Comparison of fast acquisition strategies in whole-heart four-dimensional flow cardiac MR: Two-center, 1.5 Tesla, phantom and in vivo validation study
Pankaj Garg, Jos J M Westenberg, Pieter J van den Boogaard, Peter P Swoboda, Rahoz Aziz, James R J Foley, Graham J Fent, F G J Tyl, L Coratella, Mohammed S M ElBaz, R J van der Geest, David M Higgins, John P Greenwood, Sven Plein, Pankaj Garg, Jos J M Westenberg, Pieter J van den Boogaard, Peter P Swoboda, Rahoz Aziz, James R J Foley, Graham J Fent, F G J Tyl, L Coratella, Mohammed S M ElBaz, R J van der Geest, David M Higgins, John P Greenwood, Sven Plein
Abstract
Purpose: To validate three widely-used acceleration methods in four-dimensional (4D) flow cardiac MR; segmented 4D-spoiled-gradient-echo (4D-SPGR), 4D-echo-planar-imaging (4D-EPI), and 4D-k-t Broad-use Linear Acquisition Speed-up Technique (4D-k-t BLAST).
Materials and methods: Acceleration methods were investigated in static/pulsatile phantoms and 25 volunteers on 1.5 Tesla MR systems. In phantoms, flow was quantified by 2D phase-contrast (PC), the three 4D flow methods and the time-beaker flow measurements. The later was used as the reference method. Peak velocity and flow assessment was done by means of all sequences. For peak velocity assessment 2D PC was used as the reference method. For flow assessment, consistency between mitral inflow and aortic outflow was investigated for all pulse-sequences. Visual grading of image quality/artifacts was performed on a four-point-scale (0 = no artifacts; 3 = nonevaluable).
Results: For the pulsatile phantom experiments, the mean error for 2D PC = 1.0 ± 1.1%, 4D-SPGR = 4.9 ± 1.3%, 4D-EPI = 7.6 ± 1.3% and 4D-k-t BLAST = 4.4 ± 1.9%. In vivo, acquisition time was shortest for 4D-EPI (4D-EPI = 8 ± 2 min versus 4D-SPGR = 9 ± 3 min, P < 0.05 and 4D-k-t BLAST = 9 ± 3 min, P = 0.29). 4D-EPI and 4D-k-t BLAST had minimal artifacts, while for 4D-SPGR, 40% of aortic valve/mitral valve (AV/MV) assessments scored 3 (nonevaluable). Peak velocity assessment using 4D-EPI demonstrated best correlation to 2D PC (AV:r = 0.78, P < 0.001; MV:r = 0.71, P < 0.001). Coefficient of variability (CV) for net forward flow (NFF) volume was least for 4D-EPI (7%) (2D PC:11%, 4D-SPGR: 29%, 4D-k-t BLAST: 30%, respectively).
Conclusion: In phantom, all 4D flow techniques demonstrated mean error of less than 8%. 4D-EPI demonstrated the least susceptibility to artifacts, good image quality, modest agreement with the current reference standard for peak intra-cardiac velocities and the highest consistency of intra-cardiac flow quantifications.
Level of evidence: 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2018;47:272-281.
Keywords: 4D flow cardiac MR; MR flow imaging; flow quantification; phase-contrast magnetic resonance imaging; validation.
© 2017 International Society for Magnetic Resonance in Medicine.
Figures
References
- Dyverfeldt P, Bissell M, Barker AJ, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson 2015;17:72.
- Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P. Accelerating cine phase‐contrast flow measurements using k‐t BLAST and k‐t SENSE. Magn Reson Med 2005;54:1430–1438.
- Stadlbauer A, van der Riet W, Crelier G, Salomonowitz E. Accelerated time‐resolved three‐dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k‐t BLAST. Eur J Radiol 2010;75:e15–e21.
- Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 2007;58:1182–1195.
- Davis CP, McKinnon GC, Debatin JF, et al. Normal heart: evaluation with echo‐planar MR imaging. Radiology 1994;191:691–696.
- Debatin JF, Davis CP, Felblinger J, McKinnon GC. Evaluation of ultrafast phase‐contrast imaging in the thoracic aorta. MAGMA 1995;3:59–66.
- Zaman A, Motwani M, Oliver JJ, et al. 3.0T, time‐resolved, 3D flow‐sensitive MR in the thoracic aorta: impact of k‐t BLAST acceleration using 8‐ versus 32‐channel coil arrays. J Magn Reson Imaging 2015;42:495–504.
- Lustig M, Pauly JM. SPIRiT: iterative self‐consistent parallel imaging reconstruction from arbitrary k‐space. Magn Reson Med 2010;64:457–471.
- Johnson KM, Markl M. Improved SNR in phase contrast velocimetry with five‐point balanced flow encoding. Magn Reson Med 2010;63:349–355.
- Calkoen EE, Roest AAW, Kroft LJM, et al. Characterization and improved quantification of left ventricular inflow using streamline visualization with 4DFlow MRI in healthy controls and patients after atrioventricular septal defect correction. J Magn Reson Imaging 2015;41:1512–1520.
- Calkoen EE, de Koning PJH, Blom NA, et al. Disturbed intracardiac flow organization after atrioventricular septal defect correction as assessed with 4D flow magnetic resonance imaging and quantitative particle tracing. Invest Radiol 2015;50:850–857.
- Schulz‐Menger J, Bluemke DA, Bremerich J, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing. J Cardiovasc Magn Reson 2013;15:35.
- Roes SD, Hammer S, van der Geest RJ, et al. Flow assessment through four heart valves simultaneously using 3‐dimensional 3‐directional velocity‐encoded magnetic resonance imaging with retrospective valve tracking in healthy volunteers and patients with valvular regurgitation. Invest Radiol 2009;44:669–675.
- Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase‐contrast MR imaging: basic facts and implementation. Radiographics 2002;22:651–671.
- Hsiao A, Tariq U, Alley MT, Lustig M, Vasanawala SS. Inlet and outlet valve flow and regurgitant volume may be directly and reliably quantified with accelerated, volumetric phase‐contrast MRI. J Magn Reson Imaging 2015;41:376–385.
- Frydrychowicz A, Wieben O, Niespodzany E, Reeder SB, Johnson KM, François CJ. Quantification of thoracic blood flow using volumetric magnetic resonance imaging with radial velocity encoding: in vivo validation. Invest Radiol 2013;48:819–825.
- Tariq U, Hsiao A, Alley M, Zhang T, Lustig M, Vasanawala SS. Venous and arterial flow quantification are equally accurate and precise with parallel imaging compressed sensing 4D phase contrast MRI. J Magn Reson Imaging 2013;37:1419–1426.
- Westenberg JJM, Roes SD, Ajmone Marsan N, et al. Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity‐encoded MR imaging with retrospective valve tracking. Radiology 2008;249:792–800.
- Wentland AL, Grist TM, Wieben O. Repeatability and internal consistency of abdominal 2D and 4D phase contrast MR flow measurements. Acad Radiol 2013;20:699–704.
- Hanneman K, Sivagnanam M, Nguyen ET, et al. Magnetic resonance assessment of pulmonary (QP) to systemic (QS) flows using 4D phase‐contrast imaging: pilot study comparison with standard through‐plane 2D phase‐contrast imaging. Acad Radiol 2014;21:1002–1008.
- Nett EJ, Johnson KM, Frydrychowicz A, et al. Four‐dimensional phase contrast MRI with accelerated dual velocity encoding. J Magn Reson Imaging 2012;35:1462–1471.
- Roldán‐Alzate A, Frydrychowicz A, Niespodzany E, et al. In vivo validation of 4D flow MRI for assessing the hemodynamics of portal hypertension. J Magn Reson Imaging 2013;37:1100–1108.
- Brix L, Ringgaard S, Rasmusson A, Sørensen TS, Kim WY. Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions. J Cardiovasc Magn Reson 2009;11:3.
- Carlsson M, Töger J, Kanski M, et al. Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k‐t BLAST: head to head comparison and validation at 1.5 T and 3 T. J Cardiovasc Magn Reson 2011;13:55.
- Kanski M, Töger J, Steding‐Ehrenborg K, et al. Whole‐heart four‐dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use: a head‐to‐head comparison. BMC Med Imaging 2015;15:20.
- Petersson S, Sigfridsson A, Dyverfeldt P, Carlhäll C‐J, Ebbers T. Retrospectively gated intracardiac 4D flow MRI using spiral trajectories. Magn Reson Med 2016;75:196–206.
Source: PubMed