Attain Model 4196 Left Ventricular (LV) Lead
2009년 9월 24일 업데이트: Medtronic Cardiac Rhythm and Heart Failure
Heart failure is a progressive disease that decreases the pumping action of the heart.
This may cause a backup of fluid in the heart and may result in heart beat changes.
When there are changes in the heart beat sometimes an implantable heart device is used to control the rate and rhythm of the heart beat.
In certain heart failure cases, when the two lower chambers of the heart no longer beat in a coordinated manner, cardiac resynchronization therapy (CRT) may be prescribed.
CRT is similar to a pacemaker.
It is placed (implanted) under the skin of the upper chest.
CRT is delivered as tiny electrical pulses to the right and left ventricles through three or four leads (flexible insulated wires) that are inserted through the veins to the heart.
The purpose of this study is to evaluate a new lead for delivering energy to the left ventricle (bottom left chamber of the heart).
연구 개요
연구 유형
중재적
등록 (실제)
190
단계
- 해당 없음
연락처 및 위치
이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.
연구 장소
-
-
Alaska
-
Anchorage, Alaska, 미국
-
-
Arizona
-
Phoenix, Arizona, 미국
-
-
California
-
Long Beach, California, 미국
-
Palo Alto, California, 미국
-
-
Iowa
-
Davenport, Iowa, 미국
-
-
Kentucky
-
Lexington, Kentucky, 미국
-
-
Louisiana
-
Baton Rouge, Louisiana, 미국
-
-
Michigan
-
Detroit, Michigan, 미국
-
-
Minnesota
-
Minneapolis, Minnesota, 미국
-
-
Missouri
-
Kansas City, Missouri, 미국
-
-
North Carolina
-
Raleigh, North Carolina, 미국
-
-
Ohio
-
Mayfield Heights, Ohio, 미국
-
-
Pennsylvania
-
Wynnewood, Pennsylvania, 미국
-
-
South Carolina
-
Spartanburg, South Carolina, 미국
-
-
Tennessee
-
Kingsport, Tennessee, 미국
-
Memphis, Tennessee, 미국
-
-
Texas
-
Dallas, Texas, 미국
-
Fort Worth, Texas, 미국
-
-
Vermont
-
Burlington, Vermont, 미국
-
-
-
-
Alberta
-
Calgary, Alberta, 캐나다
-
Edmonton, Alberta, 캐나다
-
-
Ontario
-
Toronto, Ontario, 캐나다
-
-
Quebec
-
Montreal, Quebec, 캐나다
-
Sainte-Foy, Quebec, 캐나다
-
-
참여기준
연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.
자격 기준
공부할 수 있는 나이
18년 이상 (성인, 고령자)
건강한 자원 봉사자를 받아들입니다
아니
연구 대상 성별
모두
설명
Inclusion Criteria:
- QRS greater than or equal to 120 milliseconds (The QRS interval is a measurement of how the electrical signal involved in a heart beat travels, or is conducted, through the ventricles. A wide QRS (more than 120) suggests that there is a conduction problem, or block, in the ventricles.)
- Left Ventricular Ejection Fraction (EF) less than or equal to 35 percent (Ejection Fraction is a measurement of how well the left ventricle pumps blood out to the rest of the body. The higher the EF the more blood the ventricle is pumping.)
- Subject has moderate to severe heart failure despite medications
Exclusion Criteria:
- Subjects with a previous lead in the left ventricle or previous implant attempt within 30 days of implant or ongoing complications from a previous unsuccessful attempt
- Subjects with chest pain or who have had a heart attack within the past month before enrollment in the study
- Subjects that have had certain surgeries on their heart within the past three months
- Subjects with chronic (permanent) fast heart beats in the upper chambers of the heart (atrial arrhythmias)
공부 계획
이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 치료
- 할당: 무작위화되지 않음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: 4196 Lead
Non-randomized study.
|
연구 장치의 임플란트 및 추적
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Safety (Subjects Without a Model 4196 Lead Related Complication)
기간: One Month
|
A subject who was free of a Model 4196 lead related complication by one month visit.
|
One Month
|
|
Efficacy (Pacing Voltage Thresholds of Distal Tip Electrode)
기간: One Month
|
Model 4196 lead distal tip electrode mean pacing voltage threshold (at 0.5 milliseconds [ms])
|
One Month
|
|
Efficacy (Pacing Voltage Threshold of Proximal Ring Electrode)
기간: Three Months
|
Model 4196 lead proximal ring electrode mean pacing voltage threshold (at 0.5 milliseconds [ms])
|
Three Months
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Subjects Successfully Implanted With Model 4196 Lead
기간: Implant
|
A successful implant occurs when the Model 4196 lead is implanted in the left ventricle of the heart and functions appropriately.
|
Implant
|
|
Subjects Successfully Implanted After Cannulation
기간: Implant
|
A successful implant occurs when the coronary sinus (CS) is successfully cannulated and a left ventricular lead is implanted in the left ventricle of the heart and functions appropriately.
|
Implant
|
|
All Left Ventricular Leads
기간: Implant
|
All left ventricular leads successfully implanted
|
Implant
|
|
All Medtronic Left Ventricular Leads (Attain Family)
기간: Implant
|
All Medtronic left ventricular leads (Lead Model Numbers included 4193, 4194, 4195 and 4196) successfully implanted
|
Implant
|
|
Cannulation Time
기간: Implant
|
Cannulation time was defined as the time from insertion of the first CS cannulation catheter to the first CS cannulation.
|
Implant
|
|
Fluoroscopy Time
기간: Implant
|
Fluoroscopy time was defined as the total time the fluoroscope was imaging.
|
Implant
|
|
Model 4196 Lead Placement Time
기간: Implant
|
Model 4196 lead placement time was defined as the time from insertion of the successfully placed lead to the time when it was placed in the first acceptable pacing location.
|
Implant
|
|
Total Implant Time
기간: Implant
|
Total implant time was defined as time from initial incision to final closure.
|
Implant
|
|
Assessment of Lead Handling Characteristics
기간: Implant
|
Lead handling characteristics assessed as "acceptable" by physicians
|
Implant
|
|
Electrical Performance - Tip Electrode: Sensing
기간: 12-month
|
Model 4196 lead tip electrode R-wave amplitude
|
12-month
|
|
Electrical Performance - Tip Electrode: LV Voltage Threshold
기간: 12-month
|
Model 4196 lead tip electrode LV voltage threshold
|
12-month
|
|
Electrical Performance -Tip Electrode: Pacing Impedance
기간: 12-month
|
Model 4196 lead tip electrode pacing impedance
|
12-month
|
|
Electrical Performance -Ring Electrode: Sensing
기간: Implant
|
Model 4196 lead ring electrode R-wave amplitude
|
Implant
|
|
Electrical Performance - Ring Electrode: LV Voltage Threshold
기간: 12-month
|
Model 4196 lead ring electrode LV voltage threshold
|
12-month
|
|
Electrical Performance -Ring Electrode: Pacing Impedance
기간: 12-Month
|
Model 4196 lead ring electrode pacing impedance
|
12-Month
|
|
Summarize All Adverse Events
기간: Up to 18 months
|
All adverse events were collected for this trial such as, but not limited to, the following: Atrial Fibrillation, Chest Pain, Pneumonia, Cold/Flu
|
Up to 18 months
|
공동 작업자 및 조사자
여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.
수사관
- 수석 연구원: Brian Ramza, MD, PHD, St. Luke's Mid America Heart Institute, Kansas City MO
연구 기록 날짜
이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.
연구 주요 날짜
연구 시작
2007년 2월 1일
기본 완료 (실제)
2007년 8월 1일
연구 완료 (실제)
2008년 12월 1일
연구 등록 날짜
최초 제출
2005년 12월 28일
QC 기준을 충족하는 최초 제출
2005년 12월 30일
처음 게시됨 (추정)
2006년 1월 2일
연구 기록 업데이트
마지막 업데이트 게시됨 (추정)
2009년 9월 28일
QC 기준을 충족하는 마지막 업데이트 제출
2009년 9월 24일
마지막으로 확인됨
2009년 9월 1일
추가 정보
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .