Comparison of corneal measurements in keratoconus using swept-source optical coherence tomography and combined Placido-Scheimpflug imaging

Tommy C Y Chan, Sayantan Biswas, Marco Yu, Vishal Jhanji, Tommy C Y Chan, Sayantan Biswas, Marco Yu, Vishal Jhanji

Abstract

Purpose: The aim of this study was to provide a comprehensive comparison of reliability of corneal topographic measurements in keratoconic eyes using swept-source optical coherence tomography (OCT) and a combined Placido-Scheimpflug imaging.

Methods: A total of 30 eyes of 30 patients were included. The mean age was 31.2 ± 8.4 years. Two consecutive topographic measurements were obtained for one eye of each patient using swept-source OCT (CASIA) and combined Placido-Scheimpflug imaging (TMS-5). Test-retest reliability of CASIA and TMS-5 measurements including central corneal thickness (CCT) and thinnest corneal thickness (TCT), keratometry at steep (Ks) and flat (Kf) axes, average keratometry (Avg K), cylinder, and, best-fit spheres (BFS) of the anterior and posterior corneal surfaces were evaluated.

Results: There was no systematic or scaling bias in any parameter in both devices. Systematic differences between CASIA and TMS-5 were found in posterior corneal Kf, Avg K and BFS, CCT and TCT (p ≤ 0.002); scaling differences between CASIA and TMS-5 were also found in CCT and TCT (p ≤ 0.002). Both machines illustrated adequate reliability. Intraclass correlation coefficients (ICC) ≥0.952 was recorded for all parameters measured with CASIA and ICC ≥ 0.914 was recorded for all parameters on TMS-5. CASIA showed significantly higher ICCs in CCT and TCT, and posterior corneal BFS (p < 0.001).

Conclusions: This study showed significant differences in posterior corneal surface and corneal thickness measurements between swept-source OCT and combined Placido-Scheimpflug imaging in eyes with keratoconus. Swept-source OCT might be preferred over Placido-Scheimpflug imaging owing to better repeatability of measurements.

Keywords: combined Placido-Scheimpflug imaging; keratoconus; repeatability; swept-source optical coherence tomography.