Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.

24 sierpnia 2020 zaktualizowane przez: Bausch & Lomb Incorporated

A Prospective Trial to Evaluate the Safety and Effectiveness of the Bausch & Lomb Technolas® 217z Laser With Zyoptix® Aspheric Algorithm When Used for the LASIK Treatment of Myopia and Myopic Astigmatism

The purpose of this study is to evaluate the safety and effectiveness of the Bausch & Lomb Technolas 217z laser with the investigational Zyoptix Aspheric Algorithm when used for LASIK treatment of myopia and myopic astigmatism.

Przegląd badań

Status

Zakończony

Typ studiów

Interwencyjne

Zapisy (Rzeczywisty)

72

Faza

  • Nie dotyczy

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Lokalizacje studiów

    • Ontario
      • Toronto, Ontario, Kanada, M5X1C8
        • Lasik MD, 130 King Street

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

18 lat i starsze (Dorosły, Starszy dorosły)

Akceptuje zdrowych ochotników

Nie

Płeć kwalifikująca się do nauki

Wszystko

Opis

Inclusion Criteria:

  • Subjects who were of legal age (at least 18) and had full legal capacity to volunteer.
  • Subjects must be willing to have both eyes treated with the laser.
  • Subjects must have up to -12.00D of absolute spherical myopia (not spherical equivalent) with up to -6.00D of refractive astigmatism, with a spherical equivalent up to -15.00D as measured by manifest subjective refraction.
  • Subjects who had less than 0.75 D of difference in the preoperative manifest spherical equivalent and cycloplegic spherical equivalent subjective refractions.

(Manifest subjective refractions adjusted for optical infinity were used.)

  • High contrast, manifest best spectacle-corrected distance logMAR visual acuity correctable to at least 0.3 in both eyes, and must not differ between eyes by more than 1 line (5 letters).
  • Contact lens wearers must have 2 Orbscan measurements and 2 manifest subjective refractions taken preoperatively at least 1 week apart.
  • Contact lens wearers must have gas permeable lenses discontinued for at least 3 weeks and soft lenses discontinued for at least 2 weeks prior to the preoperative evaluation.

Exclusion

  • Subjects with uncorrected visual acuity better than 20/40 (Snellen) or 0.3 logMAR
  • Subjects who were immuno-compromised or had a diagnosis of connective tissue disease, clinically significant atopic disease, diabetes, autoimmune diseases and other acute or chronic illnesses that increased the risk to the subject or could have confounded the outcomes of this study.
  • Subjects with uncontrolled glaucoma or glaucoma who were under treatment in the study eye or with ocular hypertension.
  • Subjects who demonstrated evidence of retinal vascular disease.
  • Subjects who have an ocular muscle disorder including strabismus or nystagmus, or other disorders that affected fixation.
  • Subjects who were at risk for angle closure or who had a potentially occludable angle, as evidenced by a shallow anterior chamber on slit lamp exam, confirmed by gonioscopy as deemed necessary by the study Investigator.
  • Subjects who had undergone previous intraocular or corneal surgery of any kind, including any type of excimer laser surgery for either refractive or therapeutic purposes.
  • Subjects who were taking systemic medications likely to affect wound healing such as corticosteroids or anti-metabolites or any other medications that may increase the risk to the subject or confound the outcomes of this study.
  • Subjects for whom the combination of their preoperative corneal thickness and the planned operative parameters for the LASIK procedure could have resulted in less than 250 microns of remaining posterior corneal thickness below the flap postoperatively.

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

  • Główny cel: Leczenie
  • Przydział: Nie dotyczy
  • Model interwencyjny: Zadanie dla jednej grupy
  • Maskowanie: Brak (otwarta etykieta)

Broń i interwencje

Grupa uczestników / Arm
Interwencja / Leczenie
Eksperymentalny: Technolas 217z Excimer Laser
Bausch & Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism.
LASIK correction of myopia and myopic astigmatism

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
Ramy czasowe: 3 months
3 months
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Ramy czasowe: 3 months
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).
3 months
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
Ramy czasowe: Baseline, 3 months
Mean change in best-corrected high contrast distance visual acuity. Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.
Baseline, 3 months
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
Ramy czasowe: 3 months
3 months
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
Ramy czasowe: 3 months
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).
3 months
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
Ramy czasowe: 3 months
Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.
3 months

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Percentage Change From Baseline in Higher Order Aberrations.
Ramy czasowe: Baseline, 3 months
The aberrations in the eye's optical system were measured over the entire pupil by the Bausch & Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device. The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II. The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms. These terms represent the wavefront aberration of the eye.
Baseline, 3 months

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Śledczy

  • Główny śledczy: Mark Cohen, MD, Lasik MD

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów

1 sierpnia 2008

Zakończenie podstawowe (Rzeczywisty)

1 marca 2010

Ukończenie studiów (Rzeczywisty)

1 marca 2010

Daty rejestracji na studia

Pierwszy przesłany

8 października 2008

Pierwszy przesłany, który spełnia kryteria kontroli jakości

8 października 2008

Pierwszy wysłany (Oszacować)

9 października 2008

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

11 września 2020

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

24 sierpnia 2020

Ostatnia weryfikacja

1 sierpnia 2020

Więcej informacji

Terminy związane z tym badaniem

Inne numery identyfikacyjne badania

  • 564

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .