Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.

24 августа 2020 г. обновлено: Bausch & Lomb Incorporated

A Prospective Trial to Evaluate the Safety and Effectiveness of the Bausch & Lomb Technolas® 217z Laser With Zyoptix® Aspheric Algorithm When Used for the LASIK Treatment of Myopia and Myopic Astigmatism

The purpose of this study is to evaluate the safety and effectiveness of the Bausch & Lomb Technolas 217z laser with the investigational Zyoptix Aspheric Algorithm when used for LASIK treatment of myopia and myopic astigmatism.

Обзор исследования

Статус

Завершенный

Вмешательство/лечение

Тип исследования

Интервенционный

Регистрация (Действительный)

72

Фаза

  • Непригодный

Контакты и местонахождение

В этом разделе приведены контактные данные лиц, проводящих исследование, и информация о том, где проводится это исследование.

Места учебы

    • Ontario
      • Toronto, Ontario, Канада, M5X1C8
        • Lasik MD, 130 King Street

Критерии участия

Исследователи ищут людей, которые соответствуют определенному описанию, называемому критериям приемлемости. Некоторыми примерами этих критериев являются общее состояние здоровья человека или предшествующее лечение.

Критерии приемлемости

Возраст, подходящий для обучения

18 лет и старше (Взрослый, Пожилой взрослый)

Принимает здоровых добровольцев

Нет

Полы, имеющие право на обучение

Все

Описание

Inclusion Criteria:

  • Subjects who were of legal age (at least 18) and had full legal capacity to volunteer.
  • Subjects must be willing to have both eyes treated with the laser.
  • Subjects must have up to -12.00D of absolute spherical myopia (not spherical equivalent) with up to -6.00D of refractive astigmatism, with a spherical equivalent up to -15.00D as measured by manifest subjective refraction.
  • Subjects who had less than 0.75 D of difference in the preoperative manifest spherical equivalent and cycloplegic spherical equivalent subjective refractions.

(Manifest subjective refractions adjusted for optical infinity were used.)

  • High contrast, manifest best spectacle-corrected distance logMAR visual acuity correctable to at least 0.3 in both eyes, and must not differ between eyes by more than 1 line (5 letters).
  • Contact lens wearers must have 2 Orbscan measurements and 2 manifest subjective refractions taken preoperatively at least 1 week apart.
  • Contact lens wearers must have gas permeable lenses discontinued for at least 3 weeks and soft lenses discontinued for at least 2 weeks prior to the preoperative evaluation.

Exclusion

  • Subjects with uncorrected visual acuity better than 20/40 (Snellen) or 0.3 logMAR
  • Subjects who were immuno-compromised or had a diagnosis of connective tissue disease, clinically significant atopic disease, diabetes, autoimmune diseases and other acute or chronic illnesses that increased the risk to the subject or could have confounded the outcomes of this study.
  • Subjects with uncontrolled glaucoma or glaucoma who were under treatment in the study eye or with ocular hypertension.
  • Subjects who demonstrated evidence of retinal vascular disease.
  • Subjects who have an ocular muscle disorder including strabismus or nystagmus, or other disorders that affected fixation.
  • Subjects who were at risk for angle closure or who had a potentially occludable angle, as evidenced by a shallow anterior chamber on slit lamp exam, confirmed by gonioscopy as deemed necessary by the study Investigator.
  • Subjects who had undergone previous intraocular or corneal surgery of any kind, including any type of excimer laser surgery for either refractive or therapeutic purposes.
  • Subjects who were taking systemic medications likely to affect wound healing such as corticosteroids or anti-metabolites or any other medications that may increase the risk to the subject or confound the outcomes of this study.
  • Subjects for whom the combination of their preoperative corneal thickness and the planned operative parameters for the LASIK procedure could have resulted in less than 250 microns of remaining posterior corneal thickness below the flap postoperatively.

Учебный план

В этом разделе представлена ​​подробная информация о плане исследования, в том числе о том, как планируется исследование и что оно измеряет.

Как устроено исследование?

Детали дизайна

  • Основная цель: Уход
  • Распределение: Н/Д
  • Интервенционная модель: Одногрупповое задание
  • Маскировка: Нет (открытая этикетка)

Оружие и интервенции

Группа участников / Армия
Вмешательство/лечение
Экспериментальный: Technolas 217z Excimer Laser
Bausch & Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism.
LASIK correction of myopia and myopic astigmatism

Что измеряет исследование?

Первичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
Временное ограничение: 3 months
3 months
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Временное ограничение: 3 months
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).
3 months
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
Временное ограничение: Baseline, 3 months
Mean change in best-corrected high contrast distance visual acuity. Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.
Baseline, 3 months
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
Временное ограничение: 3 months
3 months
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
Временное ограничение: 3 months
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).
3 months
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
Временное ограничение: 3 months
Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.
3 months

Вторичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Percentage Change From Baseline in Higher Order Aberrations.
Временное ограничение: Baseline, 3 months
The aberrations in the eye's optical system were measured over the entire pupil by the Bausch & Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device. The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II. The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms. These terms represent the wavefront aberration of the eye.
Baseline, 3 months

Соавторы и исследователи

Здесь вы найдете людей и организации, участвующие в этом исследовании.

Следователи

  • Главный следователь: Mark Cohen, MD, Lasik MD

Даты записи исследования

Эти даты отслеживают ход отправки отчетов об исследованиях и сводных результатов на сайт ClinicalTrials.gov. Записи исследований и сообщаемые результаты проверяются Национальной медицинской библиотекой (NLM), чтобы убедиться, что они соответствуют определенным стандартам контроля качества, прежде чем публиковать их на общедоступном веб-сайте.

Изучение основных дат

Начало исследования

1 августа 2008 г.

Первичное завершение (Действительный)

1 марта 2010 г.

Завершение исследования (Действительный)

1 марта 2010 г.

Даты регистрации исследования

Первый отправленный

8 октября 2008 г.

Впервые представлено, что соответствует критериям контроля качества

8 октября 2008 г.

Первый опубликованный (Оценивать)

9 октября 2008 г.

Обновления учебных записей

Последнее опубликованное обновление (Действительный)

11 сентября 2020 г.

Последнее отправленное обновление, отвечающее критериям контроля качества

24 августа 2020 г.

Последняя проверка

1 августа 2020 г.

Дополнительная информация

Термины, связанные с этим исследованием

Дополнительные соответствующие термины MeSH

Другие идентификационные номера исследования

  • 564

Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .

Клинические исследования Bausch & Lomb 217z Excimer Laser

Подписаться