NeoSpHb Study: Can a Painless Device Detect Anaemia in Babies Without Blood Tests? A Study in the Neonatal Intensive Care Unit (NICU) (NeoSpHb)

24 de agosto de 2026 atualizado por: Aliyyah Mohammad Khuzaini, University of Malaya

Validation of Non-invasive Spectrophotometric Haemoglobin (SpHb) Measurement for Detection of Anaemia and Observation of Trends in Infants Admitted to the Neonatal Intensive Care Unit (NICU): A Prospective Diagnostic Accuracy Study.

The goal of this study is to evaluate the diagnostic performance and feasibility of non-invasive spectrophotometric haemoglobin (SpHb) measurement for detecting anaemia in infants admitted to the Neonatal Intensive Care Unit (NICU). The main questions it aims to answer are:

  1. What is the agreement between non-invasive SpHb and laboratory reference total haemoglobin (tHb)?
  2. How accurately can non-invasive SpHb detect clinically significant laboratory anaemia (Hb < 10g/dL)?
  3. What is the feasibility (success rate and time-to-reading) and what physiological factors (e.g., perfusion index) moderate measurement accuracy in NICU infants? Researchers will compare non-invasive continuous SpHb readings against routine invasive laboratory haemoglobin values (tHb) obtained during standard clinical care from the same clinical encounter.

Participants will:

  1. Have a non-invasive optical sensor applied to an approved peripheral site (e.g., foot or palm) to record an SpHb reading whenever routine blood sampling is required by their attending clinician.
  2. Undergo standard routine blood sampling (venous, arterial, or capillary) for laboratory haemoglobin testing as part of usual clinical care, with no extra blood drawn for research purposes.

Visão geral do estudo

Status

Ainda não está recrutando

Condições

Descrição detalhada

What is the purpose of this study?

This study aims to find out how accurate a non-invasive device (SpHb) is in measuring haemoglobin (red blood cells) levels in babies admitted to the NICU. The study involves the use of the Masimo SpHb® (Spectrophotometric Hemoglobin), a non-invasive monitoring device currently used in clinical settings. The device is handheld and connected to a computer/tablet for recording measurements. It will be placed on the child's skin to measure the child's hemoglobin. The procedure does not involve needles, radiation, or painful interventions.

Why is this study important?

Babies in NICU often have anemia, where their red blood cells level is low. To monitor this condition, babies often need repeated blood tests, which can cause pain and blood loss. This study may help reduce unnecessary blood sampling in the future.

What type of study is this?

This is an observational study. We will compare readings from a non-invasive device with routine blood test results.

What is the procedure that is being tested? A device called Masimo SpHb® is placed on the baby's skin to measure haemoglobin using light. No blood is taken for this measurement.

Tipo de estudo

Intervencional

Inscrição (Estimado)

30

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Estude backup de contato

  • Nome: Azanna Ahmad Kamar, MBBS, MRCPCH
  • Número de telefone: +603-79492065
  • E-mail: azanna@um.edu.my

Locais de estudo

    • Kuala Lumpur
      • Kuala Lumpur, Kuala Lumpur, Malásia, 50603
        • University Malaya Medical Center
        • Contato:
        • Contato:
          • Minn Yin Choo, MBBS, MMed, MRCPCH
          • Número de telefone: +603-79492065
          • E-mail: minnyin@ummc.edu.my
        • Investigador principal:
          • Aliyyah Mohammad Khuzaini, MBChB, MRCPCH, MMEd
        • Subinvestigador:
          • Minn Yin Choo, MBBS, MMed, MRCPCH
        • Subinvestigador:
          • Azanna Ahmad Kamar, MBBS, MRCPCH
        • Subinvestigador:
          • Tengku Amatullah Madeehah Tengku Mohd, MBChB, MSc, MPH, DrPH

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

  • Filho
  • Adulto
  • Adulto mais velho

Aceita Voluntários Saudáveis

Não

Descrição

Inclusion Criteria:

  • Infants (preterm or term) admitted to the Neonatal Intensive Care Unit.
  • Gestational age ≥ 24 weeks at birth.
  • Admitted to NICU during the study period.
  • Clinical indication for laboratory haemoglobin measurement as part of routine care.
  • Written informed consent obtained from parent or legal guardian.

Exclusion Criteria:

  • Known or suspected haemoglobinopathy.
  • Cyanotic congenital heart disease.
  • Significant skin integrity issues at the proposed sensor site (e.g. burns, severe dermatitis, wounds, oedema, or medical devices preventing probe placement).
  • Conditions likely to cause unreliable peripheral perfusion at the discretion of the attending clinician.
  • Severe hypothermia (below 32°C) or hyperthermia (more than 40°C).
  • Prolonged capillary refill time indicating poor peripheral perfusion.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Diagnóstico
  • Alocação: N / D
  • Modelo Intervencional: Atribuição de grupo único
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: SpHb
Infants admitted to the NICU will undergo non-invasive SpHb measurement, with results compared against laboratory haemoglobin values obtained as part of routine clinical care.
  1. The index test utilizes multi-wavelength spectrophotometry (Masimo Radical-7® with rainbow® sensors) applied topically to peripheral tissue (e.g., foot, palm, or digit). It requires zero skin punctures, creates no additional iatrogenic blood loss, and causes no pain or distress to fragile neonates.
  2. Unlike standard diagnostic studies that require dedicated blood sampling for research, this protocol collects reference laboratory total haemoglobin (tHb) strictly when clinically indicated for routine care. No extra phlebotomy is performed solely for the study.
  3. Non-invasive SpHb measurements are acquired within a strict 5-minute window of routine clinical blood sampling to eliminate physiological variability over time and ensure paired accuracy assessment.
Outros nomes:
  • Masimo Radical-7® Pulse CO-Oximeter with rainbow® technology

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Agreement between non-invasive SpHb and laboratory haemoglobin measurements
Prazo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age

Agreement between SpHb and laboratory haemoglobin will be assessed using Bland-Altman analysis, including mean bias and 95% limits of agreement.

Unit of measure:

g/dL

During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Sensitivity and specificity of SpHb for detection of anaemia
Prazo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age

Sensitivity and specificity of SpHb for detecting anaemia, using laboratory haemoglobin as the reference standard. Anaemia will be defined according to the pre-specified haemoglobin threshold appropriate for the infant's gestational/postnatal age.

Unit of measure:

Percentage of participants

During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
Agreement between changes in SpHb and changes in laboratory haemoglobin over time
Prazo: During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age

Description:

The ability of serial SpHb measurements to detect changes in haemoglobin concentration over time will be assessed by comparing changes in SpHb with corresponding changes in laboratory haemoglobin measurements.

Unit of measure:

g/dL

During NICU hospitalization, from enrolment until hospital discharge, up to 44 weeks postmenstrual age
Regression coefficient for perfusion index and SpHb-tHb measurement difference
Prazo: From enrollment through NICU discharge (up to 44 weeks postmenstrual age).

Regression coefficient (β) and 95% confidence interval for perfusion index (measured by pulse co-oximeter perfusion index display, %) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant.

Unit of Measure: Regression coefficient (β) per 1% change in perfusion index

From enrollment through NICU discharge (up to 44 weeks postmenstrual age).
Regression coefficient for gestational age and SpHb-tHb measurement difference
Prazo: Gestational age recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).

Regression coefficient (β) and 95% confidence interval for gestational age at birth (recorded in completed weeks from obstetric dating, per medical records) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant.

Unit of Measure: Regression coefficient (β) per completed week of gestational age

Gestational age recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
Regression coefficient for birth weight and SpHb-tHb measurement difference
Prazo: Birth weight recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).

Regression coefficient (β) and 95% confidence interval for birth weight (recorded in grams at delivery, per medical records) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant.

Unit of Measure: Regression coefficient (β) per 100 grams of birth weight

Birth weight recorded at birth; association assessed from enrollment through NICU discharge (up to 44 weeks postmenstrual age).
Regression coefficient for phototherapy and SpHb-tHb measurement difference
Prazo: Assessed at each paired SpHb-tHb measurement, from enrollment through NICU discharge (up to 44 weeks postmenstrual age).

Regression coefficient (β) and 95% confidence interval for concurrent phototherapy use (yes/no, recorded from NICU treatment records at the time of each paired measurement) as a predictor of SpHb-tHb measurement difference, from a mixed-effects linear regression model with random intercept per infant.

Unit of Measure: Regression coefficient (β) for phototherapy exposure (on vs off)

Assessed at each paired SpHb-tHb measurement, from enrollment through NICU discharge (up to 44 weeks postmenstrual age).

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Publicações e links úteis

A pessoa responsável por inserir informações sobre o estudo fornece voluntariamente essas publicações. Estes podem ser sobre qualquer coisa relacionada ao estudo.

Publicações Gerais

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

1 de setembro de 2026

Conclusão Primária (Estimado)

31 de março de 2027

Conclusão do estudo (Estimado)

30 de junho de 2027

Datas de inscrição no estudo

Enviado pela primeira vez

14 de agosto de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

24 de agosto de 2026

Primeira postagem (Real)

26 de agosto de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

26 de agosto de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

24 de agosto de 2026

Última verificação

1 de agosto de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • 2026219-16355
  • ID RSCH ID-26-05390-RJH (Identificador de registro: National Medical Research Registry, Malaysia)

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Descrição do plano IPD

To safeguard participant confidentiality and privacy, individual participant-level data will not be made publicly available. Final study findings will be published in aggregate format only

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Sim

produto fabricado e exportado dos EUA

Sim

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .