Closed Loop Glucose Control in a Simulated ICU Setting

9 de junho de 2026 atualizado por: Ideal Medical Technologies

Demonstration of an Expert Rule-Based Automated Insulin Delivery (AID) System for People With Type 1 and Type 2 Diabetes

This is a proof of concept safety study of a rule-based closed loop glucose control system designed for use in the intensive care unit setting.

The persons with type 1 and type 2 diabetes in this study will have their glucose controlled to a range of 100-140 mg/dL by a fully autonomous novel rule-based closed loop glucose control system for a period of 24 hours. The subjects will consume three standardized meals and receive a continuous intravenous infusion of dextrose (D-glucose) during the 24 hour study period. Meals will not be announced to the glucose control system.

Visão geral do estudo

Descrição detalhada

Tight glucose control in the intensive care unit (ICU) setting is difficult to achieve. The investigators hypothesize that a closed loop glucose control system that is rule-based will improve upon the glucose control currently achieved by the current open loop manual methods, and that this improved glucose control may improve the outcomes of critically ill patients.

This Earl Feasibility Study will test the ability of a prototype rule-based closed loop glucose control system named FUSION, to provide safe and effective glucose control in subjects with type 1 and type 2 diabetes in a clinical research center (CRC) setting. Subjects with type 1 diabetes have been chosen as safe and effective glucose control is difficult to achieve in these subjects during meal challenges. Subjects with type 2 diabetes have been chosen as they are insulin resistant, which makes their insulin resistance profile similar to that of ICU patients. As this is a safety study of a new medical device, the controlled environment of the CRC is preferable to the less controlled environment of an ICU setting.

The prototype FUSION system to be used in this study will consist to two Dexcom G6 continuous glucose monitors (CGM), the rule-based glucose control software run on an all-in-one medical computer, and two syringe pumps. The prototype system is housed on a medical cart. Based on the average glucose value of the two Dexcom G6 CGM's, and the algorithms of the FUSION systems rule-based glucose control software, the FUSION system will make rate adjustments every 5-10 minutes to the intravenous infusion rates of short acting insulin (NovoLog) and dextrose (D10NS) under its control, in an attempt to keep the subjects glucose in the range of 100-140 mg/dL. The FUSION system only requires entry of the subjects study identification number and weight in kilograms to initiate the system.

For safety reasons, the subjects will have their blood glucose independently measured every 10-60 minutes on the point of care Nova StatStrip system, throughout the 24 hour study period.

The study has halting criteria to avoid recurrent instances of severe hypoglycemia (< 54 mg/dL).

The average of the two CGM's, that is used by the FUSION system for glucose control, will be used for statistical analysis.

Tipo de estudo

Intervencional

Inscrição (Estimado)

6

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Estude backup de contato

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

  • Adulto
  • Adulto mais velho

Aceita Voluntários Saudáveis

Não

Descrição

Inclusion Criteria:

  • Participants with type 1 and type 2 diabetes are eligible to be included in the study only if they meet all the following criteria:

    1. Are 18-70 years of age, inclusive.
    2. Can understand and sign an informed consent, communicate with the investigator, and understand and comply with the protocol requirements.
    3. Have had a diagnosis of type 1 or type 2 diabetes for a period of at least 3 months.
    4. Use insulin injections at home for glucose control.
    5. Have a hemoglobin A1c (HbA1c) in the range of 7.0 - 10.0%.
    6. Have a hemoglobin in the range for sex:

      1. Females: 7-15.5 grams/dL.
      2. Males: 7-17.5 grams/dl.
    7. Have adequate venous access sites in upper extremities.
    8. Body weight between 40 - 150 kg.

Exclusion Criteria:

  1. Have participated in an interventional medical, surgical, or pharmaceutical study within 30 days of screening.
  2. Have a known hypersensitivity to any of the components of study treatment.
  3. Have skin disease/injury at Dexcom G6 CGM insertion site(s) that would prevent insertion of the CGM.
  4. Currently abuses drugs or alcohol or has a history of abuse that in the investigator's opinion would cause the individual to be noncompliant.
  5. Have a medical condition that in the opinion of the investigator could affect study participation and/or personal well-being.
  6. Have a clinically significant history or presence of any of the following conditions:

    1. Hepatic failure or has alanine aminotransferase (ALT) greater than 3 times the upper limit of normal.
    2. Has an estimated GFR <30 ml/min/1.73 m2 or End Stage Kidney Disease on renal replacement therapy.
    3. Have congestive heart failure greater than class 1 on the NYHA classification system.
    4. Have a history of seizures.
    5. Have a history of cerebrovascular accident.
    6. Have a history of ischemic heart disease.
  7. For female subjects of potential childbearing age (age 18 to 55) they will be excluded if:

    1. Pregnant.
    2. Refuse to agree to a pregnancy test at the time of enrollment.
    3. Have a positive urine pregnancy test at the time of enrollment.
  8. Have a positive Influenza or COVID test within 14 days of visit 3.
  9. Have any viral lower respiratory related symptoms (e.g., cough, chest tightness, respiratory distress) in the 14-day period prior to visit 3.
  10. Have a known unprotected Influenza or COVID exposure in the 14-day period prior to visit 3.
  11. Are currently being treated with Hydroxyurea, a medication which may cause the Dexcom G6 CGM to produce glucose readings that are higher than the actual glucose value.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Tratamento
  • Alocação: N / D
  • Modelo Intervencional: Atribuição de grupo único
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: Estudo de segurança do sistema de controle de glicose em loop fechado de fusão
Todos os sujeitos serão tratados com o sistema de controle de glicose em circuito fechado de fusão por até 24 horas
O sistema FUSION será usado para controlar a glicose dos indivíduos em uma faixa de 100-140 mg/dL. Os dados serão coletados por até 24 horas ou após o término antecipado da sessão de estudo.

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Primary Safety Measure
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
The percentage of all glucose values <70 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours
Primary Efficacy Measure
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-180 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Percent of All Glucose Values Within Glucose Range <54 mg/dL
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are less than 54 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 54-69 mg/dL
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
The percentage of all glucose values that are within the glucose range of 54-69 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.
Number of Hypoglycemic Events (<70 mg/dL)
Prazo: From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
The total number of hypoglycemic events less than 70 mg/dL as measured by a continuous series of glucose values less than 70 mg/dL for at least 15 minutes duration.
From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.
Percent of All Glucose Values Within Glucose Range 70-100 mg/dL
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-100 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 70-140 mg/dL
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 70-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range 100-140 mg/dL
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values within the glucose range of 100-140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >140 mg/dL
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values greater than 140 mg/dL
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >180 mg/dL
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 180 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Percent of All Glucose Values Within Glucose Range >250 mg/dL
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The percentage of all glucose values that are greater than 250 mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Average glucose value in mg/dL
Prazo: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
The average of all measured glucose values in mg/dL
From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours
Measure of dispersion - Coefficient of Variation
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The Coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value by the average glucose value, then multiplying this value by 100.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Clarke Error Grid Analysis of Continuous Glucose Monitor Data
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine the percentage of the two Dexcom G6 CGM's glucose values in each zone using a Clarke error grid analysis. This same calculation will be performed on the average of the two Dexcom G6 CGM systems used by the FUSION system for purposes of glucose control.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
FDA integrated Continuous Glucose Monitor Standards
Prazo: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
The study data will be used to determine if each individual Dexcom G6 CGM system, and their average value meet FDA criteria for integrated Continuous Glucose Monitor (iCGM) accuracy for the glucose zones <70 mg/dL, 70-180 mg/dL, and >180 mg/dL.
From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours
Hemoglobin A1c Percentage Level
Prazo: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
C-Peptide Level
Prazo: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects C-peptide level will be measured prior to the start of the closed loop glucose control session.
As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session
Total Daily Dose of Insulin in Units/Kilogram
Prazo: As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
The subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms
As documented within two weeks prior to the start of the 24-hour closed loop glucose control session
Weight in Kilograms
Prazo: As measured just prior to the start of the 24-hour closed loop glucose control session
Average weight in kilograms of the subjects
As measured just prior to the start of the 24-hour closed loop glucose control session
Body Mass Index
Prazo: As measured just prior to the start of the 24-hour closed loop glucose control session
The body mass index will be calculated by dividing the subjects weight in kilograms by the square of their height in meters
As measured just prior to the start of the 24-hour closed loop glucose control session

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Investigadores

  • Diretor de estudo: Leon M DeJournett, MD, Ideal Medical Technologies
  • Investigador principal: Francisco Pasquel, MD, MPH, Emory School Of Medicine

Publicações e links úteis

A pessoa responsável por inserir informações sobre o estudo fornece voluntariamente essas publicações. Estes podem ser sobre qualquer coisa relacionada ao estudo.

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

1 de julho de 2026

Conclusão Primária (Estimado)

1 de dezembro de 2026

Conclusão do estudo (Estimado)

1 de dezembro de 2027

Datas de inscrição no estudo

Enviado pela primeira vez

9 de junho de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

9 de junho de 2026

Primeira postagem (Real)

12 de junho de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

12 de junho de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

9 de junho de 2026

Última verificação

1 de junho de 2026

Mais Informações

Termos relacionados a este estudo

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Sim

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .

Se inscrever