Enhanced Diagnostics for Early Detection of Atrial Fibrillation - Prospective Validation (DETECTAF-pro)
調査の概要
詳細な説明
Introduction This algorithm was validated earlier in a clinical trial with retrospective Data. The results were published accordingly.The algorithm is implemented in the commercial App Preventicus Heartbeats. DETECT AF is designed as prospective follow up trial.
Study design Prospective, blinded, multicenter international trial.
Methods
The subjects will be asked to position a smartphone on their index finger to allow their pulse curve to be recorded for five minutes. At the same time, an ecg will be recorded using an digital mobile ecg recorder as a reference.
Data will be collected on person-related information, comorbidities and medication (estimated overall duration/patient <20 min). No follow-up is planned. Analogous to predecessor studies, the pulse wave curve data are coded with the patient's identification number (ID) and externally analyzed at Preventicus; based on the results, the subjects will be assigned to SR/AF groups. After evaluation of all files, the grouped results will be aggregated, unblinded and evaluated under monitoring.
Primary target parameters are the app's sensitivity and specificity in correctly detecting atrial fibrillation compared to an automatically interpreted ECG.
Secondary target parameters include the proportion of non-evaluable recordings in the overall study.
研究の種類
入学 (実際)
連絡先と場所
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- Subjects of legal age,
- patients with either sinus rhythm (Group SR) or atrial fibrillation (Group AF)
- informed consent
Exclusion Criteria:
- Legally incompetent persons
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
AF
Five minutes puls wave recording will be performed in patients in atrial fibrillation at time of recruitment
|
iPhone will be positioned on the index finger for five minutes to record a pulse wave
|
|
SR
Five minutes puls wave recording will be performed in patients in sinus rhythm at time of recruitment
|
iPhone will be positioned on the index finger for five minutes to record a pulse wave
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Correct detection of atrial fibrillation
時間枠:5 minutes
|
iPhone will be positioned on the index finger for five minutes to record a pulse wave; Algorithm should correctly differentiate between AF patients and SR patients
|
5 minutes
|
協力者と研究者
捜査官
- 主任研究者:Jens Eckstein, MD, PhD、Universityhospital Basel
出版物と役立つリンク
一般刊行物
- Krivoshei L, Weber S, Burkard T, Maseli A, Brasier N, Kuhne M, Conen D, Huebner T, Seeck A, Eckstein J. Smart detection of atrial fibrillationdagger. Europace. 2017 May 1;19(5):753-757. doi: 10.1093/europace/euw125.
- Brasier N, Raichle CJ, Dorr M, Becke A, Nohturfft V, Weber S, Bulacher F, Salomon L, Noah T, Birkemeyer R, Eckstein J. Detection of atrial fibrillation with a smartphone camera: first prospective, international, two-centre, clinical validation study (DETECT AF PRO). Europace. 2019 Jan 1;21(1):41-47. doi: 10.1093/europace/euy176.
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは