Laryngeal mask airway Unique™ position in paediatric patients undergoing magnetic resonance imaging (MRI): prospective observational study

Jozef Klučka, Jan Šenkyřík, Jarmila Skotáková, Roman Štoudek, Michaela Ťoukalková, Ivo Křikava, Lukáš Mareček, Tomáš Pavlík, Alena Štouračová, Petr Štourač, Jozef Klučka, Jan Šenkyřík, Jarmila Skotáková, Roman Štoudek, Michaela Ťoukalková, Ivo Křikava, Lukáš Mareček, Tomáš Pavlík, Alena Štouračová, Petr Štourač

Abstract

Background: Laryngeal mask UNIQUE® (LMAU) is supraglottic airway device with good clinical performance and low failure rate. Little is known about the ideal position of the LMAU on the magnetic resonance imaging (MRI) and whether radiological malposition can be associated with clinical performance (audible leak) in children. The primary aim of the study was to evaluate incidence of the radiologic malposition of the LMAU according to size. The secondary outcome was the clinical performance and associated complications (1st attempt success rate, audible leak) in LMAUs in correct position vs. radiologically misplaced LMAUs.

Methods: In prospective observational study, all paediatric patients undergoing MRI of the brain under general anaesthesia with the LMAU were included (1.9.2016-16.5.2017). The radiologically correct position: LMAU in hypopharynx, proximal cuff opposite to the C1 or C2 and distance A (proximal cuff end and aditus laryngis) ≤ distance B (distal cuff end and aditus laryngis). Malposition A: LMAU outside the hypopharynx. Malposition B: proximal cuff outside C1-C2. Malposition C: distance A ≥ distance B. We measured distances on the MRI image. Malposition incidence between LMAU sizes and first attempt success rate in trainees and consultant groups was compared using Fisher exact test, difference in incidence of malpositions using McNemar test and difference in leakage according to radiological position using two-sample binomial test.

Results: Overall 202 paediatric patients were included. The incidence of radiologically defined malposition was 26.2% (n = 53). Laryngeal mask was successfully inserted on the 1st attempt in 91.1% (n = 184) cases. Audible leak was detected in 3.5% (n = 7) patients. The radiologically defined malposition was present in 42.9% (n = 3) cases with audible leak. The rate of associated complications was 1.5% (n = 3): laryngospasm, desaturation, cough. In 4.0% (n = 8) the LMAU was soiled from blood. Higher incidence of radiological malposition was in LMAU 1.0, 1.5 and LMAU 3, 4 compared to LMAU 2 or LMAU 2.5 (p < 0.001).

Conclusion: Malposition was not associated with impaired clinical performance (audible leak, complications) of the LMAU or the need for alternative airway management.

Trial registration: Clinicaltrials.gov (NCT02940652) Registered 18 October 18 2016.

Keywords: Laryngeal mask; Magnetic resonance imaging; Paediatric anaesthesia; Position.

Conflict of interest statement

Ethics approval and consent to participate

Study was approved by Local Ethics Committee of University Hospital Brno, Czech Republic (9/2016) with waiver to obtain written informed consent.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Figures

Fig. 1
Fig. 1
Laryngeal masks positions on MRI image. a Correct position of the LM; b Malposition A; c Malposition B; d Malposition C

References

    1. Goldmann K. Recent developments in airway management of the paediatric patient. Curr Opin Anaesthesiol. 2006;19:278–284. doi: 10.1097/01.aco.0000192786.93386.d5.
    1. Luce V, Harkouk H, Brasher C, et al. Supraglottic airway devices vs tracheal intubation in children: a quantitative metaanalysis of respiratory complications. Paediatr Anaesth. 2014;24:1088–1098. doi: 10.1111/pan.12495.
    1. Lopez-Gil M, Brimacombe J, Alvarez M. Safety and efficacy of the laryngeal mask airway. A prospective survey of 1400 children. Anaesthesia. 1996;51:969–972. doi: 10.1111/j.1365-2044.1996.tb14968.x.
    1. Lopez-Gil M, Brimacombe J, Cebrian J, Arranz J. Laryngeal mask airway in pediatric practice: a prospective study of skill acquisition by anesthesia residents. Anesthesiology. 1996;84:807–811. doi: 10.1097/00000542-199604000-00007.
    1. Jagannathan Narasimhan, Ramsey Melissa A., White Michelle C., Sohn Lisa. An update on newer pediatric supraglottic airways with recommendations for clinical use. Pediatric Anesthesia. 2015;25(4):334–345. doi: 10.1111/pan.12614.
    1. Verghese C, Berlet J, Kapila A, Pollard R. Clinical assessment of the single use laryngeal mask airway—the LMA-unique. British Journal of Anaesthesia. 1998;80(5):677–679. doi: 10.1093/bja/80.5.677.
    1. Brimacombe J, von Goedecke A, Keller C, Brimacombe L, Brimacombe M. The laryngeal mask airway unique versus the soft seal laryngeal mask: a randomized, crossover study in paralyzed, anesthetized patients. Anesth Analg. 2004;99(5):1560–3; table of contents.
    1. Goudsouzian Nishan G., Denman William, Cleveland Robert, Shorten George. Radiologic Localization of the Laryngeal Mask Airway in Children. Anesthesiology. 1992;77(6):1085–1089. doi: 10.1097/00000542-199212000-00007.
    1. Brain A.I.J. Laryngeal mask misplacement?causes, consequences and solutions. Anaesthesia. 1992;47(6):531–532. doi: 10.1111/j.1365-2044.1992.tb02285.x.
    1. MONCLUS ENRIC, GARCÉS ANTONIO, DE JOSE MARIA BELEN, ARTÉS DAVID, MABROCK MAGED. Study of the adjustment of the Ambu laryngeal mask under magnetic resonance imaging. Pediatric Anesthesia. 2007;17(12):1182–1186. doi: 10.1111/j.1460-9592.2007.02330.x.
    1. VIALET RENAUD, NAU ANDRÉ, CHAUMOÎTRE KATHIA, MARTIN CLAUDE. Effects of head posture on the oral, pharyngeal and laryngeal axis alignment in infants and young children by magnetic resonance imaging. Pediatric Anesthesia. 2008;18(6):525–531. doi: 10.1111/j.1460-9592.2008.02530.x.
    1. R Core Team. R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2017. URL .
    1. Pournajafian A, Alimian M, Rokhtabnak F, Ghodraty M, Mojri M. Success rate of airway devices insertion: laryngeal mask airway versus supraglottic gel device. Anesth Pain Med. 2015;5(2):e22068. 10.5812/aapm.22068.
    1. BRAIN A.I.J. THE LARYNGEAL MASK—A NEW CONCEPT IN AIRWAY MANAGEMENT. British Journal of Anaesthesia. 1983;55(8):801–805. doi: 10.1093/bja/55.8.801.