- ICH GCP
- Yhdysvaltain kliinisten tutkimusten rekisteri
- Kliininen tutkimus NCT07826039
3D-Printed PLA Guide Versus Resin Guide in Mandibular Molar Implant Placement (CGIIPMM-PLA-MR)
Computer Guided Immediate Implant Placement in Mandibular Molars Using Patient Specific 3D Printed Polylactic Acid Guide Versus Modified Resin Guide
Tutkimuksen yleiskatsaus
Tila
Yksityiskohtainen kuvaus
This randomized clinical trial compares two tooth-supported computer-guided surgical guide systems for immediate implant placement in mandibular molar extraction sockets: a patient-specific 3D-printed polylactic acid (PLA) guide system and a modified resin guide system. Preoperative CBCT and digital intraoral scans were used for prosthetically driven virtual implant planning and fabrication of the surgical guides.
Implant positioning accuracy was assessed by comparing the planned and actual implant positions using preoperative and postoperative CBCT datasets. Angular deviation, coronal deviation at the implant platform, and apical deviation were recorded.
Primary implant stability was assessed immediately after implant placement using Implant Stability Quotient (ISQ) measurements obtained with an Osstell device. Surgical time was recorded from seating of the surgical guide and initiation of osteotomy until completion of implant insertion.
The study was designed to compare the performance of the two guide systems in immediate mandibular molar implant placement.
Opintotyyppi
Ilmoittautuminen (Todellinen)
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskelupaikat
-
-
New Cairo
-
Cairo, New Cairo, Egypti, 11511
- Future University in Egypt (FUE)
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
- Aikuinen
Hyväksyy terveitä vapaaehtoisia
Kuvaus
Inclusion Criteria:
- Age 20-50 years.
- Non-restorable mandibular first molar indicated for extraction.
- Sufficient bony support for implant placement without the need for significant bone augmentation.
- Good oral hygiene and motivation.
- Adequate prosthetic space.
- Ability and willingness to undergo the surgical procedure and follow-up.
- Suitable for computer-guided, tooth-supported implant placement.
Exclusion Criteria:
- Uncontrolled systemic diseases that may impair healing (e.g., uncontrolled diabetes or hypertension).
- Severe periodontal disease.
- Patients undergoing chemotherapy or radiotherapy.
- Insufficient prosthetic space.
- Heavy smokers.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Hoito
- Jako: Satunnaistettu
- Inventiomalli: Rinnakkaistehtävä
- Naamiointi: Yksittäinen
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
|---|---|
|
Kokeellinen: patient specific 3d printed PLA guide
Immediate implant placement in mandibular molar extraction sockets using a patient-specific tooth-supported 3D-printed PLA surgical guide system.
Preoperative CBCT and digital intraoral scans were used for prosthetically driven virtual implant planning and guide fabrication.
The guide was fabricated from PLA according to the planned implant position.
Following atraumatic tooth extraction and socket preparation, the guide was seated and osteotomy preparation was performed through the guide's guiding channel according to the planned trajectory, followed by immediate implant insertion.
Primary implant stability was assessed using an Osstell device and Implant Stability Quotient (ISQ) measurements.
|
Preoperative virtual planning was performed using CBCT data and digital intraoral scans.
The datasets were used for prosthetically driven implant planning and fabrication of a patient-specific 3D-printed PLA surgical guide.
Following atraumatic tooth extraction and socket preparation, the guide was seated and osteotomy preparation was performed through the guiding channel according to the planned implant position and trajectory.
The implant was then immediately inserted.
Primary implant stability was assessed using an Osstell device and Implant Stability Quotient (ISQ) measurements.
|
|
Active Comparator: Modified Resin guide
Immediate implant placement in mandibular molar extraction sockets using a tooth-supported modified resin surgical guide system.
Preoperative CBCT and digital intraoral scans were used for prosthetically driven virtual implant planning and guide fabrication.
The guide was fabricated from light-cured surgical-guide resin according to the planned implant position.
