Comparison of Visual Outcomes After Implantation of the POD FT and the POD F (PoyPODF/FT)

29 de junho de 2021 atualizado por: Beaver-Visitec International, Inc.

Comparison of Visual Outcomes After Implantation of the POD 26P AY FT (PhysIOL) and the POD 26P AY F (PhysIOL)

There is no difference of the photic phenomena in the two groups.

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

Visão geral do estudo

Status

Concluído

Condições

Descrição detalhada

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

We used the 'TrialSize' package of the R software. Here are the parameters defined for the sample size calculation:

  1. Alpha: 0.05
  2. Power : 0.90
  3. Two sided test
  4. Equal group allocation
  5. Standard deviation for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.10
  6. Expected mean for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.89 (true difference between the means equal 0)
  7. An equivalence limit of 0.05, 0.10 and 0.15 Equivalence limit: 0.15 In order to show no difference between these two groups, 12 patients by groups are needed. With a drop-out of 15%, we obtain 30 patients (15 patients by group) to recruit in order to obtain 24 valuable patients.

Equivalence limit: 0.10 In order to show no difference between these two groups, 26 patients by groups are needed. With a drop-out of 15%, we obtain 62 patients (31 patients by group) to recruit in order to obtain 52 valuable patients.

This total number of subjects with such requirement is 26 patients in each group.

Group 1 will be made of 26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT and Group 2 will be made of 26 patients with corneal astigmatism lower than 1.0D in both eyes.

Interim analysis after the 12th patient of each group will have completed the last follow-up will be achieved to confirm the equivalence with the equivalence limit of 0.15.

The POD platform in 26 % (non toric version) is available since 2008 in the monofocal version and since 2012 in the toric one (POD 6P AY T, Ankoris®, PhysIOL).The so-called POD design is characterised by two symmetrical bifid haptics with opened loops that provide four points of support to ensure IOL stability in the capsular bag. The dual C-loop profile shortens the loops, thus improving the pliability of the implant during its injection through a micro-incision. These shortened loops should permit a clockwise or anticlockwise rotary motion and should facilitate the comfortable fitting of the implant orientation after unfolding without elastic-restoring torque. The double C-loop POD design offers 4 support points for the implant within the capsule and ensures its stability. The V-shape of the upper and lower haptics together with the flexibility of the implant allow the loop separation angle to be adjusted and ensure the most accurate adaptation to the capsular bag size. A finite element structural analysis of the implant showed that the elasticity at the distal part of the loop prevents the constraints from being transmitted onto the optical portion of the implant. The trifocal optic is associated to the POD platform and is made of 26 % raw material.

The rotational stability of this non toric version has been clinically investigated on 117 consecutive eyes 1 week, 3 weeks, 3 and 12 months after implantation. The absolute rotation of the lens averaged 2.5 ± 2.6°. The position of the center of the implant relative to the pupillary center expressed by a vector was 0.01 mm at 70°.

The investigational product (POD 26P AY FT) is CE labeled since January 2014. Main IOL features are:

  • Biconvex toric aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptic design
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Cylinders at the IOL plane: 1, 1.5, 2.25, 3, 3.75, 4.5, 5.25, 6 D
  • Addition +1.75D / +3.5 D

The control product (POD 26P AY F) is CE labeled since 2010. Main IOL features are:

  • Biconvex, aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptics design which minimizes decentration and optimizes the rotational stability
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Addition +1.75D / +3.5 D

Tipo de estudo

Observacional

Inscrição (Real)

63

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Locais de estudo

      • Madrid, Espanha, 28040
        • Hospital Clínico San Carlos

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

21 anos e mais velhos (Adulto, Adulto mais velho)

Aceita Voluntários Saudáveis

Não

Gêneros Elegíveis para o Estudo

Tudo

Método de amostragem

Amostra Não Probabilística

População do estudo

The study population consists of patients undergoing routine cataract surgery. The study participants receive a bilateral implantation of trifocal intraocular lenses. Depending on the preoperative corneal astigmatism, the patients receive either non-toric trifocal IOLs (POD F) or toric trifocal IOLs (POD FT) in both eyes.

Descrição

Inclusion Criteria:

  • Age-related cataract
  • Preoperative Corneal astigmatism < 0.75 D
  • Age 21 and older
  • Visual Acuity > 0.05
  • Normal findings in the medical history and physical examination unless the investigator considers an abnormality to be clinically irrelevant.
  • Patient willing multifocal implantation and with realistic expectations.

Exclusion Criteria:

  • Relevant other ophthalmic diseases such as pseudoexfoliation syndrome, floppy iris syndrome, corneal pathologies, retinal pathology (diabetic maculopathy, myopic maculopathy, age related macular degeneration...)
  • Previous ocular surgery or trauma.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Modelos de observação: Controle de caso
  • Perspectivas de Tempo: Prospectivo

Coortes e Intervenções

Grupo / Coorte
Intervenção / Tratamento
POD 26P AY FT
26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT
Routine cataract surgery
POD 26P AY F
26 patients with corneal astigmatism lower than 1.0D in both eyes will be implanted with the POD 26P AY F
Routine cataract surgery

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Determine the equivalence of the glare
Prazo: 3 Months
The primary endpoint is to show statistically equivalent outcomes on glare at the 3 months follow up visit. A significance level of p = 0.05 will be considered statistically significant.
3 Months

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Measurement of far, intermediate and near visual acuity
Prazo: 6 Months
The secondary objective is to measure far, intermediate and near visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Refraction
Prazo: 6 Months
The secondary objective is to measure the refraction of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Centration
Prazo: 6 Months
The secondary objective is to measure the centration of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Rotational stability
Prazo: 6 Months
The secondary objective is to measure the rotational stability of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Prazo: 6 Months
The secondary objective is to determine the equivalence of the visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Prazo: 6 Months
The secondary objective is to determine the contrast sensitivity at far of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Investigadores

  • Investigador principal: Francisco Poyales Galan, PhD, CEIC Hospital Clinical San Carlos

Publicações e links úteis

A pessoa responsável por inserir informações sobre o estudo fornece voluntariamente essas publicações. Estes podem ser sobre qualquer coisa relacionada ao estudo.

Publicações Gerais

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Real)

16 de julho de 2014

Conclusão Primária (Real)

20 de julho de 2017

Conclusão do estudo (Real)

20 de julho de 2017

Datas de inscrição no estudo

Enviado pela primeira vez

18 de junho de 2021

Enviado pela primeira vez que atendeu aos critérios de CQ

21 de junho de 2021

Primeira postagem (Real)

29 de junho de 2021

Atualizações de registro de estudo

Última Atualização Postada (Real)

1 de julho de 2021

Última atualização enviada que atendeu aos critérios de controle de qualidade

29 de junho de 2021

Última verificação

1 de junho de 2021

Mais Informações

Termos relacionados a este estudo

Termos MeSH relevantes adicionais

Outros números de identificação do estudo

  • Poyales/POD FT - POD F

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

Não

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .

Se inscrever