Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population

Fernando Grover-Páez, Ernesto G Cardona-Muñoz, David Cardona-Müller, Víctor H Guzmán-Saldívar, Mariana Rodríguez-De la Cerda, Mayra B Jiménez-Cázarez, Sylvia E Totsuka-Sutto, Guillermo A Alanis-Sánchez, Carlos G Ramos-Becerra, Fernando Grover-Páez, Ernesto G Cardona-Muñoz, David Cardona-Müller, Víctor H Guzmán-Saldívar, Mariana Rodríguez-De la Cerda, Mayra B Jiménez-Cázarez, Sylvia E Totsuka-Sutto, Guillermo A Alanis-Sánchez, Carlos G Ramos-Becerra

Abstract

Objective: The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010.

Participants and methods: The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5).

Results: The device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.

Figures

Fig. 1
Fig. 1
Plot of the systolic blood pressure device–observer difference of the 33 participants included in this study. The y-axis represents the difference of measurements between the device and the observer. The x-axis represents the mean of the device and observer measurements. SBP, systolic blood pressure.
Fig. 2
Fig. 2
Plot of the diastolic blood pressure device–observer difference of the 33 participants included in this study. The y-axis represents the difference in measurements between the device and the observer. The x-axis represents the mean of the device and observer measurements. DBP, diastolic blood pressure.

References

    1. Rojas-Martínez R, Aquilar-Salinas CA, Jiménez-Corona A, Gómez-Pérez FJ, Barquera S, Lazcano-Ponce E. Prevalence of obesity and metabolic syndrome components in Mexican adults without type 2 diabetes or hypertension. Salud Publica Mex 2012; 54:7–12.
    1. Lawes CM, van der Hoorn S, Rodgers A. Global burden of blood-pressure-related disease, 2001. The Lancet 2008; 371:1513–1518.
    1. Mendis S, Abequnde D, Oladapo O, Celletti F, Nordet P. Barriers to management of cardiovascular risk in a low-resource setting using hypertension as an entry point. J Hypertens 2004; 22:59–64.
    1. Lawes CM, vander Hoorn S, Law MR, Elliott P, MacMahon S, Rodgers A. Blood pressure and the global burden of disease 2000. Part II: estimates of attributable burden. J Hypertens 2006; 24:423–430.
    1. Kearney PM, Whelton M, Reynolds K, Whelton PK, He J. Worldwide prevalence of hypertension: a systematic review. J Hypertens 2004; 22:11–19.
    1. Lopez AD, Mathers CD, Ezzati M, Jamison DT, Murray CJ. Global and regional burden of disease and risk factors, 2001: systematic analysis of population health data. The Lancet 2006; 367:1747–1757.
    1. Whitworth JA. World Health Organization, International Society of Hypertension Writing Group. 2003 World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension. J Hypertens 2003; 21:1983–1992.
    1. Bilo G, Sala O, Perego C, Faini A, Gao L, Gluszewska A, et al. Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference? Hypertens Res 2017; 40:573–580.
    1. O’brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit 2010; 15:23–38.

Source: PubMed

3
購読する