Cette page a été traduite automatiquement et l'exactitude de la traduction n'est pas garantie. Veuillez vous référer au version anglaise pour un texte source.

Feasibility and Performance Evaluation for INVSENSOR00024

4 février 2020 mis à jour par: Masimo Corporation

Feasibility and Performance Evaluation of INVSENSOR00024

This is a prospective, non-randomized, single arm study design to evaluate SpO2 performance of the RD SET Neo investigational pulse oximeter sensors in the neonatal population using convenience sampling in a clinical environment. The initial subjects under Protocol Version 1 were enrolled as part of a feasibility study to verify the design changes to the sensor are correct. The study protocol was amended with Protocol Version 2 to enroll a group of subjects that will be used to evaluate clinical performance for submission to regulatory agencies.

Aperçu de l'étude

Statut

Complété

Intervention / Traitement

Type d'étude

Interventionnel

Inscription (Réel)

28

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

    • California
      • Stanford, California, États-Unis, 94305
        • Stanford University Medical Center

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

Pas plus vieux que 1 mois (Enfant)

Accepte les volontaires sains

Oui

Sexes éligibles pour l'étude

Tout

La description

Inclusion Criteria:

  • Less than or equal to 1 month of age
  • Weight <3kg
  • Subjects admitted with standard of care (SOC) arterial blood sampling line already in place

Exclusion Criteria:

  • Subjects with underdeveloped skin
  • Subjects with abnormalities at the planned application sites that would interfere with system measurements
  • Subjects with known allergic reactions to foam/rubber products and adhesive tape
  • Deformities of limbs, absence of feet, sever edema, and other at the discretion of the Principal Investigator

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Autre
  • Répartition: N / A
  • Modèle interventionnel: Affectation à un seul groupe
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: RD SET Neo SpO2
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Accuracy of Oxygen Saturation (SpO2) Measurement by RMS Calculation
Délai: up to 8 hours
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the root mean square (RMS) error value. In order to obtain the RMS value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for each sample, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the RMS Error value.
up to 8 hours

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

25 juin 2018

Achèvement primaire (Réel)

8 novembre 2019

Achèvement de l'étude (Réel)

8 novembre 2019

Dates d'inscription aux études

Première soumission

15 juin 2018

Première soumission répondant aux critères de contrôle qualité

15 juin 2018

Première publication (Réel)

26 juin 2018

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

20 février 2020

Dernière mise à jour soumise répondant aux critères de contrôle qualité

4 février 2020

Dernière vérification

1 février 2020

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • HAMM0002

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Oui

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .