4-electrode ECG (DigitalCardio®) vs Standard 10-electrode ECG: Evaluating Ease of Use and Recording Quality (FAST-ECG)

9. September 2026 aktualisiert von: Clinique Bizet

The goal of this clinical trial is to compare the time required to record a complete Electrocardiogram (ECG) trace using the DigitalCardio (CardioSecur®) tablet device versus a standard 12-lead ECG device in eligible patients.

The main questions it aims to answer are:

Is there a difference in diagnostic recording time (in minutes) between the DigitalCardio tablet and a standard ECG machine? How do the two devices compare in terms of trace quality (including QRS duration, QT interval, frontal axis measurements, and missing/erroneous leads)? How satisfied are healthcare professionals with the CardioSecur® device?

Researchers will compare Group A (standard 10-electrode ECG followed by 4-electrode ECG) to Group B (4-electrode ECG followed by standard 10-electrode ECG) in a 1:1 randomized crossover design to evaluate recording efficiency and trace quality.

Participants will:

Complete an initial screening visit with the principal investigator to review eligibility criteria and sign the informed consent form.

Be randomly assigned (1:1 ratio) to either Group A or Group B. Undergo both ECG recording methods sequentially (a standard 10-electrode ECG and a 4-electrode ECG using the DigitalCardio tablet), while researchers measure the total time elapsed from electrode placement to obtaining a complete trace.

Studienübersicht

Detaillierte Beschreibung

This study is a randomized, crossover clinical trial designed to compare the efficiency, trace quality, and user satisfaction of a novel 4-electrode tablet-based Electrocardiogram (ECG) system (DigitalCardio / CardioSecur®) against a standard 12-lead (10-electrode) ECG device.

Background and Rationale

An electrocardiogram (ECG) is a fundamental diagnostic tool used to record the electrical activity of the heart. Conventional 12-lead ECGs require placing 10 individual electrodes on a patient's limbs and chest while the patient is lying down. While highly detailed, electrode placement and device setup can be time-consuming in clinical workflows. The DigitalCardio tablet system utilizes a simplified 4-electrode setup designed to streamline data acquisition without compromising diagnostic accuracy.

Study Design and Procedures

After an initial screening process led by the principal investigator, eligible participants receive detailed study information and provide written informed consent. Enrolled participants are then randomly assigned in a 1:1 ratio to one of two sequence groups to control for order effects:

*Group A:Undergoes standard 10-electrode ECG acquisition first, followed immediately by 4-electrode DigitalCardio acquisition.

Group B: Undergoes 4-electrode DigitalCardio acquisition first, followed immediately by standard 10-electrode ECG acquisition.

To maintain consistency, both ECG recordings for a given participant are performed by the same healthcare operator within a minimal time interval.

Key Measurements

  1. Diagnostic Time:Research staff record the exact time elapsed (in minutes) from the initiation of electrode placement to the successful capture of a complete ECG trace for both devices.
  2. Trace Quality and Concordance:Printed and anonymized ECG traces are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads.
  3. Healthcare Professional Satisfaction:Participating healthcare professionals complete a structured questionnaire evaluating user satisfaction and usability of the CardioSecur® device in routine practice.

Studientyp

Interventionell

Einschreibung (Geschätzt)

50

Phase

  • Unzutreffend

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienkontakt

Studienorte

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

  • Erwachsene
  • Älterer Erwachsener

Akzeptiert gesunde Freiwillige

Ja

Beschreibung

Inclusion Criteria:

Hospitalized Cardiology Patients:

  • Age minimum 18 years
  • Hospitalized for an acute or chronic cardiac condition (e.g., heart failure, acute coronary syndrome, arrhythmia, valvular heart disease, etc.)
  • Ability to give informed consent
  • Hemodynamic stability permitting the performance of an ECG without clinical risk

Cardiology Outpatients:

  • Age minimum of 18 years
  • Follow-up for a known cardiac condition or suspected cardiovascular disease
  • No contraindications to performing an ECG
  • Informed consent obtained

Exclusion Criteria:

  • Minor patient (< 18 years of age)
  • Adult subject under legal protection, under temporary guardianship, curatorship, or tutelage
  • Patient refusing to participate in the study
  • Presence of an implantable cardiac device (e.g., pacemaker, defibrillator) that may interfere with ECG measurements
  • Severe skin condition preventing the placement of electrodes
  • Physical or cognitive inability to follow the study protocol
  • Current pregnancy

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Hauptzweck: Unterstützende Pflege
  • Zuteilung: Zufällig
  • Interventionsmodell: Crossover-Aufgabe
  • Maskierung: Keine (Offenes Etikett)

Waffen und Interventionen

Teilnehmergruppe / Arm
Intervention / Behandlung
Sonstiges: standard 10-electrode ECG
Undergoes standard 10-electrode ECG acquisition first
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.
Aktiver Komparator: 4-electrode DigitalCardio
Undergoes 4-electrode DigitalCardio
Patient undergoes standard 10-electrode ECG acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by 4-electrode DigitalCardio acquisition.
Patient undergoes 4-electrode DigitalCardio acquisition acquisition. Then anonymized ECG traces is printed are submitted to independent evaluators. Evaluators independently assess trace quality based on standardized criteria, comparing automated measurements (frontal axis, median QRS duration, and QT interval) and counting any missing or erroneous leads. Followed immediately by standard 10-electrode ECG acquisition.

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Comparison of Procedure Time Between 4-Electrode DigitalCardio® ECG and Standard 12-Lead ECG
Zeitfenster: 1 day

Time elapsed from the initiation of electrode placement to the obtaining of a complete ECG trace for both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Minutes

1 day

Sekundäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Comparative Quality Assessment of ECG Traces
Zeitfenster: 1 day

Proportion of paired recordings in which an independent cardiologist prefers the 4-electrode DigitalCardio® trace compared to the standard 12-lead ECG trace, based on visual evaluation of signal clarity and the presence of artifacts/interference.

Unit of Measure: Percentage of paired recordings

1 day
Intraclass Correlation Coefficient (ICC) for Quantitative ECG Parameters
Zeitfenster: 1 day

Degree of agreement for quantitative clinical parameters (e.g., QRS duration, QT interval, frontal axis measurements) automatically provided by both the 4-electrode DigitalCardio® system and the standard 12-lead ECG device.

Unit of Measure: Intraclass Correlation Coefficient (ICC) (ranging from 0 to 1, where higher values indicate better agreement)

1 day
Diagnostic Concordance for Cardiac Pathology
Zeitfenster: 1 day

Agreement in detecting the presence or absence of cardiac pathology using the 4-electrode DigitalCardio® system compared to the standard 12-lead ECG (gold standard).

Unit of Measure: Percentage of concordant diagnostic evaluations

1 day
Healthcare Professional Satisfaction Score
Zeitfenster: 1 days

Overall user satisfaction regarding the ease of use of the DigitalCardio® tablet device, evaluated by participating healthcare professionals using a structured Likert scale. The score ranges from 1 (strongly disagree / highly unsatisfied) to 5 (strongly agree / highly satisfied), where higher scores indicate a better outcome (higher satisfaction).

Unit of Measure: Score on a scale

1 days
Incidence of Device-Related Technical Incidents for 4-electrode DigitalCardio® system
Zeitfenster: 1 day

Total number of technical malfunctions, failures, or operational incidents reported for 4-electrode DigitalCardio® system devices during procedures. And how the issu is been managed.

Unit of Measure: Number of technical incidents

1 day

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Sponsor

Publikationen und hilfreiche Links

Die Bereitstellung dieser Publikationen erfolgt freiwillig durch die für die Eingabe von Informationen über die Studie verantwortliche Person. Diese können sich auf alles beziehen, was mit dem Studium zu tun hat.

Allgemeine Veröffentlichungen

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

4. September 2026

Primärer Abschluss (Geschätzt)

1. Dezember 2026

Studienabschluss (Geschätzt)

1. März 2027

Studienanmeldedaten

Zuerst eingereicht

23. Juli 2026

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

23. Juli 2026

Zuerst gepostet (Tatsächlich)

28. Juli 2026

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

11. September 2026

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

9. September 2026

Zuletzt verifiziert

1. Juli 2026

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .