Closed Loop Glucose Control in a Simulated ICU Setting

9 juni 2026 bijgewerkt door: Ideal Medical Technologies

Demonstration of an Expert Rule-Based Automated Insulin Delivery (AID) System for People With Type 1 and Type 2 Diabetes

This is a proof of concept safety study of a rule-based closed loop glucose control system designed for use in the intensive care unit setting.

The persons with type 1 and type 2 diabetes in this study will have their glucose controlled to a range of 100-140 mg/dL by a fully autonomous novel rule-based closed loop glucose control system for a period of 24 hours. The subjects will consume three standardized meals and receive a continuous intravenous infusion of dextrose (D-glucose) during the 24 hour study period. Meals will not be announced to the glucose control system.

Studie Overzicht

Gedetailleerde beschrijving

Tight glucose control in the intensive care unit (ICU) setting is difficult to achieve. The investigators hypothesize that a closed loop glucose control system that is rule-based will improve upon the glucose control currently achieved by the current open loop manual methods, and that this improved glucose control may improve the outcomes of critically ill patients.

This Earl Feasibility Study will test the ability of a prototype rule-based closed loop glucose control system named FUSION, to provide safe and effective glucose control in subjects with type 1 and type 2 diabetes in a clinical research center (CRC) setting. Subjects with type 1 diabetes have been chosen as safe and effective glucose control is difficult to achieve in these subjects during meal challenges. Subjects with type 2 diabetes have been chosen as they are insulin resistant, which makes their insulin resistance profile similar to that of ICU patients. As this is a safety study of a new medical device, the controlled environment of the CRC is preferable to the less controlled environment of an ICU setting.

The prototype FUSION system to be used in this study will consist to two Dexcom G6 continuous glucose monitors (CGM), the rule-based glucose control software run on an all-in-one medical computer, and two syringe pumps. The prototype system is housed on a medical cart. Based on the average glucose value of the two Dexcom G6 CGM's, and the algorithms of the FUSION systems rule-based glucose control software, the FUSION system will make rate adjustments every 5-10 minutes to the intravenous infusion rates of short acting insulin (NovoLog) and dextrose (D10NS) under its control, in an attempt to keep the subjects glucose in the range of 100-140 mg/dL. The FUSION system only requires entry of the subjects study identification number and weight in kilograms to initiate the system.

For safety reasons, the subjects will have their blood glucose independently measured every 10-60 minutes on the point of care Nova StatStrip system, throughout the 24 hour study period.

The study has halting criteria to avoid recurrent instances of severe hypoglycemia (< 54 mg/dL).

The average of the two CGM's, that is used by the FUSION system for glucose control, will be used for statistical analysis.

Studietype

Ingrijpend

Inschrijving (Geschat)

6

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Contact Back-up

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

  • Volwassen
  • Oudere volwassene

Accepteert gezonde vrijwilligers

Nee

Beschrijving

Inclusion Criteria:

  • Participants with type 1 and type 2 diabetes are eligible to be included in the study only if they meet all the following criteria:

    1. Are 18-70 years of age, inclusive.
    2. Can understand and sign an informed consent, communicate with the investigator, and understand and comply with the protocol requirements.
    3. Have had a diagnosis of type 1 or type 2 diabetes for a period of at least 3 months.
    4. Use insulin injections at home for glucose control.
    5. Have a hemoglobin A1c (HbA1c) in the range of 7.0 - 10.0%.
    6. Have a hemoglobin in the range for sex:

      1. Females: 7-15.5 grams/dL.
      2. Males: 7-17.5 grams/dl.
    7. Have adequate venous access sites in upper extremities.
    8. Body weight between 40 - 150 kg.

Exclusion Criteria:

  1. Have participated in an interventional medical, surgical, or pharmaceutical study within 30 days of screening.
  2. Have a known hypersensitivity to any of the components of study treatment.
  3. Have skin disease/injury at Dexcom G6 CGM insertion site(s) that would prevent insertion of the CGM.
  4. Currently abuses drugs or alcohol or has a history of abuse that in the investigator's opinion would cause the individual to be noncompliant.
  5. Have a medical condition that in the opinion of the investigator could affect study participation and/or personal well-being.
  6. Have a clinically significant history or presence of any of the following conditions:

    1. Hepatic failure or has alanine aminotransferase (ALT) greater than 3 times the upper limit of normal.
    2. Has an estimated GFR <30 ml/min/1.73 m2 or End Stage Kidney Disease on renal replacement therapy.
    3. Have congestive heart failure greater than class 1 on the NYHA classification system.
    4. Have a history of seizures.
    5. Have a history of cerebrovascular accident.
    6. Have a history of ischemic heart disease.
  7. For female subjects of potential childbearing age (age 18 to 55) they will be excluded if:

    1. Pregnant.
    2. Refuse to agree to a pregnancy test at the time of enrollment.
    3. Have a positive urine pregnancy test at the time of enrollment.
  8. Have a positive Influenza or COVID test within 14 days of visit 3.
  9. Have any viral lower respiratory related symptoms (e.g., cough, chest tightness, respiratory distress) in the 14-day period prior to visit 3.
  10. Have a known unprotected Influenza or COVID exposure in the 14-day period prior to visit 3.
  11. Are currently being treated with Hydroxyurea, a medication which may cause the Dexcom G6 CGM to produce glucose readings that are higher than the actual glucose value.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: NVT
  • Interventioneel model: Opdracht voor een enkele groep
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Fusion gesloten lus glucose controlesysteem veiligheidsstudie
Alle proefpersonen worden gedurende maximaal 24 uur behandeld met het fusie gesloten lus glucose -besturingssysteem
Het FUSION-systeem zal worden gebruikt om de glucose van de proefpersoon te reguleren tot een bereik van 100-140 mg/dL. Gegevens worden gedurende maximaal 24 uur verzameld, of bij vroegtijdige beëindiging van de studiesessie.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Primary Safety Measure
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
The percentage of all glucose values <70 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
Primary Efficacy Measure
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-180 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Percent of All Glucose Values Within Glucose Range <54 mg/dL
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are less than 54 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 54-69 mg/dL
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
The percentage of all glucose values that are within the glucose range of 54-69 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
Number of Hypoglycemic Events (<70 mg/dL)
Tijdsspanne: From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
The total number of hypoglycemic events less than 70 mg/dL as measured by a continuous series of glucose values less than 70 mg/dL for at least 15 minutes duration.
From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
Percent of All Glucose Values Within Glucose Range 70-100 mg/dL
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-100 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 70-140 mg/dL
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 100-140 mg/dL
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 100-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >140 mg/dL
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values greater than 140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >180 mg/dL
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 180 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >250 mg/dL
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 250 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Average glucose value in mg/dL
Tijdsspanne: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The average of all measured glucose values in mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Measure of dispersion - Coefficient of Variation
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The Coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value by the average glucose value, then multiplying this value by 100.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Clarke Error Grid Analysis of Continuous Glucose Monitor Data
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine the percentage of the two Dexcom G6 CGM's glucose values in each zone using a Clarke error grid analysis. This same calculation will be performed on the average of the two Dexcom G6 CGM systems used by the FUSION system for purposes of glucose control.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
FDA integrated Continuous Glucose Monitor Standards
Tijdsspanne: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine if each individual Dexcom G6 CGM system, and their average value meet FDA criteria for integrated Continuous Glucose Monitor (iCGM) accuracy for the glucose zones <70 mg/dL, 70-180 mg/dL, and >180 mg/dL.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Hemoglobin A1c Percentage Level
Tijdsspanne: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
C-Peptide Level
Tijdsspanne: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects C-peptide level will be measured prior to the start of the closed loop glucose control session.
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
Total Daily Dose of Insulin in Units/Kilogram
Tijdsspanne: As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms
As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
Weight in Kilograms
Tijdsspanne: As measured just prior to the start of the 24-hour closed loop glucose control session
Average weight in kilograms of the subjects
As measured just prior to the start of the 24-hour closed loop glucose control session
Body Mass Index
Tijdsspanne: As measured just prior to the start of the 24-hour closed loop glucose control session
The body mass index will be calculated by dividing the subjects weight in kilograms by the square of their height in meters
As measured just prior to the start of the 24-hour closed loop glucose control session

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Onderzoekers

  • Studie directeur: Leon M DeJournett, MD, Ideal Medical Technologies
  • Hoofdonderzoeker: Francisco Pasquel, MD, MPH, Emory School Of Medicine

Publicaties en nuttige links

De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Geschat)

1 juli 2026

Primaire voltooiing (Geschat)

1 december 2026

Studie voltooiing (Geschat)

1 december 2027

Studieregistratiedata

Eerst ingediend

9 juni 2026

Eerst ingediend dat voldeed aan de QC-criteria

9 juni 2026

Eerst geplaatst (Werkelijk)

12 juni 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

12 juni 2026

Laatste update ingediend die voldeed aan QC-criteria

9 juni 2026

Laatst geverifieerd

1 juni 2026

Meer informatie

Termen gerelateerd aan deze studie

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Ja

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .