Cette page a été traduite automatiquement et l'exactitude de la traduction n'est pas garantie. Veuillez vous référer au version anglaise pour un texte source.

Volumetric Visual Fields at Intermediate and Near Distances in Presbyopic Patients Using Occupational and General-Purpose Progressive Addition Lenses (CIVIUS2025PAL)

23 mai 2026 mis à jour par: José Manuel Serrano Morales, University of Seville

Ensayo clínico, unicéntrico, Transversal, Randomizado, Cruzado, Doble Ciego Para Evaluar Las Diferencias de Los Campos volumétricos de visión en Distancia Intermedia (4m) y Cerca (25cm) de Tres Tipos de Lentes oftálmicas de adición Progresiva (PAL), Dos de visión Intermedia y Cerca (PC-PAL) y Otra Generalista de Vision Lejos, Intermedia y Cerca (GPPAL), Neutros en Vision de Lejos y de 2.00 D en Vision de Cerca en Pacientes présbitas.

This single-center, randomized, double-blind, crossover clinical trial aims to compare the volumetric visual fields provided by two occupational progressive addition lenses (OPLs) optimized for intermediate and near vision in presbyopic patients. The study evaluates Vimax Zoom 1.5 FWD 4 m and Vimax PC 1.5 FWD 4 m lenses and Vimax Fit FRM, all of them with plano distance power and a +2.00 D near addition, in order to determine differences in the extent of clear vision at multiple working distances.

Participants will undergo a comprehensive optometric examination including medical and ocular history, visual acuity assessment, accommodative amplitude measurement, autorefraction, interpupillary distance measurement, ocular motility evaluation, confrontation visual fields, Amsler grid testing, and binocular vision assessment. Visual field measurements will then be performed using both study lenses under standardized conditions. Subjects will be instructed to maintain a fixed head position while identifying optotypes located at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. Both lenses will be tested in random order under masked conditions so that neither participants nor investigators know which lens design is being evaluated during each phase.

At least 23 presbyopic participants are expected to be enrolled. The study procedures are non-invasive and are anticipated to be completed in a single one-hour visit, although additional sessions may be scheduled if necessary. No significant risks are expected beyond possible transient visual fatigue associated with the duration of testing.

The primary objective is to determine which occupational progressive lens design provides a greater spatial volume of clear vision and wider horizontal and vertical visual fields at different intermediate and near distances. The findings may contribute to improving personalized recommendations of occupational lenses according to patients' visual demands and working distances.

Aperçu de l'étude

Statut

Actif, ne recrute pas

Les conditions

Type d'étude

Interventionnel

Inscription (Réel)

26

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

    • Sevilla
      • Seville, Sevilla, Espagne, 41012
        • Pharmacy school, University of Seville

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte
  • Adulte plus âgé

Accepte les volontaires sains

Oui

La description

Inclusion Criteria:

  • Spherical ametropia for distance vision ≤ ±4.00 D and astigmatic ametropia ≤ 1.00 D.
  • Decimal VA ≥ 0.8, either uncorrected (UCVA) or corrected (BCVA) with monofocal contact lenses.
  • Patients with a nasopupillary distance (NPD) of 31 mm, with a tolerance of ±2 mm in each eye

Exclusion Criteria:

  • Astigmatism greater than ±1.00 D.
  • Patients wearing progressive contact lenses.
  • Patients who have undergone cataract surgery or have monofocal or multifocal intraocular lenses.
  • Anisometropia greater than 2.50 D.
  • Decimal distance visual acuity < 0.8 (UCVA or BCVA) with monofocal contact lenses.
  • Patients without normal ocular motility or with partial dysfunction of one or more ex-traocular muscles, or with restricted visual fields or scotomas.
  • Visual suppression in one eye between 4 m and 25 cm.
  • Strabismus or nystagmus.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation croisée
  • Masquage: Double

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: OPL designed for PC use
VIMAX PC: Occupational progressive lenses designed to provide clear vision from 400 cm to 40 cm, with a wider visual field optimized for monitor-based environments with a depth of field between 60 and 80 cm. These lenses, designated L1, were custom ordered with neutral distance power and a near addition of +2.00 D, a vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in each eye.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Expérimental: OPL designed for office use
VIMAX ZOOM: Occupational progressive lenses also designed to provide clear vision from 400 cm to 40 cm, but with a more homogeneous distribution of the visual field across office and home environments, with a depth of field between 40 and 400 cm. These lenses, designated L2, were also custom ordered with neutral distance power and a near addition of +2.00 D, vertex distance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Comparateur actif: PAL
VIMAX FIT: Progressive additional lenses custom ordered with neutral distance power and a near addition of +2.00 D, vertex dis-tance of 13 mm, pantoscopic tilt of 8°, wrap angle of 6°, working distance of 40 cm, inset of 2.5 mm in both eyes, and NPD of 31 mm in both eyes.

Fitting and positioning of the spectacles : participants were instructed to move only their eyes and to verbally identify the letters indicated by the optometrist.

Measurement of clear vision limits: the horizontal and vertical boundaries of clear vision were recorded on a scaled results sheet, as defined by the participant using the optotype, at distances of 400, 200, 133, 80, 40, and 25 cm, in that order

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Visual field volume
Délai: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
It will consist of the volume defined by two parallel surfaces bounded by the visual fields perceived with clear horizontal and vertical vision by the patient between the distances of 4 m and 25 cm while wearing each pair of spectacles. The volume will be calculated using the Heronian formula for the volume of a quadrangular pyramidal frustum.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Visual fields surface
Délai: Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.
Calculation of the visual field surface areas defined by the patient with each lens type at distances of 4 m, 2 m, 1.33 m, 80 cm, 40 cm, and 25 cm. For this purpose, the areas corresponding to the visual field regions of the lens located between the fitting cross and the intermediate and near viewing zones were summed in a 1:2 proportion, respectively, for each measured distance.
Immediately after the adjustment of the frame where the lenses were placed, an average of 30 minutes per patient.

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

29 septembre 2025

Achèvement primaire (Réel)

28 novembre 2025

Achèvement de l'étude (Estimé)

1 septembre 2026

Dates d'inscription aux études

Première soumission

14 mai 2026

Première soumission répondant aux critères de contrôle qualité

23 mai 2026

Première publication (Réel)

1 juin 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

1 juin 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

23 mai 2026

Dernière vérification

1 mai 2026

Plus d'information

Termes liés à cette étude

Termes MeSH pertinents supplémentaires

Autres numéros d'identification d'étude

  • CIVIUS2025PAL
  • SICEIA-2025-001629 (Autre identifiant: SICEIA: Ethics Committee for Biomedics Research of the Junta de Andalucia)

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner