Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol

Jirar Topouchian, Davide Agnoletti, Jacques Blacher, Ahmed Youssef, Mirna N Chahine, Isabel Ibanez, Nathalie Assemani, Roland Asmar, Jirar Topouchian, Davide Agnoletti, Jacques Blacher, Ahmed Youssef, Mirna N Chahine, Isabel Ibanez, Nathalie Assemani, Roland Asmar

Abstract

Background: Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies.

Methods: The four devices measure brachial blood pressure (BP) using the oscillometric method. The Withings BP-800 has to be connected to an Apple® iOS device such as an iPhone®, iPad®, or iPod®. The ESH International Protocol Revision 2010 includes a total number of 33 subjects. The difference between observer and device BP values was calculated for each measure. Ninety-nine pairs of BP differences were classified into three categories (≤5 mmHg, ≤10 mmHg, ≤15 mmHg). The protocol procedures were followed precisely in each of the four studies.

Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were: -1.8±5.1 mmHg and -0.4±2.8 mmHg for systolic and diastolic BP, respectively, using the Omron M6 Comfort device; 2.5±4.6 mmHg and -1.2±4.3 mmHg for the Omron HEM-7420 device; -0.2±5.0 mmHg and 0.4±4.2 mmHg for the Withings BP-800 device; and 3.0±5.3 mmHg and 0.3±5.2 mmHg for the Polygreen KP-7670 device.

Conclusion: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 readings differing by less than 5 mmHg, 10 mmHg, and 15 mmHg fulfill the ESH International Protocol Revision 2010 requirements, and therefore are suitable for use by patients for SBPM, if used correctly.

Keywords: European Society of Hypertension International Protocol Revision; Omron HEM-7420; Omron M6 Comfort; Polygreen KP-7670; Withings BP-800; self-blood pressure measurement; validation.

Figures

Figure 1
Figure 1
Plots showing difference in blood pressure between the Omron M6 Comfort (HEM-7221-E) (Omron Healthcare, Kyoto, Japan) readings and the mean of two observer readings in 33 participants (n=99). (A) SBP and (B) DBP. Points in bold are multiple (superimposition). Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure.
Figure 2
Figure 2
Plots showing difference in blood pressure between the Omron HEM-7420 (Omron Healthcare, Kyoto, Japan) readings and the mean of the two observer readings in 33 participants (n=99). (A) SBP and (B) DBP. Points in bold are multiple (superimposition). Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure.
Figure 3
Figure 3
Plots showing difference in blood pressure between the Withings BP-800 (Withings, Paris, France) readings and the mean of the two observer readings in 33 participants (n=99). (A) SBP and (B) DBP. Points in bold are multiple (superimposition). Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure.
Figure 4
Figure 4
Plots showing difference in blood pressure between the Polygreen KP-7670 (Polygreen K-Jump, Taipei, Taiwan) readings and the mean of the two observer readings in 33 participants (n=99). (A) SBP and (B) DBP. Points in bold are multiple (superimposition). Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure.

References

    1. Mancia G, Fagard R, Narkiewicz K, et al. 2013 ESH/ESC Guidelines for the Management of Arterial Hypertension. The Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC) J Hypertens. 2013;31:1281–1357.
    1. Chobanian A, Bakris G, Black H, et al. Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation and Treatment of High Blood Pressure. Hypertension. 2003;42:1206–1252.
    1. Asmar R, Zanchetti A. Guidelines for the use of self-blood pressure monitoring: a summary report of the first international consensus conference. J Hypertens. 2000;18:493–508.
    1. Parati G, Stergiou G, Asmar R, et al. European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring. J Hypertens. 2008;26:1505–1530.
    1. O’Brien E, Asmar R, Beilin L, et al. on behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens. 2005;23:697–701.
    1. National Institute for Health and Care Excellence Clinical guidelines 127. Hypertension: clinical management of primary hypertension in adults. 2011. [Accessed November 12, 2013]. Available from: .
    1. Pickering T, Hall J, Lawrence A, et al. Recommendations for blood pressure measurement in humans and experimental animals. Part 1. A statement for professionals from the American Heart Association Council on high blood pressure research. Hypertension. 2005;45:142–161.
    1. O’Brien E, Petrie J, Littler WA, et al. Blood Pressure Measurement Recommendations of the British Hypertension Society. London, UK: BMJ Publishing Group; 1997.
    1. O’Brien E, Petrie J, Littler WA, et al. The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993;11(Suppl 2):S43–S62.
    1. Association for the Advancement of Medical Instrumentation American National Standard: non-invasive sphygmomanometers – part 2: clinical validation of automated measurement type; ANSI/AAMI/ISO. 2009. [Accessed December 12, 2013]. pp. 81060–81062. Available from: .
    1. O’Brien E, Pickering T, Asmar R, et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension International protocol for validation of blood pressure measuring devices in adults. Blood Press Monit. 2002;7:3–17.
    1. O’Brien E, Atkins N, Stergiou G, et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol Revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010;15:23–38.
    1. RanD, editor. Groupe d’Evaluation et de Mesure de la Société Française d’Hypertension Artérielle. La prise de la tension artérielle au cabinet edical [Measurements of blood pressure in clinic] Meudon, France: 2001.
    1. Viera A, Hinderliter A. Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International protocol. Blood Press Monit. 2007;12:335–338.
    1. Bonso E, Saladini F, Zanier A, et al. Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences. Hypertens Res. 2010;11:1186–1191.

Source: PubMed

3
購読する