Ezt az oldalt automatikusan lefordították, és a fordítás pontossága nem garantált. Kérjük, olvassa el a angol verzió forrásszöveghez.

Volumetric Visual Fields at Intermediate and Near Distances in Presbyopic Patients Using Occupational and General-Purpose Progressive Addition Lenses (CIVIUS2025PAL)

2026. május 23. frissítette: José Manuel Serrano Morales, University of Seville

Ensayo clínico, unicéntrico, Transversal, Randomizado, Cruzado, Doble Ciego Para Evaluar Las Diferencias de Los Campos volumétricos de visión en Distancia Intermedia (4m) y Cerca (25cm) de Tres Tipos de Lentes oftálmicas de adición Progresiva (PAL), Dos de visión Intermedia y Cerca (PC-PAL) y Otra Generalista de Vision Lejos, Intermedia y Cerca (GPPAL), Neutros en Vision de Lejos y de 2.00 D en Vision de Cerca en Pacientes présbitas.

This single-center, randomized, double-blind, crossover clinical trial aims to compare the volumetric visual fields provided by two occupational progressive addition lenses (OPLs) optimized for intermediate and near vision in presbyopic patients. The study evaluates Vimax Zoom 1.5 FWD 4 m and Vimax PC 1.5 FWD 4 m lenses and Vimax Fit FRM, all of them with plano distance power and a +2.00 D near addition, in order to determine differences in the extent of clear vision at multiple working distances.

Participants will undergo a comprehensive optometric examination including medical and ocular history, visual acuity assessment, accommodative amplitude measurement, autorefraction, interpupillary distance measurement, ocular motility evaluation, confrontation visual fields, Amsler grid testing, and binocular vision assessment. Visual field measurements will then be performed using both study lenses under standardized conditions. Subjects will be instructed to maintain a fixed head position while identifying optotypes located at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. Both lenses will be tested in random order under masked conditions so that neither participants nor investigators know which lens design is being evaluated during each phase.

At least 23 presbyopic participants are expected to be enrolled. The study procedures are non-invasive and are anticipated to be completed in a single one-hour visit, although additional sessions may be scheduled if necessary. No significant risks are expected beyond possible transient visual fatigue associated with the duration of testing.

The primary objective is to determine which occupational progressive lens design provides a greater spatial volume of clear vision and wider horizontal and vertical visual fields at different intermediate and near distances. The findings may contribute to improving personalized recommendations of occupational lenses according to patients' visual demands and working distances.

A tanulmány áttekintése

Állapot

Aktív, nem toborzó

Körülmények

Tanulmány típusa

Beavatkozó

Beiratkozás (Tényleges)

26

Fázis

  • Nem alkalmazható

Kapcsolatok és helyek

Ez a rész a vizsgálatot végzők elérhetőségeit, valamint a vizsgálat lefolytatásának helyére vonatkozó információkat tartalmazza.

Tanulmányi helyek

    • Sevilla
      • Seville, Sevilla, Spanyolország, 41012
        • Pharmacy school, University of Seville

Részvételi kritériumok

A kutatók olyan embereket keresnek, akik megfelelnek egy bizonyos leírásnak, az úgynevezett jogosultsági kritériumoknak. Néhány példa ezekre a kritériumokra a személy általános egészségi állapota vagy a korábbi kezelések.

Jogosultsági kritériumok

Tanulmányozható életkorok

  • Felnőtt
  • Idősebb felnőtt

Egészséges önkénteseket fogad

Igen

Leírás

Inclusion Criteria:

  • Spherical ametropia for distance vision ≤ ±4.00 D and astigmatic ametropia ≤ 1.00 D.
  • Decimal VA ≥ 0.8, either uncorrected (UCVA) or corrected (BCVA) with monofocal contact lenses.
  • Patients with a nasopupillary distance (NPD) of 31 mm, with a tolerance of ±2 mm in each eye

Exclusion Criteria:

  • Astigmatism greater than ±1.00 D.
  • Patients wearing progressive contact lenses.
  • Patients who have undergone cataract surgery or have monofocal or multifocal intraocular lenses.
  • Anisometropia greater than 2.50 D.
  • Decimal distance visual acuity < 0.8 (UCVA or BCVA) with monofocal contact lenses.
  • Patients without normal ocular motility or with partial dysfunction of one or more ex-traocular muscles, or with restricted visual fields or scotomas.
  • Visual suppression in one eye between 4 m and 25 cm.
  • Strabismus or nystagmus.

