Tympanic, infrared skin, and temporal artery scan thermometers compared with rectal measurement in children: a real-life assessment

Karel Allegaert, Kristina Casteels, Ilse van Gorp, Guy Bogaert, Karel Allegaert, Kristina Casteels, Ilse van Gorp, Guy Bogaert

Abstract

Introduction: Body temperature measurement in children is of clinical relevance. Although rectal measurement is the gold standard, less invasive tools have become available. We aimed to describe the accuracy of tympanic, infrared skin, or temporal artery scan thermometers compared with rectal measurement to reflect core temperature.

Methods: Rectal (Filac 3000; Covidien, Mechelen, Belgium), tympanic (AccuSystem Genius2 Typmanic Infrared Ear Thermometer, Covidien, Mechelen, Belgium), temporal artery scan (Exergen, Exergen Corp, Watertown, Massachusetts), and infrared (ThermoFlash Contactless Medical Electronic Thermometer, Visiomedlab, Paris, France) body temperature measurements were randomly performed and readings were collected once. Temperature readings were described as median and range, and observations were compared with rectal temperature readings (using Wilcoxon, Bland-Altman, sensitivity, and specificity tests). The child's comfort was assessed by the child, parent, and nurse (using Likert scales) and ease of use was assessed by nurses (using visual analog scale).

Results: Based on observations in 294 (median age = 3.2 years, range = 0.02-17 years) children, the mean difference was 0.49°C (tympanic scan; P < 0.0001), 0.34°C (infrared skin scan; P < 0.0001), and 0°C (temporal artery scan; P = 0.9288), respectively, when compared with rectal temperature readings. Based on visual inspection of Bland-Altman plots, all tools overestimated the temperature at lower body temperature and underestimated the temperature at higher body temperature, resulting in a sensitivity of 22% to 41% and a specificity of 98% to 100% for rectal temperatures above 38°C. The Likert scale scores and the visual analog scale scores for rectal measurement were only slightly higher when compared with the other methods.

Conclusions: All noninvasive techniques underperformed compared with rectal measurement. The temporal artery scan deviations were smallest, but all noninvasive techniques overestimate lower temperatures and underestimate higher temperatures compared with rectal measurement. In our hands, temporal artery scan measurement seems to be second best, but not yet ideal.

Keywords: Bland-Altman; body temperature; infrared skin scan; rectal; temporal artery scan; tympanic scan.

Figures

Figure 1
Figure 1
Bland-Altman plot illustrating a mean difference of 0.49°C, with a temperature-related influence on the direction and extent of the difference between rectal or tympanic scan temperature measurement.
Figure 2
Figure 2
Bland-Altman plot illustrating a mean difference of 0.34°C, with a temperature-related influence on the direction and extent of the difference between rectal or infrared skin scan temperature measurement.
Figure 3
Figure 3
Bland-Altman plot illustrating a mean difference of 0°C, with a temperature-related influence on the direction and extent of the difference between rectal or temporal artery scan temperature measurement.

References

    1. Davis T. NICE guideline: feverish illness in children—assessment and initial management in children younger than 5 years. Arch Dis Child Educ Pract Ed. 2013;98:232–235.
    1. Walson P.D., Jones J., Chesney R. Antipyretic efficacy and tolerability of a single intravenous dose of the acetaminophen prodrug propacetamol in children: a randomized, double-blind, placebo-controlled trial. Clin Ther. 2006;28:762–769.
    1. Allegaert K., van den Anker J. Pharmacokinetics and pharmacodynamics of intravenous acetaminophen in neonates. Expert Rev Clin Pharmacol. 2011;4:713–718.
    1. Craig J.V., Lancaster G.A., Taylor S. Infrared ear thermometry compared with rectal thermometry in children: a systematic review. Lancet. 2002;360:603–609.
    1. Craig J.V., Lancaster G.A., Williamson P.R. Temperature measured at the axilla compared with rectum in children and young people: a systematic review. BMJ. 2000;320:1174–1178.
    1. Paes B.F., Vermuelen K., Brohet R.M. Accuracy of tympanic and infrared skin thermometers in children. Arch Dis Child. 2010;95:974–978.
    1. Batra P., Goval S. Comparison of rectal, axillary, tympanic and temporal artery thermometer in the pediatric emergency room. Pediatr Emerg Care. 2013;29:63–66.
    1. Dodd S.R., Lancaster G.A., Craig J.V. In a systematic review, infrared ear thermometry for fever diagnosis in children finds poor sensitivity. J Clin Epidemiol. 2006;59:354–357.
    1. Eyelade O.R., Orimadegun A.E., Akinyemi O.A. Esophageal, tympanic, rectal and skin temperatures in children undergoing surgery with general anesthesia. J Perianesth Nurs. 2011;26:151–159.
    1. Devrim I., Kara A., Ceyhan M. Measurement accuracy of fever by tympanic and axillary thermometry. Pediatr Emerg Care. 2007;23:16–19.
    1. Dew P.L. Is tympanic membrane thermometry the best method for recording temperature in children? J Child Health Care. 2006;10:96–110.
    1. Myny D., de Waele J., Defloor T. Temporal scanner thermometry: a new method of core temperature estimation in ICU patients. Scott Med J. 2005;50:15–18.
    1. Selent M.U., Molinari N.M., Baxter A. Mass screening for fever in children: a comparison of 3 infrared thermal detection systems. Pediatr Emerg Care. 2013;29:305–313.
    1. Fortuna E.L., Carney M.M., Macy M. Accuracy of non-contact infrared thermometry versus rectal thermometry in young children evaluated in the emergency department for fever. J Emerg Nurs. 2010;36:101–104.
    1. Chiappini E., Sollai S., Longhi R. Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings. J Clin Nurs. 2011;20:1311–1318.
    1. Reynolds M., Bonham L., Gueck M. Are temporal artery temperatures accurate enough to replace rectal temperature measurement in pediatric ED patients? J Emerg Nurs. 2012 Nov 8 [Epub ahead of print]
    1. Bahorski J., Repasky T., Ranner D. Temperature measurement in pediatrics: a comparison of the rectal method versus the temporal artery method. J Pediatr Nurs. 2012;27:243–247.
    1. Hopchet L., Kulo A., Rayyan M. Does intravenous paracetamol administration affect body temperature in neonates? Arch Dis Child. 2011;96:301–304.
    1. Van den Anker J.N. Optimising the management of fever and pain in children. Int J Clin Pract Suppl. 2013:26–32.

Source: PubMed

3
구독하다