- ICH GCP
- Rejestr badań klinicznych w USA
- Badanie kliniczne NCT00769808
Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.
24 sierpnia 2020 zaktualizowane przez: Bausch & Lomb Incorporated
A Prospective Trial to Evaluate the Safety and Effectiveness of the Bausch & Lomb Technolas® 217z Laser With Zyoptix® Aspheric Algorithm When Used for the LASIK Treatment of Myopia and Myopic Astigmatism
The purpose of this study is to evaluate the safety and effectiveness of the Bausch & Lomb Technolas 217z laser with the investigational Zyoptix Aspheric Algorithm when used for LASIK treatment of myopia and myopic astigmatism.
Przegląd badań
Status
Zakończony
Warunki
Interwencja / Leczenie
Typ studiów
Interwencyjne
Zapisy (Rzeczywisty)
72
Faza
- Nie dotyczy
Kontakty i lokalizacje
Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.
Lokalizacje studiów
-
-
Ontario
-
Toronto, Ontario, Kanada, M5X1C8
- Lasik MD, 130 King Street
-
-
Kryteria uczestnictwa
Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.
Kryteria kwalifikacji
Wiek uprawniający do nauki
18 lat i starsze (Dorosły, Starszy dorosły)
Akceptuje zdrowych ochotników
Nie
Płeć kwalifikująca się do nauki
Wszystko
Opis
Inclusion Criteria:
- Subjects who were of legal age (at least 18) and had full legal capacity to volunteer.
- Subjects must be willing to have both eyes treated with the laser.
- Subjects must have up to -12.00D of absolute spherical myopia (not spherical equivalent) with up to -6.00D of refractive astigmatism, with a spherical equivalent up to -15.00D as measured by manifest subjective refraction.
- Subjects who had less than 0.75 D of difference in the preoperative manifest spherical equivalent and cycloplegic spherical equivalent subjective refractions.
(Manifest subjective refractions adjusted for optical infinity were used.)
- High contrast, manifest best spectacle-corrected distance logMAR visual acuity correctable to at least 0.3 in both eyes, and must not differ between eyes by more than 1 line (5 letters).
- Contact lens wearers must have 2 Orbscan measurements and 2 manifest subjective refractions taken preoperatively at least 1 week apart.
- Contact lens wearers must have gas permeable lenses discontinued for at least 3 weeks and soft lenses discontinued for at least 2 weeks prior to the preoperative evaluation.
Exclusion
- Subjects with uncorrected visual acuity better than 20/40 (Snellen) or 0.3 logMAR
- Subjects who were immuno-compromised or had a diagnosis of connective tissue disease, clinically significant atopic disease, diabetes, autoimmune diseases and other acute or chronic illnesses that increased the risk to the subject or could have confounded the outcomes of this study.
- Subjects with uncontrolled glaucoma or glaucoma who were under treatment in the study eye or with ocular hypertension.
- Subjects who demonstrated evidence of retinal vascular disease.
- Subjects who have an ocular muscle disorder including strabismus or nystagmus, or other disorders that affected fixation.
- Subjects who were at risk for angle closure or who had a potentially occludable angle, as evidenced by a shallow anterior chamber on slit lamp exam, confirmed by gonioscopy as deemed necessary by the study Investigator.
- Subjects who had undergone previous intraocular or corneal surgery of any kind, including any type of excimer laser surgery for either refractive or therapeutic purposes.
- Subjects who were taking systemic medications likely to affect wound healing such as corticosteroids or anti-metabolites or any other medications that may increase the risk to the subject or confound the outcomes of this study.
- Subjects for whom the combination of their preoperative corneal thickness and the planned operative parameters for the LASIK procedure could have resulted in less than 250 microns of remaining posterior corneal thickness below the flap postoperatively.
Plan studiów
Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.
Jak projektuje się badanie?
Szczegóły projektu
- Główny cel: Leczenie
- Przydział: Nie dotyczy
- Model interwencyjny: Zadanie dla jednej grupy
- Maskowanie: Brak (otwarta etykieta)
Broń i interwencje
Grupa uczestników / Arm |
Interwencja / Leczenie |
|---|---|
|
Eksperymentalny: Technolas 217z Excimer Laser
Bausch & Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism.
|
LASIK correction of myopia and myopic astigmatism
|
Co mierzy badanie?
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
Ramy czasowe: 3 months
|
3 months
|
|
|
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Ramy czasowe: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
Ramy czasowe: Baseline, 3 months
|
Mean change in best-corrected high contrast distance visual acuity.
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
Baseline, 3 months
|
|
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
Ramy czasowe: 3 months
|
3 months
|
|
|
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
Ramy czasowe: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
Ramy czasowe: 3 months
|
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
3 months
|
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Percentage Change From Baseline in Higher Order Aberrations.
Ramy czasowe: Baseline, 3 months
|
The aberrations in the eye's optical system were measured over the entire pupil by the Bausch & Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device.
The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II.
The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms.
These terms represent the wavefront aberration of the eye.
|
Baseline, 3 months
|
Współpracownicy i badacze
Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.
Sponsor
Śledczy
- Główny śledczy: Mark Cohen, MD, Lasik MD
Daty zapisu na studia
Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.
Główne daty studiów
Rozpoczęcie studiów
1 sierpnia 2008
Zakończenie podstawowe (Rzeczywisty)
1 marca 2010
Ukończenie studiów (Rzeczywisty)
1 marca 2010
Daty rejestracji na studia
Pierwszy przesłany
8 października 2008
Pierwszy przesłany, który spełnia kryteria kontroli jakości
8 października 2008
Pierwszy wysłany (Oszacować)
9 października 2008
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
11 września 2020
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
24 sierpnia 2020
Ostatnia weryfikacja
1 sierpnia 2020
Więcej informacji
Terminy związane z tym badaniem
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
- 564
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .