- ICH GCP
- 미국 임상 시험 레지스트리
- 임상시험 NCT05052242
Evaluation of Ocular Biometry in PACD With Two Swept Source Optical Coherence Tomography Devices
Evaluation of Ocular Biometry in Primary Angle-closure Disease With Two Swept Source Optical Coherence Tomography Devices
Purpose: To investigate agreement between 2 swept source OCT biometers, IOL M.aster700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD).
Methods: This was a prospective study conducted in a tertiary eye care centre involving biometric measurements obtained with 2 devices in phakic eye with diagnosis of PACD. Mean difference and intraclass correlation coefficient (ICC) with confidence limits were assessed, and calculations of estimated residual refraction of the IOL were analysed using Barrett's formula.
연구 개요
상세 설명
Primary angle-closure disease (PACD) is the leading cause of blindness worldwide. PACD is an anatomic disorder, typically of relatively small eye with shallow anterior chamber depth (ACD), thick lens, shorter axial length (AL), small corneal curvature, and relatively anterior lens position. Pupillary block, plateau iris and phacomorphic mechanisms play major roles in the pathophysiology of the disease, and pupillary block is thought to be the most common causal mechanism.
For decades, phacoemulsification and intraocular lens implantation (PEI) has been proposed for initial management of PACD.4 PEI relieves pupillary block, deepening ACD and widening irido-corneal angle.
Biometry is essential in evaluating ocular dimensions and calculating IOL power in PACD, and ocular biometers have been evolving in recent years. Optical coherence tomography (OCT) includes partial coherence interferometry (PCI), and swept source OCT (SS-OCT) is in general use in clinical practice. The agreement of the calculations of these devices has been studied by many investigators; however, research into their agreement in PACD patients has been sparse so far. A newer SS-OCT, Anterion, is now available and its performance needs to be evaluated.
연구 유형
등록 (실제)
연락처 및 위치
연구 장소
-
-
-
Bangkok, 태국, 10400
- Somporn Chantra
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
샘플링 방법
연구 인구
설명
Inclusion Criteria:
- Inclusion criteria were age > 40 years old
- phakic eye
- diagnosis of PACD, including primary angle-closure suspect (PACS), primary angle-closure (PAC) and primary angle-closure glaucoma (PACG) as classified by Foster et al
Exclusion Criteria:
- patients who had opaque optical media
- dense cataract
- anterior and/or posterior segment diseases such as advanced pterygium, diabetic retinopathy, or maculopathy
- Participants who had previous history of ocular trauma, ocular surgery
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
코호트 및 개입
그룹/코호트 |
개입 / 치료 |
|---|---|
|
one group compared two measurement methods
|
2 swept source OCT biometers, IOL Master700 and Anterion
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Intraclass correlation coefficient of IOL power calculation
기간: 30 minutes per person
|
agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in intraocular lens (IOL) calculations
|
30 minutes per person
|
|
Intraclass correlation coefficient of various ocular biometry
기간: 30 minutes per person
|
agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry( ACD, AL, LT, K1, K2, WTW)
|
30 minutes per person
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Somporn Chantra, MD, Rajavithi Hospital
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
기타 연구 ID 번호
- Rajavithi Hospital (기타 보조금/기금 번호: Rajavithi Hospital)
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .