Comparison of Visual Outcomes After Implantation of the POD FT and the POD F (PoyPODF/FT)

29 juni 2021 uppdaterad av: Beaver-Visitec International, Inc.

Comparison of Visual Outcomes After Implantation of the POD 26P AY FT (PhysIOL) and the POD 26P AY F (PhysIOL)

There is no difference of the photic phenomena in the two groups.

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

Studieöversikt

Status

Avslutad

Betingelser

Detaljerad beskrivning

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

We used the 'TrialSize' package of the R software. Here are the parameters defined for the sample size calculation:

  1. Alpha: 0.05
  2. Power : 0.90
  3. Two sided test
  4. Equal group allocation
  5. Standard deviation for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.10
  6. Expected mean for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.89 (true difference between the means equal 0)
  7. An equivalence limit of 0.05, 0.10 and 0.15 Equivalence limit: 0.15 In order to show no difference between these two groups, 12 patients by groups are needed. With a drop-out of 15%, we obtain 30 patients (15 patients by group) to recruit in order to obtain 24 valuable patients.

Equivalence limit: 0.10 In order to show no difference between these two groups, 26 patients by groups are needed. With a drop-out of 15%, we obtain 62 patients (31 patients by group) to recruit in order to obtain 52 valuable patients.

This total number of subjects with such requirement is 26 patients in each group.

Group 1 will be made of 26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT and Group 2 will be made of 26 patients with corneal astigmatism lower than 1.0D in both eyes.

Interim analysis after the 12th patient of each group will have completed the last follow-up will be achieved to confirm the equivalence with the equivalence limit of 0.15.

The POD platform in 26 % (non toric version) is available since 2008 in the monofocal version and since 2012 in the toric one (POD 6P AY T, Ankoris®, PhysIOL).The so-called POD design is characterised by two symmetrical bifid haptics with opened loops that provide four points of support to ensure IOL stability in the capsular bag. The dual C-loop profile shortens the loops, thus improving the pliability of the implant during its injection through a micro-incision. These shortened loops should permit a clockwise or anticlockwise rotary motion and should facilitate the comfortable fitting of the implant orientation after unfolding without elastic-restoring torque. The double C-loop POD design offers 4 support points for the implant within the capsule and ensures its stability. The V-shape of the upper and lower haptics together with the flexibility of the implant allow the loop separation angle to be adjusted and ensure the most accurate adaptation to the capsular bag size. A finite element structural analysis of the implant showed that the elasticity at the distal part of the loop prevents the constraints from being transmitted onto the optical portion of the implant. The trifocal optic is associated to the POD platform and is made of 26 % raw material.

The rotational stability of this non toric version has been clinically investigated on 117 consecutive eyes 1 week, 3 weeks, 3 and 12 months after implantation. The absolute rotation of the lens averaged 2.5 ± 2.6°. The position of the center of the implant relative to the pupillary center expressed by a vector was 0.01 mm at 70°.

The investigational product (POD 26P AY FT) is CE labeled since January 2014. Main IOL features are:

  • Biconvex toric aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptic design
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Cylinders at the IOL plane: 1, 1.5, 2.25, 3, 3.75, 4.5, 5.25, 6 D
  • Addition +1.75D / +3.5 D

The control product (POD 26P AY F) is CE labeled since 2010. Main IOL features are:

  • Biconvex, aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptics design which minimizes decentration and optimizes the rotational stability
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Addition +1.75D / +3.5 D

Studietyp

Observationell

Inskrivning (Faktisk)

63

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

      • Madrid, Spanien, 28040
        • Hospital Clinico San Carlos

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

21 år och äldre (Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Testmetod

Icke-sannolikhetsprov

Studera befolkning

The study population consists of patients undergoing routine cataract surgery. The study participants receive a bilateral implantation of trifocal intraocular lenses. Depending on the preoperative corneal astigmatism, the patients receive either non-toric trifocal IOLs (POD F) or toric trifocal IOLs (POD FT) in both eyes.

Beskrivning

Inclusion Criteria:

  • Age-related cataract
  • Preoperative Corneal astigmatism < 0.75 D
  • Age 21 and older
  • Visual Acuity > 0.05
  • Normal findings in the medical history and physical examination unless the investigator considers an abnormality to be clinically irrelevant.
  • Patient willing multifocal implantation and with realistic expectations.

Exclusion Criteria:

  • Relevant other ophthalmic diseases such as pseudoexfoliation syndrome, floppy iris syndrome, corneal pathologies, retinal pathology (diabetic maculopathy, myopic maculopathy, age related macular degeneration...)
  • Previous ocular surgery or trauma.

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Observationsmodeller: Case-Control
  • Tidsperspektiv: Blivande

Kohorter och interventioner

Grupp / Kohort
Intervention / Behandling
POD 26P AY FT
26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT
Routine cataract surgery
POD 26P AY F
26 patients with corneal astigmatism lower than 1.0D in both eyes will be implanted with the POD 26P AY F
Routine cataract surgery

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Determine the equivalence of the glare
Tidsram: 3 Months
The primary endpoint is to show statistically equivalent outcomes on glare at the 3 months follow up visit. A significance level of p = 0.05 will be considered statistically significant.
3 Months

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Measurement of far, intermediate and near visual acuity
Tidsram: 6 Months
The secondary objective is to measure far, intermediate and near visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Refraction
Tidsram: 6 Months
The secondary objective is to measure the refraction of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Centration
Tidsram: 6 Months
The secondary objective is to measure the centration of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Rotational stability
Tidsram: 6 Months
The secondary objective is to measure the rotational stability of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Tidsram: 6 Months
The secondary objective is to determine the equivalence of the visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Tidsram: 6 Months
The secondary objective is to determine the contrast sensitivity at far of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Utredare

  • Huvudutredare: Francisco Poyales Galan, PhD, CEIC Hospital Clinical San Carlos

Publikationer och användbara länkar

Den som ansvarar för att lägga in information om studien tillhandahåller frivilligt dessa publikationer. Dessa kan handla om allt som har med studien att göra.

Allmänna publikationer

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

16 juli 2014

Primärt slutförande (Faktisk)

20 juli 2017

Avslutad studie (Faktisk)

20 juli 2017

Studieregistreringsdatum

Först inskickad

18 juni 2021

Först inskickad som uppfyllde QC-kriterierna

21 juni 2021

Första postat (Faktisk)

29 juni 2021

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

1 juli 2021

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

29 juni 2021

Senast verifierad

1 juni 2021

Mer information

Termer relaterade till denna studie

Ytterligare relevanta MeSH-villkor

Andra studie-ID-nummer

  • Poyales/POD FT - POD F

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

Nej

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .