Effect of Dry Eye Treatment on Corneal Astigmatism Measurements
Effect of Dry Eye and Ocular Surface Treatment on Agreement and Repeatability of Corneal Astigmatism Measurements Among Three Devices
This prospective single-center study evaluates the effect of dry eye disease and short-term ocular surface treatment on the agreement and repeatability of corneal astigmatism measurements in patients with age-related cataract. Participants undergoing preoperative cataract evaluation are classified into a control group or a dry eye disease group according to ocular surface symptoms and objective tear film findings.
All participants undergo ocular surface assessment and repeated corneal measurements using three commonly used devices: IOLMaster 700, Pentacam AXL, and iTrace Prime. Measurements include keratometric corneal astigmatism, total corneal astigmatism, and vector components of astigmatism. Patients with dry eye disease receive 0.3% sodium hyaluronate eye drops four times daily for 2 weeks, after which ocular surface assessment and device measurements are repeated.
The primary objective is to determine whether dry eye disease reduces interdevice agreement and test-retest repeatability of corneal astigmatism measurements, and whether short-term ocular surface treatment improves measurement stability. Interdevice agreement is assessed using Bland-Altman analysis, and repeatability is evaluated using within-subject standard deviation, test-retest repeatability, and intraclass correlation coefficient.
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Studietype
Inschrijving (Geschat)
Fase
- Niet toepasbaar
Contacten en locaties
Studiecontact
- Naam: Baoxian Zhuo
- E-mail: 1532483480@qq.com
Studie Contact Back-up
- Naam: Xuanqiao Lin, Doctor
- Telefoonnummer: +8615088920668
- E-mail: 1532483480@qq.com
Studie Locaties
-
-
Shanghai Municipality
-
Shanghai, Shanghai Municipality, China, 200000
- Werving
- Eye and ENT Hospital of Fudan University
-
Contact:
- Xuanqiao Lin
- Telefoonnummer: 15088920668
- E-mail: 1532483480@qq.com
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
- Kind
- Volwassen
- Oudere volwassene
Accepteert gezonde vrijwilligers
Beschrijving
Inclusion Criteria:
- Patients with age-related cataract undergoing preoperative ocular examination.
- Ability to complete ocular surface evaluation, including the Ocular Surface
- Disease Index questionnaire, non-invasive tear film break-up time, tear meniscus height, Schirmer test, and ocular surface staining.
- Ability to complete repeated corneal astigmatism measurements using IOLMaster 700, Pentacam AXL, and iTrace Prime.
- Ability to maintain stable fixation during device measurements. Provision of written informed consent.
Exclusion Criteria:
- Previous ocular surgery or ocular trauma.
- Corneal degeneration, corneal dystrophy, or other corneal diseases that could affect corneal measurements.
- Active corneal or conjunctival infection.
- Uveitis.
- Use of any topical eye drops within 24 hours before examination.
- Use of antiglaucoma medications.
- Contact lens wear within 1 month before examination.
- Previous or ongoing treatment for dry eye disease.
- Poor fixation during device measurements.
- Poor-quality measurements from any study device
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Behandeling
- Toewijzing: Niet-gerandomiseerd
- Interventioneel model: Parallelle opdracht
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Geen tussenkomst: Control Cohort
Patients with age-related cataract who did not meet the diagnostic criteria for dry eye disease.
Participants underwent ocular surface evaluation and repeated corneal astigmatism measurements using IOLMaster 700, Pentacam AXL, and iTrace Prime.
No ocular surface treatment was administered in this cohort.
|
|
|
Experimenteel: Dry Eye Disease Cohort
Patients with age-related cataract who met the diagnostic criteria for dry eye disease.
Participants underwent baseline ocular surface evaluation and repeated corneal astigmatism measurements using IOLMaster 700, Pentacam AXL, and iTrace Prime.
They then received 0.3% sodium hyaluronate eye drops four times daily for 2 weeks, followed by repeat ocular surface evaluation and device measurements using the same protocol.
|
0.3% sodium hyaluronate eye drops administered four times daily for 2 weeks in participants with dry eye disease.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Interdevice Agreement of Corneal Astigmatism Measurements
Tijdsspanne: Baseline and 2 weeks after ocular surface treatment in the dry eye disease arm.
|
Interdevice agreement of corneal astigmatism measurements among IOLMaster 700, Pentacam AXL, and iTrace Prime will be assessed using Bland-Altman analysis.
Pairwise comparisons will be performed for IOLMaster 700 versus Pentacam AXL, IOLMaster 700 versus iTrace Prime, and Pentacam AXL versus iTrace Prime.
The main parameters include keratometric corneal astigmatism, total corneal astigmatism, and their vector components J0 and J45.
The mean difference and 95% limits of agreement will be calculated.
|
Baseline and 2 weeks after ocular surface treatment in the dry eye disease arm.
|
|
Test-Retest Repeatability of Corneal Astigmatism Measurements
Tijdsspanne: Baseline and 2 weeks after ocular surface treatment in the dry eye disease arm.
|
Test-retest repeatability of corneal astigmatism measurements obtained using IOLMaster 700, Pentacam AXL, and iTrace Prime will be evaluated using three consecutive valid measurements from each device.
The main parameters include keratometric corneal astigmatism, total corneal astigmatism, and their vector components J0 and J45.
Repeatability will be assessed using within-subject standard deviation, test-retest repeatability calculated as 2.77 times the within-subject standard deviation, and intraclass correlation coefficient with 95% confidence intervals.
|
Baseline and 2 weeks after ocular surface treatment in the dry eye disease arm.
|
Medewerkers en onderzoekers
Sponsor
Onderzoekers
- Studie stoel: Xuanqiao Lin, Doctor, Fudan University
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- DryEye-2026-001
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Beschrijving IPD-plan
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .