Denne siden ble automatisk oversatt og nøyaktigheten av oversettelsen er ikke garantert. Vennligst referer til engelsk versjon for en kildetekst.

Controlled Insulin Delivery: Combining Technology With Treatment

18. mars 2021 oppdatert av: Sansum Diabetes Research Institute

Hypothesis: Closed-loop control systems for an artificial pancreas using multi-parametric model predictive control can be developed and evaluated safely in patients with Type 1 Diabetes Mellitus (T1DM) to control blood glucose concentrations.

This study seeks to combine real-time continuous glucose sensing with automated insulin delivery in a closed-loop system that will achieve euglycemia in patients with T1DM. The end result of this line of research will be an artificial pancreas that will provide around-the-clock glucose regulation through controlled insulin delivery in response to detected patterns of change in glucose levels.

Studieoversikt

Status

Fullført

Intervensjon / Behandling

Detaljert beskrivelse

The goal of the JDRF Artificial Pancreas Project is to produce an autonomous artificial pancreas that can safely and effectively regulate glycemia in people with type 1 diabetes mellitus. In our work, this fully automated closed-loop system combines a subcutaneous continuous glucose monitor (CGM) and a continuous subcutaneous insulin infusion (CSII) pump with a sophisticated control algorithm. This is a proof-of-concept study to demonstrate that the controller could bring the patient back to a relatively normal glucose concentration after an unannounced meal and from mild hyperglycemia. Once the system is initiated, all insulin delivery is calculated automatically. There was no outside intervention either by the subject or medical personnel. An artificial pancreas system that aims at replicating normal beta-cell function by using the subcutaneous-subcutaneous (sc-sc) route needs to address inherent delays in both glucose sensing and insulin delivery. Our strategic approach is that a closed-loop system should operate safely without any knowledge of meals or other disturbances. We have developed the Artificial Pancreas System (APS©) and used it to clinically evaluate a control strategy that allows efficient glycemic control without any a priori meal information. The Artificial Pancreas device uses the Artificial Pancreas System (APS©) platform with the OmniPod insulin pump, the DexCom SEVEN PLUS CGM and a multi-parametric model predictive control algorithm (mpMPC) with an insulin-on-board (IOB) safety constraint.

Studietype

Intervensjonell

Registrering (Faktiske)

19

Fase

  • Tidlig fase 1

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

    • California
      • Santa Barbara, California, Forente stater, 93105
        • Sansum Diabetes Research Institute
      • Petah Tikva, Israel
        • Schneider Children's Medical Center of Israel

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

18 år og eldre (Voksen, Eldre voksen)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  1. Willing to sign the consent form
  2. Type 1 diabetes for at least 1 year prior to the study
  3. Using continuous subcutaneous insulin infusion pump
  4. Above 21 years of age
  5. Willing to follow the study requirements

Exclusion Criteria:

  1. Allergy to the sensor or to one of its components
  2. Psychiatric disorders
  3. Reported diabetic ketoacidosis within last 3 months
  4. Abnormal liver function (Transaminase > 2 times the upper limit of normal)
  5. Heart failure
  6. Any carcinogenic disease
  7. Any other chronic abnormality
  8. Unwilling to perform or to follow the research protocol
  9. Participation in any other study concurrent with the proposed study
  10. Creatinine concentration above the upper limit of normal for age and sex
  11. Active coronary artery disease
  12. Active gastroparesis
  13. History of uncontrolled seizures
  14. Pregnancy
  15. Untreated adrenal insufficiency
  16. Hypokalemia
  17. Uncontrolled thyroid disease.
  18. Condition, which in the opinion of the investigator, would interfere with patient safety

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Behandling
  • Tildeling: N/A
  • Intervensjonsmodell: Enkeltgruppeoppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: Closed-loop control system

The objective of this study is to automate glucose control in subjects with type 1 diabetes using a computer control algorithm in a controlled in-clinic research setting.

The controller will be evaluated under two conditions:

  1. restoring euglycemia (80-140 mg/dL) when the controller is initiated during a period when the subject's glucose is above the euglycemic range;
  2. restoring euglycemia (80-140 mg/dL) when the controller is challenged with a small unannounced meal (~25 g CHO).

Subjects will arrive fasting at 7am unless needed for hypoglycemia (glucose < 70 mg/dL) and no extra bolus insulin after 3am. An IV catheter will be inserted for blood samples and for IV administration of glucose if necessary. Blood samples will be analyzed for glucose by YSI 2300Stat every 30 minutes.

Breakfast consisting of 25g of CHO will be eaten at 7:30 am and the subject will bolus for this amount of CHO. The controller is switched "on" on the down slope of the meal response, and the subject is brought to a basal steady-state by the controller. Target blood glucose is 110 ± 30-mg/dL.

After approximately 3 hours a small lunch will be eaten consisting of 25g of CHO (unannounced meal challenge). The subject will be monitored until blood glucose returns to euglycemia.

Andre navn:
  • Closed-loop Artificial Pancreas Device

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Restoration of Euglycemia
Tidsramme: 12 hours
The primary endpoint of this pilot study is successful restoration of euglycemia from the two perturbed conditions using closed loop control with minimal hypo- and hyperglycemia exposure. The first condition involves restoration of euglycemia when the subject is in a hyperglycemic state. The second condition involves restoration of euglycemia following consumption of a meal containing 25g CHO with no insulin bolus. The patients will be followed for the duration of the 12 hour study.
12 hours

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Average percent-of-time-in-range (80 - 180 mg/dL)
Tidsramme: 12 hours
All reported blood glucose values per both CGM and YSI ware analyzed for average percent-of-time-in-range (80 - 180 mg/dL). The participants will be followed for the duration of the 12 hour study.
12 hours

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Etterforskere

  • Hovedetterforsker: Lois Jovanovic, M.D., Sansum Diabetes Research Institute

Publikasjoner og nyttige lenker

Den som er ansvarlig for å legge inn informasjon om studien leverer frivillig disse publikasjonene. Disse kan handle om alt relatert til studiet.

Generelle publikasjoner

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart

1. juli 2008

Primær fullføring (Faktiske)

1. januar 2010

Studiet fullført (Faktiske)

1. januar 2010

Datoer for studieregistrering

Først innsendt

15. november 2011

Først innsendt som oppfylte QC-kriteriene

30. november 2011

Først lagt ut (Anslag)

2. desember 2011

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

23. mars 2021

Siste oppdatering sendt inn som oppfylte QC-kriteriene

18. mars 2021

Sist bekreftet

1. mars 2021

Mer informasjon

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .

Abonnere