Following atraumatic tooth extraction and socket preparation, the guide was seated and osteotomy preparation was performed through the guide's guiding channel according to the planned trajectory, followed by immediate implant insertion.
Primary implant stability was assessed using an Osstell device and Implant Stability Quotient (ISQ) measurements.
|
Preoperative virtual planning was performed using CBCT data and digital intraoral scans.
The datasets were used for prosthetically driven implant planning and fabrication of a tooth-supported modified resin surgical guide.
Following atraumatic tooth extraction and socket preparation, the guide was seated and osteotomy preparation was performed through the guiding channel according to the planned implant position and trajectory.
The implant was then immediately inserted.
Primary implant stability was assessed using an Osstell device and Implant Stability Quotient (ISQ) measurements.
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
accuracy of implant position
Aikaikkuna: Assessed within 1 week postoperatively, following acquisition of the postoperative CBCT scan.
|
Accuracy of implant placement will be evaluated by comparing the planned implant position from the preoperative digital plan with the actual implant position obtained from the postoperative CBCT scan.
Deviations will be assessed in three dimensions, including angular deviation (degrees), coronal deviation (platform) (mm), and apical deviation (mm).
Measurements will be performed using specialized implant-planning software by superimposing the preoperative and postoperative datasets.
Lower deviation values indicate greater agreement between the planned and actual implant positions.
|
Assessed within 1 week postoperatively, following acquisition of the postoperative CBCT scan.
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
primary implant stability by osstell
Aikaikkuna: Measured immediately at the time of implant placement during the surgical procedure.
|
Primary implant stability will be assessed using an Osstell device, which measures Implant Stability Quotient (ISQ) values through resonance frequency analysis.
A SmartPeg will be attached to the implant, and ISQ readings will be taken in two perpendicular directions (buccolingual and mesiodistal).
Higher ISQ values indicate greater mechanical stability.
The mean ISQ value for each implant will be recorded and used for analysis
|
Measured immediately at the time of implant placement during the surgical procedure.
|
|
surgical time
Aikaikkuna: Assessed intraoperatively, measured during the surgical appointment, and recorded immediately after surgery
|
Surgical time will be recorded to evaluate the procedural efficiency of guided immediate implant placement.
Time will be measured from seating of the surgical guide and initiation of osteotomy until completion of implant insertion.
The total duration will be recorded in minutes.
|
Assessed intraoperatively, measured during the surgical appointment, and recorded immediately after surgery
|
Yhteistyökumppanit ja tutkijat
Sponsori
Julkaisuja ja hyödyllisiä linkkejä
Yleiset julkaisut
- Shi Y, Wang J, Ma C, Shen J, Dong X, Lin D. A systematic review of the accuracy of digital surgical guides for dental implantation. Int J Implant Dent. 2023 Oct 25;9(1):38. doi: 10.1186/s40729-023-00507-w.
- Ragucci GM, Elnayef B, Criado-Camara E, Del Amo FS, Hernandez-Alfaro F. Immediate implant placement in molar extraction sockets: a systematic review and meta-analysis. Int J Implant Dent. 2020 Oct 13;6(1):40. doi: 10.1186/s40729-020-00235-5.
- Ku JK, Lee J, Lee HJ, Yun PY, Kim YK. Accuracy of dental implant placement with computer-guided surgery: a retrospective cohort study. BMC Oral Health. 2022 Jan 16;22(1):8. doi: 10.1186/s12903-022-02046-z.
- Pieralli S, Spies BC, Hromadnik V, Nicic R, Beuer F, Wesemann C. How Accurate Is Oral Implant Installation Using Surgical Guides Printed from a Degradable and Steam-Sterilized Biopolymer? J Clin Med. 2020 Jul 22;9(8):2322. doi: 10.3390/jcm9082322.
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Todellinen)
Ensisijainen valmistuminen (Todellinen)
Opintojen valmistuminen (Todellinen)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muut tutkimustunnusnumerot
- FUE.REC(78)/10-2025
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
IPD-suunnitelman kuvaus
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .