4-electrode ECG (DigitalCardio®) vs Standard 10-electrode ECG: Evaluating Ease of Use and Recording Quality (FAST-ECG)

9 września 2026 zaktualizowane przez: Clinique Bizet

The goal of this clinical trial is to compare the time required to record a complete Electrocardiogram (ECG) trace using the DigitalCardio (CardioSecur®) tablet device versus a standard 12-lead ECG device in eligible patients.

The main questions it aims to answer are:

Is there a difference in diagnostic recording time (in minutes) between the DigitalCardio tablet and a standard ECG machine? How do the two devices compare in terms of trace quality (including QRS duration, QT interval, frontal axis measurements, and missing/erroneous leads)? How satisfied are healthcare professionals with the CardioSecur® device?

Researchers will compare Group A (standard 10-electrode ECG followed by 4-electrode ECG) to Group B (4-electrode ECG followed by standard 10-electrode ECG) in a 1:1 randomized crossover design to evaluate recording efficiency and trace quality.

Participants will:

Complete an initial screening visit with the principal investigator to review eligibility criteria and sign the informed consent form.

Be randomly assigned (1:1 ratio) to either Group A or Group B. Undergo both ECG recording methods sequentially (a standard 10-electrode ECG and a 4-electrode ECG using the DigitalCardio tablet), while researchers measure the total time elapsed from electrode placement to obtaining a complete trace.

Przegląd badań

Szczegółowy opis

This study is a randomized, crossover clinical trial designed to compare the efficiency, trace quality, and user satisfaction of a novel 4-electrode tablet-based Electrocardiogram (ECG) system (DigitalCardio / CardioSecur®) against a standard 12-lead (10-electrode) ECG device.

Background and Rationale

An electrocardiogram (ECG) is a fundamental diagnostic tool used to record the electrical activity of the heart. Conventional 12-lead ECGs require placing 10 individual electrodes on a patient's limbs and chest while the patient is lying down. While highly detailed, electrode placement and device setup can be time-consuming in clinical workflows. The DigitalCardio tablet system utilizes a simplified 4-electrode setup designed to streamline data acquisition without compromising diagnostic accuracy.

Study Design and Procedures

After an initial screening process led by the principal investigator, eligible participants receive detailed study information and provide written informed consent. Enrolled participants are then randomly assigned in a 1:1 ratio to one of two sequence groups to control for order effects:

*Group A:Undergoes standard 10-electrode ECG acquisition first, followed immediately by 4-electrode DigitalCardio acquisition.

Group B: Undergoes 4-electrode DigitalCardio acquisition first, followed immediately by standard 10-electrode ECG acquisition.

To maintain consistency, both ECG recordings for a given participant are performed by the same healthcare operator within a minimal time interval.

Key Measurements

  1. Diagnostic Time:Research staff record the exact time elapsed (in minutes) from the initiation of electrode placement to the successful capture of a complete ECG trace for both devices.
  2. Trace Quality and Concordance:Printed and anonymized ECG traces are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads.
  3. Healthcare Professional Satisfaction:Participating healthcare professionals complete a structured questionnaire evaluating user satisfaction and usability of the CardioSecur® device in routine practice.

Typ studiów

Interwencyjne

Zapisy (Szacowany)

50

Faza

  • Nie dotyczy

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Kontakt w sprawie studiów

Lokalizacje studiów

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

  • Dorosły
  • Starszy dorosły

Akceptuje zdrowych ochotników

Tak

Opis

Inclusion Criteria:

Hospitalized Cardiology Patients:

  • Age minimum 18 years
  • Hospitalized for an acute or chronic cardiac condition (e.g., heart failure, acute coronary syndrome, arrhythmia, valvular heart disease, etc.)
  • Ability to give informed consent
  • Hemodynamic stability permitting the performance of an ECG without clinical risk

Cardiology Outpatients:

  • Age minimum of 18 years
  • Follow-up for a known cardiac condition or suspected cardiovascular disease
  • No contraindications to performing an ECG
  • Informed consent obtained

Exclusion Criteria:

  • Minor patient (< 18 years of age)
  • Adult subject under legal protection, under temporary guardianship, curatorship, or tutelage
  • Patient refusing to participate in the study
  • Presence of an implantable cardiac device (e.g., pacemaker, defibrillator) that may interfere with ECG measurements
  • Severe skin condition preventing the placement of electrodes
  • Physical or cognitive inability to follow the study protocol
  • Current pregnancy

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

  • Główny cel: Leczenie podtrzymujące
  • Przydział: Randomizowane
  • Model interwencyjny: Zadanie krzyżowe
  • Maskowanie: Brak (otwarta etykieta)

Broń i interwencje

Grupa uczestników / Arm
Interwencja / Leczenie
Inny: standard 10-electrode ECG
Undergoes standard 10-electrode ECG acquisition first
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.
Aktywny komparator: 4-electrode DigitalCardio
Undergoes 4-electrode DigitalCardio
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Comparison of Procedure Time Between 4-Electrode DigitalCardio® ECG and Standard 12-Lead ECG
Ramy czasowe: 1 day

Time elapsed from the initiation of electrode placement to the obtaining of a complete ECG trace for both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Minutes

1 day

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Comparative Quality Assessment of ECG Traces
Ramy czasowe: 1 day

Proportion of paired recordings in which an independent cardiologist prefers the 4-electrode DigitalCardio® trace compared to the standard 12-lead ECG trace, based on visual evaluation of signal clarity and the presence of artifacts/interference.

Unit of Measure: Percentage of paired recordings

1 day
Intraclass Correlation Coefficient (ICC) for Quantitative ECG Parameters
Ramy czasowe: 1 day

Degree of agreement for quantitative clinical parameters (e.g., QRS duration, QT interval, frontal axis measurements) automatically provided by both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Intraclass Correlation Coefficient (ICC) (ranging from 0 to 1, where higher values indicate better agreement)

1 day
Diagnostic Concordance for Cardiac Pathology
Ramy czasowe: 1 day

Agreement in detecting the presence or absence of cardiac pathology using the 4-electrode DigitalCardio® system compared to the standard 12-lead ECG (gold standard).

Unit of Measure: Percentage of concordant diagnostic evaluations

1 day
Healthcare Professional Satisfaction Score
Ramy czasowe: 1 days

Overall user satisfaction regarding the ease of use of the DigitalCardio® tablet device, evaluated by participating healthcare professionals using a structured Likert scale. The score ranges from 1 (strongly disagree / highly unsatisfied) to 5 (strongly agree / highly satisfied), where higher scores indicate a better outcome (higher satisfaction).

Unit of Measure: Score on a scale

1 days
Incidence of Device-Related Technical Incidents for 4-electrode DigitalCardio® system
Ramy czasowe: 1 day

Total number of technical malfunctions, failures, or operational incidents reported for 4-electrode DigitalCardio® system devices during procedures. And how the issu is been managed.

Unit of Measure: Number of technical incidents

1 day

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Sponsor

Publikacje i pomocne linki

Osoba odpowiedzialna za wprowadzenie informacji o badaniu dobrowolnie udostępnia te publikacje. Mogą one dotyczyć wszystkiego, co jest związane z badaniem.

Publikacje ogólne

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Rzeczywisty)

4 września 2026

Zakończenie podstawowe (Szacowany)

1 grudnia 2026

Ukończenie studiów (Szacowany)

1 marca 2027

Daty rejestracji na studia

Pierwszy przesłany

23 lipca 2026

Pierwszy przesłany, który spełnia kryteria kontroli jakości

23 lipca 2026

Pierwszy wysłany (Rzeczywisty)

28 lipca 2026

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

11 września 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

9 września 2026

Ostatnia weryfikacja

1 lipca 2026

Więcej informacji

Terminy związane z tym badaniem

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Nie

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .

Subskrybuj