NeoSpHb Study: Can a Painless Device Detect Anaemia in Babies Without Blood Tests? A Study in the Neonatal Intensive Care Unit (NICU) (NeoSpHb)
Validation of Non-invasive Spectrophotometric Haemoglobin (SpHb) Measurement for Detection of Anaemia and Observation of Trends in Infants Admitted to the Neonatal Intensive Care Unit (NICU): A Prospective Diagnostic Accuracy Study.
The goal of this study is to evaluate the diagnostic performance and feasibility of non-invasive spectrophotometric haemoglobin (SpHb) measurement for detecting anaemia in infants admitted to the Neonatal Intensive Care Unit (NICU). The main questions it aims to answer are:
- What is the agreement between non-invasive SpHb and laboratory reference total haemoglobin (tHb)?
- How accurately can non-invasive SpHb detect clinically significant laboratory anaemia (Hb < 10g/dL)?
- What is the feasibility (success rate and time-to-reading) and what physiological factors (e.g., perfusion index) moderate measurement accuracy in NICU infants? Researchers will compare non-invasive continuous SpHb readings against routine invasive laboratory haemoglobin values (tHb) obtained during standard clinical care from the same clinical encounter.
Participants will:
- Have a non-invasive optical sensor applied to an approved peripheral site (e.g., foot or palm) to record an SpHb reading whenever routine blood sampling is required by their attending clinician.
- Undergo standard routine blood sampling (venous, arterial, or capillary) for laboratory haemoglobin testing as part of usual clinical care, with no extra blood drawn for research purposes.
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
What is the purpose of this study?
This study aims to find out how accurate a non-invasive device (SpHb) is in measuring haemoglobin (red blood cells) levels in babies admitted to the NICU. The study involves the use of the Masimo SpHb® (Spectrophotometric Hemoglobin), a non-invasive monitoring device currently used in clinical settings. The device is handheld and connected to a computer/tablet for recording measurements. It will be placed on the child's skin to measure the child's hemoglobin. The procedure does not involve needles, radiation, or painful interventions.
Why is this study important?
Babies in NICU often have anemia, where their red blood cells level is low. To monitor this condition, babies often need repeated blood tests, which can cause pain and blood loss. This study may help reduce unnecessary blood sampling in the future.
What type of study is this?
This is an observational study. We will compare readings from a non-invasive device with routine blood test results.
What is the procedure that is being tested? A device called Masimo SpHb® is placed on the baby's skin to measure haemoglobin using light. No blood is taken for this measurement.
Tipo di studio
Iscrizione (Stimato)
Fase
- Non applicabile
Contatti e Sedi
Contatto studio
- Nome: Aliyyah Mohammad Khuzaini, MBChB, MRCPCH, MMEd
- Numero di telefono: +60132134616
- Email: aliyyahkhuzaini@usim.edu.my
Backup dei contatti dello studio
- Nome: Azanna Ahmad Kamar, MBBS, MRCPCH
- Numero di telefono: +603-79492065
- Email: azanna@um.edu.my
Luoghi di studio
-
-
Kuala Lumpur
-
Kuala Lumpur, Kuala Lumpur, Malaysia, 50603
- University Malaya Medical Center
-
Contatto:
- Aliyyah Mohammad Khuzaini, MBChB, MRCPCH, MMEd
- Numero di telefono: +60132134616
- Email: aliyyahkhuzaini@usim.edu.my
-
Contatto:
- Minn Yin Choo, MBBS, MMed, MRCPCH
- Numero di telefono: +603-79492065
- Email: minnyin@ummc.edu.my
-
Investigatore principale:
- Aliyyah Mohammad Khuzaini, MBChB, MRCPCH, MMEd
-
Sub-investigatore:
- Minn Yin Choo, MBBS, MMed, MRCPCH
-
Sub-investigatore:
- Azanna Ahmad Kamar, MBBS, MRCPCH
-
Sub-investigatore:
- Tengku Amatullah Madeehah Tengku Mohd, MBChB, MSc, MPH, DrPH
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Bambino
- Adulto
- Adulto più anziano
Accetta volontari sani
Descrizione
Inclusion Criteria:
- Infants (preterm or term) admitted to the Neonatal Intensive Care Unit.
- Gestational age ≥ 24 weeks at birth.
- Admitted to NICU during the study period.
- Clinical indication for laboratory haemoglobin measurement as part of routine care.
- Written informed consent obtained from parent or legal guardian.
Exclusion Criteria:
- Known or suspected haemoglobinopathy.
- Cyanotic congenital heart disease.
- Significant skin integrity issues at the proposed sensor site (e.g. burns, severe dermatitis, wounds, oedema, or medical devices preventing probe placement).
- Conditions likely to cause unreliable peripheral perfusion at the discretion of the attending clinician.
- Severe hypothermia (below 32°C) or hyperthermia (more than 40°C).
- Prolonged capillary refill time indicating poor peripheral perfusion.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Diagnostico
- Assegnazione: N / A
- Modello interventistico: Assegnazione di gruppo singolo
- Mascheramento: Nessuno (etichetta aperta)
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: SpHb
Infants admitted to the NICU will undergo non-invasive SpHb measurement, with results compared against laboratory haemoglobin values obtained as part of routine clinical care.
|
Altri nomi:
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Agreement between non-invasive SpHb and laboratory haemoglobin measurements
Lasso di tempo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
Agreement between SpHb and laboratory haemoglobin will be assessed using Bland-Altman analysis, including mean bias and 95% limits of agreement. Unit of measure: g/dL |
During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Sensitivity and specificity of SpHb for detection of anaemia
Lasso di tempo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
Sensitivity and specificity of SpHb for detecting anaemia, using laboratory haemoglobin as the reference standard. Anaemia will be defined according to the pre-specified haemoglobin threshold appropriate for the infant's gestational/postnatal age. Unit of measure: Percentage of participants |
During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
|
Agreement between changes in SpHb and changes in laboratory haemoglobin over time
Lasso di tempo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
Description: The ability of serial SpHb measurements to detect changes in haemoglobin concentration over time will be assessed by comparing changes in SpHb with corresponding changes in laboratory haemoglobin measurements. Unit of measure: g/dL |
During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
|
|
Regression coefficient for perfusion index and SpHb-tHb measurement difference
Lasso di tempo: From enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
Regression coefficient (β) and 95% confidence interval for perfusion index (measured by pulse co-oximeter perfusion index display, %) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) per 1% change in perfusion index |
From enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
|
Regression coefficient for gestational age and SpHb-tHb measurement difference
Lasso di tempo: Gestational age recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
Regression coefficient (β) and 95% confidence interval for gestational age at birth (recorded in completed weeks from obstetric dating, per medical records) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) per completed week of gestational age |
Gestational age recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
|
Regression coefficient for birth weight and SpHb-tHb measurement difference
Lasso di tempo: Birth weight recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
Regression coefficient (β) and 95% confidence interval for birth weight (recorded in grams at delivery, per medical records) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) per 100 grams of birth weight |
Birth weight recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
|
Regression coefficient for phototherapy and SpHb-tHb measurement difference
Lasso di tempo: Assessed at each paired SpHb-tHb measurement, from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
Regression coefficient (β) and 95% confidence interval for concurrent phototherapy use (yes/no, recorded from NICU treatment records at the time of each paired measurement) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant. Unit of Measure: Regression coefficient (β) for phototherapy exposure (on vs off) |
Assessed at each paired SpHb-tHb measurement, from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
|
Collaboratori e investigatori
Sponsor
Collaboratori
Pubblicazioni e link utili
Pubblicazioni generali
- Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg. 2011 May;201(5):592-8. doi: 10.1016/j.amjsurg.2011.01.020.
- Nicholas C, George R, Sardesai S, Durand M, Ramanathan R, Cayabyab R. Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants. J Perinatol. 2015 Aug;35(8):617-20. doi: 10.1038/jp.2015.12. Epub 2015 Mar 5.
- Saengnipanthkul S, Tawiangnan C, Sirikarn P, Supasai S, Sitthikarnkha P, Techasatian L, Uppala R. Accuracy of noninvasive total haemoglobin measurement for detecting anaemia in infants: a cross-sectional study in a well-child care clinic, Khon Kaen, Thailand. BMJ Open. 2024 Dec 30;14(12):e083356. doi: 10.1136/bmjopen-2023-083356.
- Arai Y, Shoji H, Awata K, Inage E, Ikuse T, Shimizu T. Evaluation of the use of non-invasive hemoglobin measurement in early childhood. Pediatr Res. 2023 Mar;93(4):1036-1040. doi: 10.1038/s41390-022-02204-7. Epub 2022 Jul 29.
- Hsu DP, French AJ, Madson SL, Palmer JM, Gidvani-Diaz V. Evaluation of a Noninvasive Hemoglobin Measurement Device to Screen for Anemia in Infancy. Matern Child Health J. 2016 Apr;20(4):827-32. doi: 10.1007/s10995-015-1913-9.
- Noiri E, Kobayashi N, Takamura Y, Iijima T, Takagi T, Doi K, Nakao A, Yamamoto T, Takeda S, Fujita T. Pulse total-hemoglobinometer provides accurate noninvasive monitoring. Crit Care Med. 2005 Dec;33(12):2831-5. doi: 10.1097/01.ccm.0000190430.96750.51.
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- 2026219-16355
- ID RSCH ID-26-05390-RJH (Identificatore di registro: National Medical Research Registry, Malaysia)
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Descrizione del piano IPD
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .