Esta página foi traduzida automaticamente e a precisão da tradução não é garantida. Por favor, consulte o versão em inglês para um texto fonte.

Feasibility and Performance Evaluation for INVSENSOR00024

4 de fevereiro de 2020 atualizado por: Masimo Corporation

Feasibility and Performance Evaluation of INVSENSOR00024

This is a prospective, non-randomized, single arm study design to evaluate SpO2 performance of the RD SET Neo investigational pulse oximeter sensors in the neonatal population using convenience sampling in a clinical environment. The initial subjects under Protocol Version 1 were enrolled as part of a feasibility study to verify the design changes to the sensor are correct. The study protocol was amended with Protocol Version 2 to enroll a group of subjects that will be used to evaluate clinical performance for submission to regulatory agencies.

Visão geral do estudo

Status

Concluído

Intervenção / Tratamento

Tipo de estudo

Intervencional

Inscrição (Real)

28

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Locais de estudo

    • California
      • Stanford, California, Estados Unidos, 94305
        • Stanford University Medical Center

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

Não mais velho que 1 mês (Filho)

Aceita Voluntários Saudáveis

Sim

Gêneros Elegíveis para o Estudo

Tudo

Descrição

Inclusion Criteria:

  • Less than or equal to 1 month of age
  • Weight <3kg
  • Subjects admitted with standard of care (SOC) arterial blood sampling line already in place

Exclusion Criteria:

  • Subjects with underdeveloped skin
  • Subjects with abnormalities at the planned application sites that would interfere with system measurements
  • Subjects with known allergic reactions to foam/rubber products and adhesive tape
  • Deformities of limbs, absence of feet, sever edema, and other at the discretion of the Principal Investigator

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Outro
  • Alocação: N / D
  • Modelo Intervencional: Atribuição de grupo único
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: RD SET Neo SpO2
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Accuracy of Oxygen Saturation (SpO2) Measurement by RMS Calculation
Prazo: up to 8 hours
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the root mean square (RMS) error value. In order to obtain the RMS value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for each sample, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the RMS Error value.
up to 8 hours

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Real)

25 de junho de 2018

Conclusão Primária (Real)

8 de novembro de 2019

Conclusão do estudo (Real)

8 de novembro de 2019

Datas de inscrição no estudo

Enviado pela primeira vez

15 de junho de 2018

Enviado pela primeira vez que atendeu aos critérios de CQ

15 de junho de 2018

Primeira postagem (Real)

26 de junho de 2018

Atualizações de registro de estudo

Última Atualização Postada (Real)

20 de fevereiro de 2020

Última atualização enviada que atendeu aos critérios de controle de qualidade

4 de fevereiro de 2020

Última verificação

1 de fevereiro de 2020

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • HAMM0002

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Sim

produto fabricado e exportado dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .

Se inscrever