In Vivo-Ex Vivo Evaluation of the Root ZX3 for Working-Length Determination
Accuracy and Precision of the Root ZX3 Electronic Apex Locator for Working-Length Determination in Teeth With Different Pulpal and Periapical Conditions: A Mixed In Vivo-Ex Vivo Diagnostic Accuracy Study
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
Accurate determination of root canal working length is essential for successful endodontic treatment because instrumentation and obturation should terminate at an appropriate apical endpoint. Electronic apex locators have become widely used to improve working-length determination and reduce reliance on radiographic methods. The Root ZX3 is a third-generation electronic apex locator that has demonstrated promising accuracy; however, limited evidence exists regarding its performance under different pulpal and periapical conditions using microscopic actual working length as the reference standard.
This prospective mixed in vivo-ex vivo diagnostic accuracy study evaluates the accuracy and repeatability of the Root ZX3 electronic apex locator in single-rooted permanent teeth indicated for extraction. Adult participants requiring extraction of teeth for prosthodontic, periodontal, orthodontic, or non-restorable reasons and requiring primary nonsurgical root canal treatment are enrolled after providing informed consent.
Before tooth extraction, electronic working length is determined using the Root ZX3 according to the manufacturer's instructions with a standardized coronal reference point. Pulpal and periapical diagnoses are recorded clinically. After extraction, limited field-of-view CBCT imaging is performed, and working length is measured using standardized imaging protocols. The actual working length is then determined microscopically by directly identifying the apical foramen and is used as the reference standard.
The primary outcome is the diagnostic accuracy of the Root ZX3 compared with the microscopic actual working length. Accuracy is evaluated as the proportion of measurements within ±0.5 mm and ±1.0 mm of the reference standard. Agreement between measurement methods is assessed using Bland-Altman analysis, Lin's concordance correlation coefficient, and the intraclass correlation coefficient. Repeatability of electronic measurements is also evaluated.
Tipo di studio
Iscrizione (Effettivo)
Fase
- Non applicabile
Contatti e Sedi
Luoghi di studio
-
-
Riyadh Region
-
Riyadh, Riyadh Region, Arabia Saudita, 12755
- Abier alhekeir
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Descrizione
Inclusion Criteria:
* Adults (≥18 years) requiring extraction of a single-rooted permanent tooth for prosthodontic, periodontal, orthodontic, or non-restorable reasons.
- Tooth requiring primary nonsurgical root canal treatment.
- Fully formed root with a closed apex.
- Ability to provide written informed consent.
Exclusion Criteria:
Teeth with root resorption (internal or external).
- Teeth with root perforation or previous endodontic treatment.
- Teeth with immature (open) apices.
- Teeth with root fractures.
- Teeth in which apical patency could not be achieved.
- Teeth with no electronic working-length measurement obtained using the Root ZX3.
- Teeth that fractured during extraction, preventing microscopic determination of the actual working length.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Diagnostico
- Assegnazione: N / A
- Modello interventistico: Assegnazione di gruppo singolo
- Mascheramento: Nessuno (etichetta aperta)
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: Root ZX3 Working-Length Determination
Participants underwent electronic working-length determination using the Root ZX3 electronic apex locator during primary nonsurgical root canal treatment.
Electronic working-length measurements were obtained before tooth extraction and subsequently compared with cone-beam computed tomography (CBCT) measurements and microscopic actual working length, with the microscopic measurement serving as the reference standard for evaluating diagnostic accuracy and precision.
|
The Root ZX3 electronic apex locator was used to determine electronic working length in single-rooted permanent teeth during primary nonsurgical root canal treatment according to the manufacturer's instructions.
A size 10 or 15 K-file was advanced within the root canal until the device indicated the apical foramen (red bar), and the working length was recorded using a standardized coronal reference point.
Electronic measurements were compared with microscopic actual working length after tooth extraction, which served as the reference standard.
Cone-beam computed tomography (CBCT) measurements were also obtained for comparison.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Diagnostic accuracy of the Root ZX3 electronic apex locator for working-length determination
Lasso di tempo: during endodontic treatment before tooth extraction
|
The diagnostic accuracy of the Root ZX3 electronic apex locator was evaluated by comparing electronic working-length measurements with the microscopic actual working length, which served as the reference standard.
Accuracy was assessed by the proportion of measurements within ±0.5 mm and ±1.0 mm of the reference standard.
Agreement was further evaluated using mean differences and Bland-Altman analysis.
|
during endodontic treatment before tooth extraction
|
Misure di risultato secondarie
Misura del risultato |
Lasso di tempo |
|---|---|
|
Repeatability of Root ZX3 measurements
Lasso di tempo: before tooth extrac
|
before tooth extrac
|
Collaboratori e investigatori
Sponsor
Investigatori
- Investigatore principale: Abier Fahad Alhekeir, Riyadh Elm University
Pubblicazioni e link utili
Pubblicazioni generali
- Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals. J Endod. 1998 Jan;24(1):48-50. doi: 10.1016/S0099-2399(98)80214-8.
- Izadi A, Golmakani F, Kazeminejad E, Mahdavi Asl A. Accuracy of Working Length Measurement Using Cone Beam Computed Tomography at Three Field of View Settings, Conventional Radiography, and Electronic Apex Locator: An Ex-vivo Study. Eur Endod J. 2024 Aug 22;9(3):266-272. doi: 10.14744/eej.2023.97769.
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- HAP-01-R-14
- 2529 (Altro identificatore: Saudi food and drug authority)
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Root ZX3 Electronic Apex Locator
-
National and Kapodistrian University of AthensAttivo, non reclutanteIndagine clinica sui parametri chimici per una disinfezione efficiente del canale radicolare (ENDOA)Parodontite apicale | Infezione del canale radicolareGrecia