Closed Loop Glucose Control in a Simulated ICU Setting

9 de junio de 2026 actualizado por: Ideal Medical Technologies

Demonstration of an Expert Rule-Based Automated Insulin Delivery (AID) System for People With Type 1 and Type 2 Diabetes

This is a proof of concept safety study of a rule-based closed loop glucose control system designed for use in the intensive care unit setting.

The persons with type 1 and type 2 diabetes in this study will have their glucose controlled to a range of 100-140 mg/dL by a fully autonomous novel rule-based closed loop glucose control system for a period of 24 hours. The subjects will consume three standardized meals and receive a continuous intravenous infusion of dextrose (D-glucose) during the 24 hour study period. Meals will not be announced to the glucose control system.

Descripción general del estudio

Descripción detallada

Tight glucose control in the intensive care unit (ICU) setting is difficult to achieve. The investigators hypothesize that a closed loop glucose control system that is rule-based will improve upon the glucose control currently achieved by the current open loop manual methods, and that this improved glucose control may improve the outcomes of critically ill patients.

This Earl Feasibility Study will test the ability of a prototype rule-based closed loop glucose control system named FUSION, to provide safe and effective glucose control in subjects with type 1 and type 2 diabetes in a clinical research center (CRC) setting. Subjects with type 1 diabetes have been chosen as safe and effective glucose control is difficult to achieve in these subjects during meal challenges. Subjects with type 2 diabetes have been chosen as they are insulin resistant, which makes their insulin resistance profile similar to that of ICU patients. As this is a safety study of a new medical device, the controlled environment of the CRC is preferable to the less controlled environment of an ICU setting.

The prototype FUSION system to be used in this study will consist to two Dexcom G6 continuous glucose monitors (CGM), the rule-based glucose control software run on an all-in-one medical computer, and two syringe pumps. The prototype system is housed on a medical cart. Based on the average glucose value of the two Dexcom G6 CGM's, and the algorithms of the FUSION systems rule-based glucose control software, the FUSION system will make rate adjustments every 5-10 minutes to the intravenous infusion rates of short acting insulin (NovoLog) and dextrose (D10NS) under its control, in an attempt to keep the subjects glucose in the range of 100-140 mg/dL. The FUSION system only requires entry of the subjects study identification number and weight in kilograms to initiate the system.

For safety reasons, the subjects will have their blood glucose independently measured every 10-60 minutes on the point of care Nova StatStrip system, throughout the 24 hour study period.

The study has halting criteria to avoid recurrent instances of severe hypoglycemia (< 54 mg/dL).

The average of the two CGM's, that is used by the FUSION system for glucose control, will be used for statistical analysis.

Tipo de estudio

Intervencionista

Inscripción (Estimado)

6

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Copia de seguridad de contactos de estudio

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Descripción

Inclusion Criteria:

  • Participants with type 1 and type 2 diabetes are eligible to be included in the study only if they meet all the following criteria:

    1. Are 18-70 years of age, inclusive.
    2. Can understand and sign an informed consent, communicate with the investigator, and understand and comply with the protocol requirements.
    3. Have had a diagnosis of type 1 or type 2 diabetes for a period of at least 3 months.
    4. Use insulin injections at home for glucose control.
    5. Have a hemoglobin A1c (HbA1c) in the range of 7.0 - 10.0%.
    6. Have a hemoglobin in the range for sex:

      1. Females: 7-15.5 grams/dL.
      2. Males: 7-17.5 grams/dl.
    7. Have adequate venous access sites in upper extremities.
    8. Body weight between 40 - 150 kg.

Exclusion Criteria:

  1. Have participated in an interventional medical, surgical, or pharmaceutical study within 30 days of screening.
  2. Have a known hypersensitivity to any of the components of study treatment.
  3. Have skin disease/injury at Dexcom G6 CGM insertion site(s) that would prevent insertion of the CGM.
  4. Currently abuses drugs or alcohol or has a history of abuse that in the investigator's opinion would cause the individual to be noncompliant.
  5. Have a medical condition that in the opinion of the investigator could affect study participation and/or personal well-being.
  6. Have a clinically significant history or presence of any of the following conditions:

    1. Hepatic failure or has alanine aminotransferase (ALT) greater than 3 times the upper limit of normal.
    2. Has an estimated GFR <30 ml/min/1.73 m2 or End Stage Kidney Disease on renal replacement therapy.
    3. Have congestive heart failure greater than class 1 on the NYHA classification system.
    4. Have a history of seizures.
    5. Have a history of cerebrovascular accident.
    6. Have a history of ischemic heart disease.
  7. For female subjects of potential childbearing age (age 18 to 55) they will be excluded if:

    1. Pregnant.
    2. Refuse to agree to a pregnancy test at the time of enrollment.
    3. Have a positive urine pregnancy test at the time of enrollment.
  8. Have a positive Influenza or COVID test within 14 days of visit 3.
  9. Have any viral lower respiratory related symptoms (e.g., cough, chest tightness, respiratory distress) in the 14-day period prior to visit 3.
  10. Have a known unprotected Influenza or COVID exposure in the 14-day period prior to visit 3.
  11. Are currently being treated with Hydroxyurea, a medication which may cause the Dexcom G6 CGM to produce glucose readings that are higher than the actual glucose value.

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Tratamiento
  • Asignación: N / A
  • Modelo Intervencionista: Asignación de un solo grupo
  • Enmascaramiento: Ninguno (etiqueta abierta)

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: Estudio de seguridad del sistema de control de glucosa de bucle cerrado de fusión
Todos los sujetos serán tratados con el sistema de control de glucosa de bucle cerrado de fusión por hasta 24 horas
El sistema FUSION se utilizará para controlar la glucosa de los sujetos en un rango de 100-140 mg/dL. Los datos se recopilarán durante un máximo de 24 horas, o al término anticipado de la sesión de estudio.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Primary Safety Measure
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
The percentage of all glucose values <70 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
Primary Efficacy Measure
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-180 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Percent of All Glucose Values Within Glucose Range <54 mg/dL
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are less than 54 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 54-69 mg/dL
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
The percentage of all glucose values that are within the glucose range of 54-69 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
Number of Hypoglycemic Events (<70 mg/dL)
Periodo de tiempo: From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
The total number of hypoglycemic events less than 70 mg/dL as measured by a continuous series of glucose values less than 70 mg/dL for at least 15 minutes duration.
From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
Percent of All Glucose Values Within Glucose Range 70-100 mg/dL
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-100 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 70-140 mg/dL
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 100-140 mg/dL
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 100-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >140 mg/dL
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values greater than 140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >180 mg/dL
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 180 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >250 mg/dL
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 250 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Average glucose value in mg/dL
Periodo de tiempo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The average of all measured glucose values in mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Measure of dispersion - Coefficient of Variation
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The Coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value by the average glucose value, then multiplying this value by 100.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Clarke Error Grid Analysis of Continuous Glucose Monitor Data
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine the percentage of the two Dexcom G6 CGM's glucose values in each zone using a Clarke error grid analysis. This same calculation will be performed on the average of the two Dexcom G6 CGM systems used by the FUSION system for purposes of glucose control.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
FDA integrated Continuous Glucose Monitor Standards
Periodo de tiempo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine if each individual Dexcom G6 CGM system, and their average value meet FDA criteria for integrated Continuous Glucose Monitor (iCGM) accuracy for the glucose zones <70 mg/dL, 70-180 mg/dL, and >180 mg/dL.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Hemoglobin A1c Percentage Level
Periodo de tiempo: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
C-Peptide Level
Periodo de tiempo: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects C-peptide level will be measured prior to the start of the closed loop glucose control session.
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
Total Daily Dose of Insulin in Units/Kilogram
Periodo de tiempo: As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms
As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
Weight in Kilograms
Periodo de tiempo: As measured just prior to the start of the 24-hour closed loop glucose control session
Average weight in kilograms of the subjects
As measured just prior to the start of the 24-hour closed loop glucose control session
Body Mass Index
Periodo de tiempo: As measured just prior to the start of the 24-hour closed loop glucose control session
The body mass index will be calculated by dividing the subjects weight in kilograms by the square of their height in meters
As measured just prior to the start of the 24-hour closed loop glucose control session

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Investigadores

  • Director de estudio: Leon M DeJournett, MD, Ideal Medical Technologies
  • Investigador principal: Francisco Pasquel, MD, MPH, Emory School Of Medicine

Publicaciones y enlaces útiles

La persona responsable de ingresar información sobre el estudio proporciona voluntariamente estas publicaciones. Estos pueden ser sobre cualquier cosa relacionada con el estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Estimado)

1 de julio de 2026

Finalización primaria (Estimado)

1 de diciembre de 2026

Finalización del estudio (Estimado)

1 de diciembre de 2027

Fechas de registro del estudio

Enviado por primera vez

9 de junio de 2026

Primero enviado que cumplió con los criterios de control de calidad

9 de junio de 2026

Publicado por primera vez (Actual)

12 de junio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

12 de junio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

9 de junio de 2026

Última verificación

1 de junio de 2026

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

Sí

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .