Electrical Activation Mapping Guided Tailor Made Approach for Cardiac Resynchronization Therapy

2021年10月18日 更新者:Professor Bryan Ping Yen YAN、Chinese University of Hong Kong

Background

Cardiac Resynchronization Therapy (CRT) is proven to improve survival and heart function of patient with certain electrical conduction abnormality and heart failure. However, in patient with certain electrical conduction abnormality, a good response is observed in less than 40% in patient receiving CRT. Conventionally the surgical approach of CRT is to implant one pacing lead in the right heart and one in the left heart to resynchronize the contraction and the pacing lead in the left heart is usually placed in the posterior or lateral portion of the left heart. However, this single approach may not be optimal, especially for those patients with conduction abnormality known to have poor response to CRT.

Purpose of the clinical investigation

The purpose of the Electrical Activation Guided CRT Study is to study the effectiveness of a tailored made approach to CRT procedure by using a noninvasive globally mapping system studying the electrical conduction under different approaches to delivery CRT. The pacing approach that optimally corrects conduction abnormality will be determined before the actual implantation procedure.

Conduct of the Investigation

This study will include 93 patients with conduction abnormality that known to have a poor response to CRT from Prince of Wales Hospital, Hong Kong.Subjects will be followed up at 3 months and 6 months visit.

調査の概要

詳細な説明

The purpose is to prospectively study the feasibility to optimize configuration of CRT delivery for acute correction of electrical dyssynchrony using a noninvasive mapping of global electrical activation.

Study Hypothesis: Tailor-made configuration of CRT delivery is feasible and able to improve responder rate compare to single method of CRT delivery in candidates with known poor response to CRT.

Primary outcome measure: Responder rate of greater than 10% of LV end systolic volume reduction in patients undergoing tailor-made approach of CRT delivery at 6 months. The responder rate is to compare with pre-defined level of 40% for single method of CRT delivery namely biventricular pacing with LV lead in coronary sinus.

Sample Size: The total required sample size is 93 patients with device implanted.

研究の種類

介入

入学 (予想される)

93

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

      • Hong Kong、香港
        • 募集
        • The Chinese University of Hong Kong
        • コンタクト:
        • 副調査官:
          • Joseph YS Chan
        • 主任研究者:
          • Bryan PY Yan
        • 副調査官:
          • Gary CP Chan

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • Adult (aged 18 or above) of both sexes
  • Ischemic or non-ischemic cause of heart failure
  • QRS duration > 120 ms, non -LBBB type of conduction disturbance
  • NYHA class III or above
  • Sinus rhythm
  • Informed consent by the patient
  • Already received stable dose of guideline directed medical therapy for at least 3 months

Exclusion Criteria:

  • LBBB* patients
  • Pregnant women
  • Participation in another study
  • Patient with contraindication to left ventricle catheterization by a retrograde aortic approach (eg mechanical aortic valve, severe aortic stenosis and aortic dissection) *The definitions of LBBB (QRS duration ≥130 ms; QS or rS in lead V1; broad R waves in leads I, aVL, V5, or V6; and absent q waves in leads I, V5, and V6).

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:処理
  • 割り当て:なし
  • 介入モデル:単一グループの割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:Tailor-made CRT delivery
Patient undergoes acute noninvasive electrical dyssynchrony study with various CRT configurations. CRT device is then implanted with optimal configuration.

Ventricular activation maps will be acquired simultaneously with hemodynamic measurements using noninvasive mapping system (ECVUE, Medtronic Inc, USA). A thoracic computed tomographic scan will be acquired with the electrodes attached to the patient. Local ventricular activation times will be defined as the onset of the QRS complex or the pacing spike to the maximal negative slope of each unipolar electrogram.

Pacing leads will be placed in high right atrium, His-bundle region, right ventricular apex, high septal RV, coronary sinus posterior/lateral branch, coronary sinus anterior branch, lateral and septal region of endocardial LV in order to deliver CRT in 8 different configurations.

他の名前:
  • Noninvasive electrical activation
  • Electrical Activation Mapping
  • Noninvasive mapping system
  • ECVUE

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Responder rate of greater than 10% of LV end systolic volume reduction in patients undergoing tailor-made approach of CRT delivery at 6 months.
時間枠:6 months
Responder rate of greater than 10% of LV end systolic volume reduction measured by echocardiogram comparing baseline and 6 months post implant in patients undergoing tailor-made approach of CRT delivery. Responder rate of greater than 10% of LV end systolic volume reduction measured by echocardiogram comparing baseline and 6 months post implant in patients undergoing tailor-made approach of CRT delivery. The responder rate is to compare with pre-defined level of 40% for single method of CRT delivery namely biventricular pacing with LV lead in coronary sinus.
6 months

二次結果の測定

結果測定
メジャーの説明
時間枠
The acute electrical dyssynchrony indices of different methods of CRT delivery.
時間枠:during procedure
The acute electrical dyssynchrony indices of different methods of CRT delivery.
during procedure
The hemodynamic responses of different methods of CRT delivery.
時間枠:during procedure
The hemodynamic responses of different methods of CRT delivery. The hemodynamic response will be maximal dp/dt as measured by pressure wire introduced into the left ventricle during the procedure.
during procedure
Procedure duration of the optimal CRT delivery method
時間枠:during procedure
Procedure duration of the optimal CRT delivery method as determined by the best improvement in electrical dyssynchrony indices.
during procedure
Implantation success rate of the optimal CRT delivery method
時間枠:during procedure
Implantation success rate of the optimal CRT delivery method as determined by the best improvement in electrical dyssynchrony indices.
during procedure
Cine images (PA, LAO 300, RAO 300) and Chest X ray (PA view)
時間枠:during procedure
Cine images (PA, LAO 300, RAO 300) and Chest X ray (PA view)
during procedure
Peri-operative and 6 months follow-up complications rate:
時間枠:Peri-operative and 6 months
  1. Thromboembolic event
  2. Dislodgement and migration of pacing leads
  3. Phrenic nerve stimulation
  4. Others
Peri-operative and 6 months
Left ventricular systolic and diastolic volume at baseline and 6 months:
時間枠:baseline and 6 months
Echocardiogram parameter of left ventricular systolic and diastolic volume at baseline and 6 months.
baseline and 6 months
Left ventricular ejection fraction at baseline and 6 months:
時間枠:baseline and 6 months
Echocardiogram parameter of left ventricular ejection fraction at baseline and 6 months.
baseline and 6 months
Degree of mitral regurgitation at baseline and 6 months:
時間枠:baseline and 6 months
Echocardiogram parameter of degree of mitral regurgitation at baseline and 6 months.
baseline and 6 months
Strain imaging at baseline and 6 months:
時間枠:baseline and 6 months
Echocardiogram parameter of strain imaging at baseline and 6 months.
baseline and 6 months
NYHA class
時間枠:baseline and 6 months
NYHA class at baseline and 6 months.
baseline and 6 months
6 minute hall walk test
時間枠:baseline and 6 months
6 minute hall walk test at baseline and 6 months.
baseline and 6 months
Quality of life using Minnesota's questionnaire
時間枠:baseline and 6 months
Quality of life using Minnesota's questionnaire at baseline and 6 months.
baseline and 6 months
Electrical parameters including threshold, sensitivity and lead impedance of pacing leads at implant and 6 months follow-up.
時間枠:during procedure and 6 months
Electrical parameters including threshold, sensitivity and lead impedance of pacing leads at implant and 6 months follow-up.
during procedure and 6 months

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:Bryan PY YAN、Chinese University of Hong Kong

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

一般刊行物

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2018年3月8日

一次修了 (予想される)

2024年8月31日

研究の完了 (予想される)

2024年8月31日

試験登録日

最初に提出

2017年11月8日

QC基準を満たした最初の提出物

2017年11月23日

最初の投稿 (実際)

2017年11月29日

学習記録の更新

投稿された最後の更新 (実際)

2021年10月19日

QC基準を満たした最後の更新が送信されました

2021年10月18日

最終確認日

2021年10月1日

詳しくは

本研究に関する用語

追加の関連 MeSH 用語

その他の研究ID番号

  • CUHK-MAPCRT-1

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する