4-electrode ECG (DigitalCardio®) vs Standard 10-electrode ECG: Evaluating Ease of Use and Recording Quality (FAST-ECG)

keskiviikko 9. syyskuuta 2026 päivittänyt: Clinique Bizet

The goal of this clinical trial is to compare the time required to record a complete Electrocardiogram (ECG) trace using the DigitalCardio (CardioSecur®) tablet device versus a standard 12-lead ECG device in eligible patients.

The main questions it aims to answer are:

Is there a difference in diagnostic recording time (in minutes) between the DigitalCardio tablet and a standard ECG machine? How do the two devices compare in terms of trace quality (including QRS duration, QT interval, frontal axis measurements, and missing/erroneous leads)? How satisfied are healthcare professionals with the CardioSecur® device?

Researchers will compare Group A (standard 10-electrode ECG followed by 4-electrode ECG) to Group B (4-electrode ECG followed by standard 10-electrode ECG) in a 1:1 randomized crossover design to evaluate recording efficiency and trace quality.

Participants will:

Complete an initial screening visit with the principal investigator to review eligibility criteria and sign the informed consent form.

Be randomly assigned (1:1 ratio) to either Group A or Group B. Undergo both ECG recording methods sequentially (a standard 10-electrode ECG and a 4-electrode ECG using the DigitalCardio tablet), while researchers measure the total time elapsed from electrode placement to obtaining a complete trace.

Tutkimuksen yleiskatsaus

Yksityiskohtainen kuvaus

This study is a randomized, crossover clinical trial designed to compare the efficiency, trace quality, and user satisfaction of a novel 4-electrode tablet-based Electrocardiogram (ECG) system (DigitalCardio / CardioSecur®) against a standard 12-lead (10-electrode) ECG device.

Background and Rationale

An electrocardiogram (ECG) is a fundamental diagnostic tool used to record the electrical activity of the heart. Conventional 12-lead ECGs require placing 10 individual electrodes on a patient's limbs and chest while the patient is lying down. While highly detailed, electrode placement and device setup can be time-consuming in clinical workflows. The DigitalCardio tablet system utilizes a simplified 4-electrode setup designed to streamline data acquisition without compromising diagnostic accuracy.

Study Design and Procedures

After an initial screening process led by the principal investigator, eligible participants receive detailed study information and provide written informed consent. Enrolled participants are then randomly assigned in a 1:1 ratio to one of two sequence groups to control for order effects:

*Group A:Undergoes standard 10-electrode ECG acquisition first, followed immediately by 4-electrode DigitalCardio acquisition.

Group B: Undergoes 4-electrode DigitalCardio acquisition first, followed immediately by standard 10-electrode ECG acquisition.

To maintain consistency, both ECG recordings for a given participant are performed by the same healthcare operator within a minimal time interval.

Key Measurements

  1. Diagnostic Time:Research staff record the exact time elapsed (in minutes) from the initiation of electrode placement to the successful capture of a complete ECG trace for both devices.
  2. Trace Quality and Concordance:Printed and anonymized ECG traces are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads.
  3. Healthcare Professional Satisfaction:Participating healthcare professionals complete a structured questionnaire evaluating user satisfaction and usability of the CardioSecur® device in routine practice.

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

50

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

Opiskelupaikat

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

  • Aikuinen
  • Vanhempi Aikuinen

Hyväksyy terveitä vapaaehtoisia

Joo

Kuvaus

Inclusion Criteria:

Hospitalized Cardiology Patients:

  • Age minimum 18 years
  • Hospitalized for an acute or chronic cardiac condition (e.g., heart failure, acute coronary syndrome, arrhythmia, valvular heart disease, etc.)
  • Ability to give informed consent
  • Hemodynamic stability permitting the performance of an ECG without clinical risk

Cardiology Outpatients:

  • Age minimum of 18 years
  • Follow-up for a known cardiac condition or suspected cardiovascular disease
  • No contraindications to performing an ECG
  • Informed consent obtained

Exclusion Criteria:

  • Minor patient (< 18 years of age)
  • Adult subject under legal protection, under temporary guardianship, curatorship, or tutelage
  • Patient refusing to participate in the study
  • Presence of an implantable cardiac device (e.g., pacemaker, defibrillator) that may interfere with ECG measurements
  • Severe skin condition preventing the placement of electrodes
  • Physical or cognitive inability to follow the study protocol
  • Current pregnancy

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Tukevaa hoitoa
  • Jako: Satunnaistettu
  • Inventiomalli: Crossover-tehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Muut: standard 10-electrode ECG
Undergoes standard 10-electrode ECG acquisition first
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.
Active Comparator: 4-electrode DigitalCardio
Undergoes 4-electrode DigitalCardio
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Comparison of Procedure Time Between 4-Electrode DigitalCardio® ECG and Standard 12-Lead ECG
Aikaikkuna: 1 day

Time elapsed from the initiation of electrode placement to the obtaining of a complete ECG trace for both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Minutes

1 day

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Comparative Quality Assessment of ECG Traces
Aikaikkuna: 1 day

Proportion of paired recordings in which an independent cardiologist prefers the 4-electrode DigitalCardio® trace compared to the standard 12-lead ECG trace, based on visual evaluation of signal clarity and the presence of artifacts/interference.

Unit of Measure: Percentage of paired recordings

1 day
Intraclass Correlation Coefficient (ICC) for Quantitative ECG Parameters
Aikaikkuna: 1 day

Degree of agreement for quantitative clinical parameters (e.g., QRS duration, QT interval, frontal axis measurements) automatically provided by both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Intraclass Correlation Coefficient (ICC) (ranging from 0 to 1, where higher values indicate better agreement)

1 day
Diagnostic Concordance for Cardiac Pathology
Aikaikkuna: 1 day

Agreement in detecting the presence or absence of cardiac pathology using the 4-electrode DigitalCardio® system compared to the standard 12-lead ECG (gold standard).

Unit of Measure: Percentage of concordant diagnostic evaluations

1 day
Healthcare Professional Satisfaction Score
Aikaikkuna: 1 days

Overall user satisfaction regarding the ease of use of the DigitalCardio® tablet device, evaluated by participating healthcare professionals using a structured Likert scale. The score ranges from 1 (strongly disagree / highly unsatisfied) to 5 (strongly agree / highly satisfied), where higher scores indicate a better outcome (higher satisfaction).

Unit of Measure: Score on a scale

1 days
Incidence of Device-Related Technical Incidents for 4-electrode DigitalCardio® system
Aikaikkuna: 1 day

Total number of technical malfunctions, failures, or operational incidents reported for 4-electrode DigitalCardio® system devices during procedures. And how the issu is been managed.

Unit of Measure: Number of technical incidents

1 day

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Sponsori

Julkaisuja ja hyödyllisiä linkkejä

Tutkimusta koskevien tietojen syöttämisestä vastaava henkilö toimittaa nämä julkaisut vapaaehtoisesti. Nämä voivat koskea mitä tahansa tutkimukseen liittyvää.

Yleiset julkaisut

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Perjantai 4. syyskuuta 2026

Ensisijainen valmistuminen (Arvioitu)

Tiistai 1. joulukuuta 2026

Opintojen valmistuminen (Arvioitu)

Maanantai 1. maaliskuuta 2027

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Torstai 23. heinäkuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Torstai 23. heinäkuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Tiistai 28. heinäkuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Perjantai 11. syyskuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Keskiviikko 9. syyskuuta 2026

Viimeksi vahvistettu

Keskiviikko 1. heinäkuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .

Tilaa