Tanulási terv

Ez a rész a vizsgálati terv részleteit tartalmazza, beleértve a vizsgálat megtervezését és a vizsgálat mérését.

Hogyan készül a tanulmány?

Tervezési részletek

  • Elsődleges cél: Kezelés
  • Kiosztás: Véletlenszerűsített
  • Beavatkozó modell: Crossover kiosztás
  • Maszkolás: Kettős

Fegyverek és beavatkozások

Résztvevő csoport / kar
Beavatkozás / kezelés
Kísérleti: OPL designed for PC use
VIMAX PC: Occupational progressive lenses designed to provide clear vision from 400 cm to 40 cm, with a wider visual field optimized for monitor-based environments with a depth of field between 60 and 80 cm. These lenses, designated L1, were custom ordered with neutral distance power and a near addition of +2.00 D, a vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in each eye.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Kísérleti: OPL designed for office use
VIMAX ZOOM: Occupational progressive lenses also designed to provide clear vision from 400 cm to 40 cm, but with a more homogeneous distribution of the visual field across office and home environments, with a depth of field between 40 and 400 cm. These lenses, designated L2, were also custom ordered with neutral distance power and a near addition of +2.00 D, vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Aktív összehasonlító: PAL
VIMAX FIT: Progressive additional lenses custom ordered with neutral distance power and a near addition of +2.00 D, vertex dis-tance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Mit mér a tanulmány?

Elsődleges eredményintézkedések

Eredménymérő
Intézkedés leírása
Időkeret
Visual field volume
Időkeret: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
It will consist of the volume defined by two parallel surfaces bounded by the visual fields perceived with clear horizontal and vertical vision by the patient between the distances of 4 m and 25 cm while wearing each pair of spectacles. The volume will be calculated using the Heronian formula for the volume of a quadrangular pyramidal frustum.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Másodlagos eredményintézkedések

Eredménymérő
Intézkedés leírása
Időkeret
Visual fields surface
Időkeret: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
Calculation of the visual field surface areas defined by the patient with each lens type at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. For this purpose, the areas corresponding to the visual field regions of the lens located between the fitting cross and the intermediate and near viewing zones were summed in a 1:2 proportion, respectively, for each measured distance.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Együttműködők és nyomozók

Itt találhatja meg a tanulmányban érintett személyeket és szervezeteket.

Tanulmányi rekorddátumok

Ezek a dátumok nyomon követik a ClinicalTrials.gov webhelyre benyújtott vizsgálati rekordok és összefoglaló eredmények benyújtásának folyamatát. A vizsgálati feljegyzéseket és a jelentett eredményeket a Nemzeti Orvostudományi Könyvtár (NLM) felülvizsgálja, hogy megbizonyosodjon arról, hogy megfelelnek-e az adott minőség-ellenőrzési szabványoknak, mielőtt közzéteszik őket a nyilvános weboldalon.

Tanulmány főbb dátumok

Tanulmány kezdete (Tényleges)

2025. szeptember 29.

Elsődleges befejezés (Tényleges)

2025. november 28.

A tanulmány befejezése (Becsült)

2026. szeptember 1.

Tanulmányi regisztráció dátumai

Először benyújtva

2026. május 14.

Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak

2026. május 23.

Első közzététel (Tényleges)

2026. június 1.

Tanulmányi rekordok frissítései

Utolsó frissítés közzétéve (Tényleges)

2026. június 1.

Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak

2026. május 23.

Utolsó ellenőrzés

2026. május 1.

Több információ

A tanulmányhoz kapcsolódó kifejezések

További vonatkozó MeSH feltételek

Egyéb vizsgálati azonosító számok

  • CIVIUS2025PAL
  • SICEIA-2025-001629 (Egyéb azonosító: SICEIA: Ethics Committee for Biomedics Research of the Junta de Andalucia)

Terv az egyéni résztvevői adatokhoz (IPD)

Tervezi megosztani az egyéni résztvevői adatokat (IPD)?

NEM

Gyógyszer- és eszközinformációk, tanulmányi dokumentumok

Egy amerikai FDA által szabályozott gyógyszerkészítményt tanulmányoz

Nem

Egy amerikai FDA által szabályozott eszközterméket tanulmányoz

Nem

az Egyesült Államokban gyártott és onnan exportált termék

Nem

Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .

Iratkozz fel