Questa pagina è stata tradotta automaticamente e l'accuratezza della traduzione non è garantita. Si prega di fare riferimento al Versione inglese per un testo di partenza.

Feasibility and Performance Evaluation for INVSENSOR00024

4 febbraio 2020 aggiornato da: Masimo Corporation

Feasibility and Performance Evaluation of INVSENSOR00024

This is a prospective, non-randomized, single arm study design to evaluate SpO2 performance of the RD SET Neo investigational pulse oximeter sensors in the neonatal population using convenience sampling in a clinical environment. The initial subjects under Protocol Version 1 were enrolled as part of a feasibility study to verify the design changes to the sensor are correct. The study protocol was amended with Protocol Version 2 to enroll a group of subjects that will be used to evaluate clinical performance for submission to regulatory agencies.

Panoramica dello studio

Stato

Completato

Intervento / Trattamento

Tipo di studio

Interventistico

Iscrizione (Effettivo)

28

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

    • California
      • Stanford, California, Stati Uniti, 94305
        • Stanford University Medical Center

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

Non più vecchio di 1 mese (Bambino)

Accetta volontari sani

Sessi ammissibili allo studio

Tutto

Descrizione

Inclusion Criteria:

  • Less than or equal to 1 month of age
  • Weight <3kg
  • Subjects admitted with standard of care (SOC) arterial blood sampling line already in place

Exclusion Criteria:

  • Subjects with underdeveloped skin
  • Subjects with abnormalities at the planned application sites that would interfere with system measurements
  • Subjects with known allergic reactions to foam/rubber products and adhesive tape
  • Deformities of limbs, absence of feet, sever edema, and other at the discretion of the Principal Investigator

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Altro
  • Assegnazione: N / A
  • Modello interventistico: Assegnazione di gruppo singolo
  • Mascheramento: Nessuno (etichetta aperta)

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Sperimentale: RD SET Neo SpO2
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Accuracy of Oxygen Saturation (SpO2) Measurement by RMS Calculation
Lasso di tempo: up to 8 hours
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the root mean square (RMS) error value. In order to obtain the RMS value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for each sample, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the RMS Error value.
up to 8 hours

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

25 giugno 2018

Completamento primario (Effettivo)

8 novembre 2019

Completamento dello studio (Effettivo)

8 novembre 2019

Date di iscrizione allo studio

Primo inviato

15 giugno 2018

Primo inviato che soddisfa i criteri di controllo qualità

15 giugno 2018

Primo Inserito (Effettivo)

26 giugno 2018

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

20 febbraio 2020

Ultimo aggiornamento inviato che soddisfa i criteri QC

4 febbraio 2020

Ultimo verificato

1 febbraio 2020

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • HAMM0002

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .