Validation of Correction Algorithms for Near-IR Analysis of Human Milk in an Independent Sample Set-Effect of Pasteurization

Gynter Kotrri, Gerhard Fusch, Celia Kwan, Dasol Choi, Arum Choi, Nisreen Al Kafi, Niels Rochow, Christoph Fusch, Gynter Kotrri, Gerhard Fusch, Celia Kwan, Dasol Choi, Arum Choi, Nisreen Al Kafi, Niels Rochow, Christoph Fusch

Abstract

Commercial infrared (IR) milk analyzers are being increasingly used in research settings for the macronutrient measurement of breast milk (BM) prior to its target fortification. These devices, however, may not provide reliable measurement if not properly calibrated. In the current study, we tested a correction algorithm for a Near-IR milk analyzer (Unity SpectraStar, Brookfield, CT, USA) for fat and protein measurements, and examined the effect of pasteurization on the IR matrix and the stability of fat, protein, and lactose. Measurement values generated through Near-IR analysis were compared against those obtained through chemical reference methods to test the correction algorithm for the Near-IR milk analyzer. Macronutrient levels were compared between unpasteurized and pasteurized milk samples to determine the effect of pasteurization on macronutrient stability. The correction algorithm generated for our device was found to be valid for unpasteurized and pasteurized BM. Pasteurization had no effect on the macronutrient levels and the IR matrix of BM. These results show that fat and protein content can be accurately measured and monitored for unpasteurized and pasteurized BM. Of additional importance is the implication that donated human milk, generally low in protein content, has the potential to be target fortified.

Keywords: analyzer; donor milk; fat; lactose; milk; nutrition; preterm; protein; term.

Figures

Figure 1
Figure 1
Study Design: To validate the Near-IR human milk analyzer in unpasteurized breast milk, group B samples were compared to the corresponding group A samples. The correction algorithm was also tested for pasteurized breast milk, through a comparison of all group D samples with the corresponding samples in group C. The effect of pasteurization on macronutrient levels was determined through two separate comparisons: one between groups A and C, and another between all group D samples and the corresponding samples in group B.
Figure 2
Figure 2
Fat (A) (n = 20) and protein (B) (n = 20) values as measured through chemical reference methods correlated against corrected Near-IR human milk analyzer values. Bland–Altman plots for fat (C) and protein (D) indicating the difference obtained by the reference method (x-axis) and the corrected Near-IR human milk analyzer values represented in y-axis.
Figure 3
Figure 3
Fat (A) (n = 10) and protein (B) (n = 10) values in pasteurized breast milk as measured through chemical reference methods correlated against corrected Near-IR human milk analyzer values; Bland–Altman plots for fat (C) and protein (D) indicating the difference in pasteurized breast milk obtained by the reference method (x-axis) and the corrected Near-IR human milk analyzer values represented in y-axis.
Figure 4
Figure 4
(A) shows the correlation between samples of pasteurized and unpasteurized breast milk (n = 50) as measured through Near-IR milk analyzers and subsequently adjusted with the correction algorithm; (B) shows a similar correlation between pasteurized and unpasteurized samples (n = 10) as measured through validated chemical reference methods.

References

    1. American Academy of Pediatrics Policy Statement Breastfeeding and the use of human milk. Pediatrics. 2012;129:827–841.
    1. Polberger S.K., Axelsson I.A., Raiha N.C. Growth of very low birth weight infants on varying amounts of human milk protein. Pediatr. Res. 1989;25:414–419. doi: 10.1203/00006450-198904000-00022.
    1. Kashyap S., Schulze K.F., Forsyth M., Dell R.B., Ramakrishnan R., Heird W.C. Growth, nutrient retention, and metabolic response of low-birth-weight infants fed supplemented and unsupplemented preterm human milk. Am. J. Clin. Nutr. 1990;52:254–262.
    1. Clark R., Thomas P., Peabody J. Extrauterine Growth Restriction Remains a Serious Problem in Prematurely Born Neonates. Pediatrics. 2003;111:986–990. doi: 10.1542/peds.111.5.986.
    1. Rochow N., Fusch G., Choi A., Chessell L., Elliott L., McDonald K., Kuiper E., Purcha M., Turner S., Chan E., et al. Target Fortification of Breast Milk with Fat, Protein, and Carbohydrates for Preterm Infants. J. Pediatr. 2013;163:1001–1007. doi: 10.1016/j.jpeds.2013.04.052.
    1. Fusch G., Rochow N., Choi A., Fusch S., Poeschl S., Ubah A., Lee S.Y., Raja P., Fusch C. Rapid measurement of macronutrients in breast milk: How reliable are infrared milk analyzers? Clin. Nutr. 2014;34:465–476. doi: 10.1016/j.clnu.2014.05.005.
    1. Gartner L.M., Morton J., Lawrence R.A., Naylor A.J., O’Hare D., Schanler R.J., Eidelman A.I. American Academy of Pediatrics Section on Breastfeeding. Breastfeeding and the use of human milk. Pediatrics. 2005;115:496–506.
    1. Vieira A.A., Soares F.V., Pimenta H.P., Abranches A.D., Moreira M.E. Analysis of the influence of pasteurization, freezing/thawing, and offer processes on human milk’s macronutrient concentrations. Early Hum. Dev. 2011;87:577–580. doi: 10.1016/j.earlhumdev.2011.04.016.
    1. Lepri L., Del Bubba M., Maggini R., Donzelli G., Galvan P. Effect of pasteurization and storage on some components of pooled human milk. J. Chromatogr. B. 1997;704:1–10. doi: 10.1016/S0378-4347(97)00439-8.
    1. Fidler N., Sauerwald T., Koletzko B., Demmelmair H. Effects of Human Milk Pasteurization and Sterilization on Available Fat Content and Fatty Acid Composition. J. Pediatr. Gastroenterol. Nutr. 1998;27:317–322. doi: 10.1097/00005176-199809000-00009.
    1. Wardell J., Hill C., D’Souza S. Effect of Pasteurization and of Freezing and Thawing Human Milk on its Triglyceride Content. Acta Paediatr. 1981;70:467–471. doi: 10.1111/j.1651-2227.1981.tb05724.x.
    1. Henderson T., Fay T., Hamosh M. Effect of pasteurization on long chain polyunsaturated fatty acid levels and enzyme activities of human milk. J. Pediatr. 1998;132:876–878. doi: 10.1016/S0022-3476(98)70323-3.
    1. Choi A., Fusch G., Rochow N., Sheikh N., Fusch C. Establishment of micromethods for macronutrient contents analysis in breast milk. Mater. Child Nutr. 2013;34:465–476. doi: 10.1111/mcn.12053.
    1. Fusch G., Choi A., Rochow N., Fusch C. Quantification of lactose content in human and cow’s milk using UPLC-tandem mass spectrometry. J. Chromatog. B. 2011;879:3759–3762. doi: 10.1016/j.jchromb.2011.09.053.
    1. Krouwer J.S. Why Bland-Altman plots should use X, not (Y+X)/2 when X is a reference method. Stat. Med. 2008;27:778–780. doi: 10.1002/sim.3086.
    1. Sauer C.W., Kim J.H. Human milk macronutrient analysis using point-of-care near-infrared spectrophotometry. J. Perinatol. 2011;31:339–343. doi: 10.1038/jp.2010.123.
    1. Smilowitz J.T., Gho D.S., Mirmiran M., German J.B., Underwood M.A. Rapid measurement of human milk macronutrients in the neonatal intensive care unit: Accuracy and precision of Fourier transform mid-infrared spectroscopy. J. Hum. Lact. 2014;30:180–189. doi: 10.1177/0890334413517941.
    1. Billard H., Simon L., Desnots E., Sochard A., Boscher C., Riaublanc A., Alexandre-Gouabau M.C., Boquien C.Y. Calibration Adjustment of the Mid-infrared Analyzer for an Accurate Determination of the Macronutrient Composition of Human Milk. J. Hum. Lact. 2015 doi: 10.1177/0890334415588513.
    1. Menjo A., Mizurno K., Murase M., Nishida Y., Taki M., Itabashi K., Shimono T., Namba K. Bedside analysis of human milk for adjustable nutrition strategy. Acta Pediatr. 2009;98:380–384. doi: 10.1111/j.1651-2227.2008.01042.x.
    1. Casadio Y.S., Williams T.M., Lai C.T., Olsson S.E., Hepworth A.R., Harmann P.E. Evaluation of a mid-infrared analyzer for the determination of the macronutrient composition of human milk. J. Hum. Lact. 2010;26:376–383. doi: 10.1177/0890334410376948.
    1. Silvestre D., Fraga M., Gormaz M., Torres E., Vento M. Comparison of mid-infrared transmission spectroscopy with biochemical methods for the determination of macronutrients in human milk. Mater. Child Nutr. 2012;10:373–382. doi: 10.1111/j.1740-8709.2012.00431.x.
    1. Corvaglia L., Battistini B., Paoletti V., Aceti A., Capretti M.G., Faldella G. Near-infrared reflectance analysis to evaluate the nitrogen and fat content of human milk in neonatal intensive care units. Arch. Dis. Child. Fetal Neonatal Ed. 2008;93:F372–F375. doi: 10.1136/adc.2007.133280.
    1. ISO 22662:2007 (IDF 198: 2007). Milk and Milk Products—Determination of Lactose Content by High-Performance Liquid Chromatography (Reference Method) [(accessed on 1 December 2015)]. Available online: .
    1. Espinosa-Martos I., Montilla A., Segura A., Escuder D., Bustos G., Pallás C., Rodríguez J.M., Corzo N., Fernández L. Bacteriological, Biochemical, and Immunological Modifications in Human Colostrum after Holder Pasteurisation. J. Pediatr. Gastroenterol. Nutr. 2013;56:560–568. doi: 10.1097/MPG.0b013e31828393ed.
    1. Valentine C., Morrow G., Fernandez S., Gulati P., Bartholomew D., Long D., Welty S.E., Morrow A.L., Rogers L.K. Docosahexaenoic Acid and Amino Acid Contents in Pasteurized Donor Milk are Low for Preterm Infants. J. Pediatr. 2010;157:906–910. doi: 10.1016/j.jpeds.2010.06.017.
    1. Boyd C., Quigley M., Brocklehurst P. Donor breast milk versus infant formula for preterm infants: Systematic review and meta-analysis. Arch. Dis. Child. Fetal Neonatal Ed. 2007;92:F169–F175. doi: 10.1136/adc.2005.089490.
    1. Wojcik K., Rechtman D., Lee M., Montoya A., Medo E. Macronutrient Analysis of a Nationwide Sample of Donor Breast Milk. J. Am. Diet. Assoc. 2009;109:137–140. doi: 10.1016/j.jada.2008.10.008.

Source: PubMed

3
購読する