4-electrode ECG (DigitalCardio®) vs Standard 10-electrode ECG: Evaluating Ease of Use and Recording Quality (FAST-ECG)

9. september 2026 oppdatert av: Clinique Bizet

The goal of this clinical trial is to compare the time required to record a complete Electrocardiogram (ECG) trace using the DigitalCardio (CardioSecur®) tablet device versus a standard 12-lead ECG device in eligible patients.

The main questions it aims to answer are:

Is there a difference in diagnostic recording time (in minutes) between the DigitalCardio tablet and a standard ECG machine? How do the two devices compare in terms of trace quality (including QRS duration, QT interval, frontal axis measurements, and missing/erroneous leads)? How satisfied are healthcare professionals with the CardioSecur® device?

Researchers will compare Group A (standard 10-electrode ECG followed by 4-electrode ECG) to Group B (4-electrode ECG followed by standard 10-electrode ECG) in a 1:1 randomized crossover design to evaluate recording efficiency and trace quality.

Participants will:

Complete an initial screening visit with the principal investigator to review eligibility criteria and sign the informed consent form.

Be randomly assigned (1:1 ratio) to either Group A or Group B. Undergo both ECG recording methods sequentially (a standard 10-electrode ECG and a 4-electrode ECG using the DigitalCardio tablet), while researchers measure the total time elapsed from electrode placement to obtaining a complete trace.

Studieoversikt

Detaljert beskrivelse

This study is a randomized, crossover clinical trial designed to compare the efficiency, trace quality, and user satisfaction of a novel 4-electrode tablet-based Electrocardiogram (ECG) system (DigitalCardio / CardioSecur®) against a standard 12-lead (10-electrode) ECG device.

Background and Rationale

An electrocardiogram (ECG) is a fundamental diagnostic tool used to record the electrical activity of the heart. Conventional 12-lead ECGs require placing 10 individual electrodes on a patient's limbs and chest while the patient is lying down. While highly detailed, electrode placement and device setup can be time-consuming in clinical workflows. The DigitalCardio tablet system utilizes a simplified 4-electrode setup designed to streamline data acquisition without compromising diagnostic accuracy.

Study Design and Procedures

After an initial screening process led by the principal investigator, eligible participants receive detailed study information and provide written informed consent. Enrolled participants are then randomly assigned in a 1:1 ratio to one of two sequence groups to control for order effects:

*Group A:Undergoes standard 10-electrode ECG acquisition first, followed immediately by 4-electrode DigitalCardio acquisition.

Group B: Undergoes 4-electrode DigitalCardio acquisition first, followed immediately by standard 10-electrode ECG acquisition.

To maintain consistency, both ECG recordings for a given participant are performed by the same healthcare operator within a minimal time interval.

Key Measurements

  1. Diagnostic Time:Research staff record the exact time elapsed (in minutes) from the initiation of electrode placement to the successful capture of a complete ECG trace for both devices.
  2. Trace Quality and Concordance:Printed and anonymized ECG traces are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads.
  3. Healthcare Professional Satisfaction:Participating healthcare professionals complete a structured questionnaire evaluating user satisfaction and usability of the CardioSecur® device in routine practice.

Studietype

Intervensjonell

Registrering (Antatt)

50

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiekontakt

Studiesteder

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

  • Voksen
  • Eldre voksen

Tar imot friske frivillige

Ja

Beskrivelse

Inclusion Criteria:

Hospitalized Cardiology Patients:

  • Age minimum 18 years
  • Hospitalized for an acute or chronic cardiac condition (e.g., heart failure, acute coronary syndrome, arrhythmia, valvular heart disease, etc.)
  • Ability to give informed consent
  • Hemodynamic stability permitting the performance of an ECG without clinical risk

Cardiology Outpatients:

  • Age minimum of 18 years
  • Follow-up for a known cardiac condition or suspected cardiovascular disease
  • No contraindications to performing an ECG
  • Informed consent obtained

Exclusion Criteria:

  • Minor patient (< 18 years of age)
  • Adult subject under legal protection, under temporary guardianship, curatorship, or tutelage
  • Patient refusing to participate in the study
  • Presence of an implantable cardiac device (e.g., pacemaker, defibrillator) that may interfere with ECG measurements
  • Severe skin condition preventing the placement of electrodes
  • Physical or cognitive inability to follow the study protocol
  • Current pregnancy

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Støttende omsorg
  • Tildeling: Randomisert
  • Intervensjonsmodell: Crossover-oppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Annen: standard 10-electrode ECG
Undergoes standard 10-electrode ECG acquisition first
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.
Aktiv komparator: 4-electrode DigitalCardio
Undergoes 4-electrode DigitalCardio
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Comparison of Procedure Time Between 4-Electrode DigitalCardio® ECG and Standard 12-Lead ECG
Tidsramme: 1 day

Time elapsed from the initiation of electrode placement to the obtaining of a complete ECG trace for both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Minutes

1 day

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Comparative Quality Assessment of ECG Traces
Tidsramme: 1 day

Proportion of paired recordings in which an independent cardiologist prefers the 4-electrode DigitalCardio® trace compared to the standard 12-lead ECG trace, based on visual evaluation of signal clarity and the presence of artifacts/interference.

Unit of Measure: Percentage of paired recordings

1 day
Intraclass Correlation Coefficient (ICC) for Quantitative ECG Parameters
Tidsramme: 1 day

Degree of agreement for quantitative clinical parameters (e.g., QRS duration, QT interval, frontal axis measurements) automatically provided by both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Intraclass Correlation Coefficient (ICC) (ranging from 0 to 1, where higher values indicate better agreement)

1 day
Diagnostic Concordance for Cardiac Pathology
Tidsramme: 1 day

Agreement in detecting the presence or absence of cardiac pathology using the 4-electrode DigitalCardio® system compared to the standard 12-lead ECG (gold standard).

Unit of Measure: Percentage of concordant diagnostic evaluations

1 day
Healthcare Professional Satisfaction Score
Tidsramme: 1 days

Overall user satisfaction regarding the ease of use of the DigitalCardio® tablet device, evaluated by participating healthcare professionals using a structured Likert scale. The score ranges from 1 (strongly disagree / highly unsatisfied) to 5 (strongly agree / highly satisfied), where higher scores indicate a better outcome (higher satisfaction).

Unit of Measure: Score on a scale

1 days
Incidence of Device-Related Technical Incidents for 4-electrode DigitalCardio® system
Tidsramme: 1 day

Total number of technical malfunctions, failures, or operational incidents reported for 4-electrode DigitalCardio® system devices during procedures. And how the issu is been managed.

Unit of Measure: Number of technical incidents

1 day

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Sponsor

Publikasjoner og nyttige lenker

Den som er ansvarlig for å legge inn informasjon om studien leverer frivillig disse publikasjonene. Disse kan handle om alt relatert til studiet.

Generelle publikasjoner

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Faktiske)

4. september 2026

Primær fullføring (Antatt)

1. desember 2026

Studiet fullført (Antatt)

1. mars 2027

Datoer for studieregistrering

Først innsendt

23. juli 2026

Først innsendt som oppfylte QC-kriteriene

23. juli 2026

Først lagt ut (Faktiske)

28. juli 2026

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

11. september 2026

Siste oppdatering sendt inn som oppfylte QC-kriteriene

9. september 2026

Sist bekreftet

1. juli 2026

Mer informasjon

Begreper knyttet til denne studien

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Nei

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .

Abonnere