Tourniquet pressure settings based on limb occlusion pressure determination or arterial occlusion pressure estimation in total knee arthroplasty? A prospective, randomized, double blind trial

Bahattin Tuncali, Hakan Boya, Zeynep Kayhan, Sukru Arac, Bahattin Tuncali, Hakan Boya, Zeynep Kayhan, Sukru Arac

Abstract

Objective: The aim of this study was to compare the limb occlusion pressure (LOP) determination and arterial occlusion pressure (AOP) estimation methods for tourniquet pressure setting in adult patients undergoing knee arthroplasty under combined spinal-epidural anesthesia.

Methods: Ninety-three patients were randomized into two groups. Pneumatic tourniquet inflation pressures were adjusted based either on LOP determination or AOP estimation in Group 1 (46 patients, 38 female and 8 male; mean age: 67.71 ± 9.17) and Group 2 (47 patients, 40 female and 7 male; mean age: 70.31 ± 8.27), respectively. Initial and maximal systolic blood pressures, LOP/AOP levels, required time to estimate AOP/determinate LOP and set the cuff pressure, initial and maximal tourniquet pressures and tourniquet time were recorded. The effectiveness of the tourniquet was assessed by the orthopedic surgeons using a Likert scale.

Results: Initial and maximal systolic blood pressures, determined LOP, estimated AOP, duration of tourniquet and the performance of the tourniquet were not different between groups. However, the initial (182.44 ± 14.59 mm Hg vs. 200.69 ± 15.55 mm Hg) and maximal tourniquet pressures (186.91 ± 12.91 mm Hg vs. 200.69 ± 15.55 mm Hg) were significantly lower, the time required to estimate AOP and set the tourniquet cuff pressure was significantly less (23.91 ± 4.77 s vs. 178.81 ± 25.46 s) in Group II (p = 0.000). No complications that could be related to the tourniquet were observed during or after surgery.

Conclusion: Tourniquet inflation pressure setting based on AOP estimation method provides a bloodless surgical field that is comparable to that of LOP determination method with lower pneumatic inflation pressure and less required time for cuff pressure adjustment in adult patients undergoing total knee arthroplasty under combined spinal epidural anesthesia.

Keywords: Arterial occlusion pressure; Estimation method; Knee arthroplasty; Obesity; Pneumatic tourniquet.

Copyright © 2018 Turkish Association of Orthopaedics and Traumatology. Production and hosting by Elsevier B.V. All rights reserved.

References

    1. Noordin S., McEwen J.A., Kragh J.F., Jr., Eisen A., Masri B.A. Surgical tourniquets in orthopaedics. J Bone Jointt Surg Am. 2009;91:2958–2967.
    1. Odinsson A., Finsen V. Tourniquet use and its complications in Norway. J Bone Joint Surg Br. 2006;88:1090–1092.
    1. Olivecrona C., Ponzer S., Hamberg P., Blomfeldt R. Lower tourniquet cuff pressure reduces postoperative wound complications after total knee arthroplasty: a randomized controlled study of 164 patients. J Bone Joint Surg Am. 2012;94:2216–2221.
    1. Olivecrona C., Blomfeldt R., Ponzer S., Stanford B.R., Nilsson B.Y. Tourniquet cuff pressure and nerve injury in knee arthroplasty in a bloodless field: a neurophysiological study. Acta Orthop. 2013;84:159–164.
    1. Brin Y.S., Feldman V., Ron Gal I., Markushevitch M., Regev A., Stern A. The sterile elastic exsanguination tourniquet vs. the pneumatic tourniquet for total knee arthroplasty. J Arthroplast. 2015;30:595–599.
    1. Drolet B.C., Okhah Z., Phillips B.Z. Evidence for safe tourniquet use in 500 consecutive upper extremity procedures. Hand (N Y) 2014;9:494–498.
    1. Tejwani N.C., Immerman I., Achan P., Egol K.A., McLaurin T. Tourniquet cuff pressure: the gulf between science and practice. J Trauma. 2006;61:1415–1418.
    1. Pauers R.S., Carocci M.A. Low pressure pneumatic tourniquets: effectiveness at minimum recommended inflation pressures. J Foot Ankle Surg. 1994;33:605–609.
    1. Shaw J.A., Murray D.G. The relationship between tourniquet pressure and underlying soft-tissue pressure in the thigh. J Bone Joint Surg Am. 1982;64:1148–1152.
    1. McLaren A.C., Rorabeck C.H. The pressure distribution under tourniquets. J Bone Joint Surg Am. 1985;67:433–438.
    1. Hargens A.R., McClure A.G., Skyhar M.J., Lieber R.L., Gershuni D.H., Akeson W.H. Local compression patterns beneath pneumatic tourniquets applied to arms and thighs of human cadavera. J Orthop Res. 1987;5:247–252.
    1. Graham B., Breault M.J., McEwen J.A., McGraw R.W. Perineural pressures under the pneumatic tourniquet in the upper extremity. J Hand Surg Br. 1992;17:262–266.
    1. Estebe J.P., Davies J.M., Richebe P. The pneumatic tourniquet: mechanical, ischaemia-reperfusion and systemic effects. Eur J Anaesthesiol. 2011;28:404–411.
    1. Estersohn H.S., Sourifman H.A. The minimum effective midthigh tourniquet pressure. J Foot Surg. 1982;21:281–284.
    1. Newman R.J., Muirhead A. A safe and effective low pressure tourniquet. A prospective evaluation. J Bone Jt Surg Br. 1986;68:625–628.
    1. Sadri A., Braithwaite I.J., Abdul-Jabar H.B., Sarraf K.M. Understanding of intra-operative tourniquets amongst orthopedic surgeons and theatre staff- a questionnaire study. Ann R Coll Surg Engl. 2010;92:243–245.
    1. Yalçınkaya M., Sökücü S., Erdoğan S., Kabukçuoğlu Y.S. Tourniquet use in orthopedic surgery: a descriptive survey study among Turkish orthopedic surgeons and residents in Istanbul. Acta Orthop Traumatol Turc. 2014;48:483–490.
    1. Boya H., Tuncalı B., Özcan O., Araç Ş., Tuncay C. Practice of tourniquets use in Turkey: a pilot study. Acta Orthop Traumatol Turc. 2016;50:162–170.
    1. Sharma J.P., Salhotra R. Tourniquets in orthopedic surgery. Indian J Orthop. 2012;46:377–383.
    1. Reid H.S., Camp R.A., Jacob W.H. Tourniquet hemostasis. A clinical study. Clin Orthop Relat Res. 1983;177:230–234.
    1. McEwen J.A., Inkpen K.B., Younger A. Thigh tourniquet safety: limb occlusion pressure measurement and a wide contoured cuff allow lower cuff pressures. Surg Technol. 2002;34:8–18.
    1. AORN Recommended Practices Committee Recommended practices for the use of the pneumatic tourniquet in the perioperative practice setting. AORN J. 2007;86:640–655.
    1. Younger A.S., Manzary M., Wing K.J., Stothers K. Automated cuff occlusion pressure effect on quality of operative fields in foot and ankle surgery: a randomized prospective study. Foot Ankle Int. 2011;32:239–243.
    1. Masri B.A., Day B., Younger A.S.E., Jeyasurya J. Technique for measuring limb occlusion pressure that facilitates personalized tourniquet systems: a randomized trial. J Med Biol Eng. 2016;36:644–650.
    1. Tuncali B., Karci A., Tuncali B.E. A new method for estimating arterial occlusion pressure in optimizing pneumatic tourniquet inflation pressure. Anesth Analg. 2006;102:1752–1757.
    1. Unver B., Karatosun V., Tuncali B. Effects of tourniquet pressure on rehabilitation outcomes in patients undergoing total knee arthroplasty. Orthop Nurs. 2013;32:217–222.
    1. Tuncali B., Boya H., Kayhan Z., Arac S., Camurdan M.A. Clinical utilization of arterial occlusion pressure estimation method in lower limb surgery: effectiveness of tourniquet pressures. Acta Orthop Traumatol Turc. 2016;50:171–177.
    1. Tuncali B., Karci A., Bacakoglu A.K., Tuncali B.E., Ekin A. Controlled hypotension and minimal inflation pressure: a new approach for pneumatic tourniquet application in upper limb surgery. Anesth Analg. 2003;97:1529–1532.
    1. Crenshaw A.G., Hargens A.R., Gershuni D.H., Rydevik B. Wide tourniquet cuffs more effective at lower inflation pressures. Acta Orthop Scand. 1988;59:447–451.
    1. Mittal P., Shenoy S., Sandhu J.S. Effect of different cuff widths on the motor nerve conduction of the median nerve: an experimental study. J Orthop Surg Res. 2008;3:1.
    1. Kovar F.M., Jaindl M., Oberleitner G. Nerve compression and pain in human volunteers with narrow vs wide tourniquets. World J Orthop. 2015;6:394–399.

Source: PubMed

3
구독하다