Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.
24 augustus 2020 bijgewerkt door: Bausch & Lomb Incorporated
A Prospective Trial to Evaluate the Safety and Effectiveness of the Bausch & Lomb Technolas® 217z Laser With Zyoptix® Aspheric Algorithm When Used for the LASIK Treatment of Myopia and Myopic Astigmatism
The purpose of this study is to evaluate the safety and effectiveness of the Bausch & Lomb Technolas 217z laser with the investigational Zyoptix Aspheric Algorithm when used for LASIK treatment of myopia and myopic astigmatism.
Studie Overzicht
Toestand
Voltooid
Conditie
Interventie / Behandeling
Studietype
Ingrijpend
Inschrijving (Werkelijk)
72
Fase
- Niet toepasbaar
Contacten en locaties
In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.
Studie Locaties
-
-
Ontario
-
Toronto, Ontario, Canada, M5X1C8
- Lasik MD, 130 King Street
-
-
Deelname Criteria
Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
18 jaar en ouder (Volwassen, Oudere volwassene)
Accepteert gezonde vrijwilligers
Nee
Geslachten die in aanmerking komen voor studie
Allemaal
Beschrijving
Inclusion Criteria:
- Subjects who were of legal age (at least 18) and had full legal capacity to volunteer.
- Subjects must be willing to have both eyes treated with the laser.
- Subjects must have up to -12.00D of absolute spherical myopia (not spherical equivalent) with up to -6.00D of refractive astigmatism, with a spherical equivalent up to -15.00D as measured by manifest subjective refraction.
- Subjects who had less than 0.75 D of difference in the preoperative manifest spherical equivalent and cycloplegic spherical equivalent subjective refractions.
(Manifest subjective refractions adjusted for optical infinity were used.)
- High contrast, manifest best spectacle-corrected distance logMAR visual acuity correctable to at least 0.3 in both eyes, and must not differ between eyes by more than 1 line (5 letters).
- Contact lens wearers must have 2 Orbscan measurements and 2 manifest subjective refractions taken preoperatively at least 1 week apart.
- Contact lens wearers must have gas permeable lenses discontinued for at least 3 weeks and soft lenses discontinued for at least 2 weeks prior to the preoperative evaluation.
Exclusion
- Subjects with uncorrected visual acuity better than 20/40 (Snellen) or 0.3 logMAR
- Subjects who were immuno-compromised or had a diagnosis of connective tissue disease, clinically significant atopic disease, diabetes, autoimmune diseases and other acute or chronic illnesses that increased the risk to the subject or could have confounded the outcomes of this study.
- Subjects with uncontrolled glaucoma or glaucoma who were under treatment in the study eye or with ocular hypertension.
- Subjects who demonstrated evidence of retinal vascular disease.
- Subjects who have an ocular muscle disorder including strabismus or nystagmus, or other disorders that affected fixation.
- Subjects who were at risk for angle closure or who had a potentially occludable angle, as evidenced by a shallow anterior chamber on slit lamp exam, confirmed by gonioscopy as deemed necessary by the study Investigator.
- Subjects who had undergone previous intraocular or corneal surgery of any kind, including any type of excimer laser surgery for either refractive or therapeutic purposes.
- Subjects who were taking systemic medications likely to affect wound healing such as corticosteroids or anti-metabolites or any other medications that may increase the risk to the subject or confound the outcomes of this study.
- Subjects for whom the combination of their preoperative corneal thickness and the planned operative parameters for the LASIK procedure could have resulted in less than 250 microns of remaining posterior corneal thickness below the flap postoperatively.
Studie plan
Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Behandeling
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: Technolas 217z Excimer Laser
Bausch & Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism.
|
LASIK correction of myopia and myopic astigmatism
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
Tijdsspanne: 3 months
|
3 months
|
|
|
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Tijdsspanne: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
Tijdsspanne: Baseline, 3 months
|
Mean change in best-corrected high contrast distance visual acuity.
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
Baseline, 3 months
|
|
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
Tijdsspanne: 3 months
|
3 months
|
|
|
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
Tijdsspanne: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
Tijdsspanne: 3 months
|
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
3 months
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Percentage Change From Baseline in Higher Order Aberrations.
Tijdsspanne: Baseline, 3 months
|
The aberrations in the eye's optical system were measured over the entire pupil by the Bausch & Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device.
The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II.
The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms.
These terms represent the wavefront aberration of the eye.
|
Baseline, 3 months
|
Medewerkers en onderzoekers
Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.
Sponsor
Onderzoekers
- Hoofdonderzoeker: Mark Cohen, MD, Lasik MD
Studie record data
Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.
Bestudeer belangrijke data
Studie start
1 augustus 2008
Primaire voltooiing (Werkelijk)
1 maart 2010
Studie voltooiing (Werkelijk)
1 maart 2010
Studieregistratiedata
Eerst ingediend
8 oktober 2008
Eerst ingediend dat voldeed aan de QC-criteria
8 oktober 2008
Eerst geplaatst (Schatting)
9 oktober 2008
Updates van studierecords
Laatste update geplaatst (Werkelijk)
11 september 2020
Laatste update ingediend die voldeed aan QC-criteria
24 augustus 2020
Laatst geverifieerd
1 augustus 2020
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 564
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .