interventio

Ilmoittautuminen (Todellinen)

26

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskelupaikat

    • Sevilla
      • Seville, Sevilla, Espanja, 41012
        • Pharmacy school, University of Seville

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

  • Aikuinen
  • Vanhempi Aikuinen

Hyväksyy terveitä vapaaehtoisia

Joo

Kuvaus

Inclusion Criteria:

  • Spherical ametropia for distance vision ≤ ±4.00 D and astigmatic ametropia ≤ 1.00 D.
  • Decimal VA ≥ 0.8, either uncorrected (UCVA) or corrected (BCVA) with monofocal contact lenses.
  • Patients with a nasopupillary distance (NPD) of 31 mm, with a tolerance of ±2 mm in each eye

Exclusion Criteria:

  • Astigmatism greater than ±1.00 D.
  • Patients wearing progressive contact lenses.
  • Patients who have undergone cataract surgery or have monofocal or multifocal intraocular lenses.
  • Anisometropia greater than 2.50 D.
  • Decimal distance visual acuity < 0.8 (UCVA or BCVA) with monofocal contact lenses.
  • Patients without normal ocular motility or with partial dysfunction of one or more ex-traocular muscles, or with restricted visual fields or scotomas.
  • Visual suppression in one eye between 4 m and 25 cm.
  • Strabismus or nystagmus.

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Hoito
  • Jako: Satunnaistettu
  • Inventiomalli: Crossover-tehtävä
  • Naamiointi: Kaksinkertainen

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: OPL designed for PC use
VIMAX PC: Occupational progressive lenses designed to provide clear vision from 400 cm to 40 cm, with a wider visual field optimized for monitor-based environments with a depth of field between 60 and 80 cm. These lenses, designated L1, were custom ordered with neutral distance power and a near addition of +2.00 D, a vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in each eye.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Kokeellinen: OPL designed for office use
VIMAX ZOOM: Occupational progressive lenses also designed to provide clear vision from 400 cm to 40 cm, but with a more homogeneous distribution of the visual field across office and home environments, with a depth of field between 40 and 400 cm. These lenses, designated L2, were also custom ordered with neutral distance power and a near addition of +2.00 D, vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Active Comparator: PAL
VIMAX FIT: Progressive additional lenses custom ordered with neutral distance power and a near addition of +2.00 D, vertex dis-tance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Visual field volume
Aikaikkuna: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
It will consist of the volume defined by two parallel surfaces bounded by the visual fields perceived with clear horizontal and vertical vision by the patient between the distances of 4 m and 25 cm while wearing each pair of spectacles. The volume will be calculated using the Heronian formula for the volume of a quadrangular pyramidal frustum.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Visual fields surface
Aikaikkuna: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
Calculation of the visual field surface areas defined by the patient with each lens type at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. For this purpose, the areas corresponding to the visual field regions of the lens located between the fitting cross and the intermediate and near viewing zones were summed in a 1:2 proportion, respectively, for each measured distance.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Maanantai 29. syyskuuta 2025

Ensisijainen valmistuminen (Todellinen)

Perjantai 28. marraskuuta 2025

Opintojen valmistuminen (Arvioitu)

Tiistai 1. syyskuuta 2026

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Torstai 14. toukokuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Lauantai 23. toukokuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Maanantai 1. kesäkuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Maanantai 1. kesäkuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Lauantai 23. toukokuuta 2026

Viimeksi vahvistettu

Perjantai 1. toukokuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muita asiaankuuluvia MeSH-ehtoja

Muut tutkimustunnusnumerot

  • CIVIUS2025PAL
  • SICEIA-2025-001629 (Muu tunniste: SICEIA: Ethics Committee for Biomedics Research of the Junta de Andalucia)

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

EI

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Yhdysvalloissa valmistettu ja sieltä viety tuote

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .

Tilaa