Comparison of Visual Outcomes After Implantation of the POD FT and the POD F (PoyPODF/FT)

29 giugno 2021 aggiornato da: Beaver-Visitec International, Inc.

Comparison of Visual Outcomes After Implantation of the POD 26P AY FT (PhysIOL) and the POD 26P AY F (PhysIOL)

There is no difference of the photic phenomena in the two groups.

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

Panoramica dello studio

Stato

Completato

Condizioni

Descrizione dettagliata

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

We used the 'TrialSize' package of the R software. Here are the parameters defined for the sample size calculation:

  1. Alpha: 0.05
  2. Power : 0.90
  3. Two sided test
  4. Equal group allocation
  5. Standard deviation for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.10
  6. Expected mean for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.89 (true difference between the means equal 0)
  7. An equivalence limit of 0.05, 0.10 and 0.15 Equivalence limit: 0.15 In order to show no difference between these two groups, 12 patients by groups are needed. With a drop-out of 15%, we obtain 30 patients (15 patients by group) to recruit in order to obtain 24 valuable patients.

Equivalence limit: 0.10 In order to show no difference between these two groups, 26 patients by groups are needed. With a drop-out of 15%, we obtain 62 patients (31 patients by group) to recruit in order to obtain 52 valuable patients.

This total number of subjects with such requirement is 26 patients in each group.

Group 1 will be made of 26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT and Group 2 will be made of 26 patients with corneal astigmatism lower than 1.0D in both eyes.

Interim analysis after the 12th patient of each group will have completed the last follow-up will be achieved to confirm the equivalence with the equivalence limit of 0.15.

The POD platform in 26 % (non toric version) is available since 2008 in the monofocal version and since 2012 in the toric one (POD 6P AY T, Ankoris®, PhysIOL).The so-called POD design is characterised by two symmetrical bifid haptics with opened loops that provide four points of support to ensure IOL stability in the capsular bag. The dual C-loop profile shortens the loops, thus improving the pliability of the implant during its injection through a micro-incision. These shortened loops should permit a clockwise or anticlockwise rotary motion and should facilitate the comfortable fitting of the implant orientation after unfolding without elastic-restoring torque. The double C-loop POD design offers 4 support points for the implant within the capsule and ensures its stability. The V-shape of the upper and lower haptics together with the flexibility of the implant allow the loop separation angle to be adjusted and ensure the most accurate adaptation to the capsular bag size. A finite element structural analysis of the implant showed that the elasticity at the distal part of the loop prevents the constraints from being transmitted onto the optical portion of the implant. The trifocal optic is associated to the POD platform and is made of 26 % raw material.

The rotational stability of this non toric version has been clinically investigated on 117 consecutive eyes 1 week, 3 weeks, 3 and 12 months after implantation. The absolute rotation of the lens averaged 2.5 ± 2.6°. The position of the center of the implant relative to the pupillary center expressed by a vector was 0.01 mm at 70°.

The investigational product (POD 26P AY FT) is CE labeled since January 2014. Main IOL features are:

  • Biconvex toric aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptic design
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Cylinders at the IOL plane: 1, 1.5, 2.25, 3, 3.75, 4.5, 5.25, 6 D
  • Addition +1.75D / +3.5 D

The control product (POD 26P AY F) is CE labeled since 2010. Main IOL features are:

  • Biconvex, aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptics design which minimizes decentration and optimizes the rotational stability
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Addition +1.75D / +3.5 D

Tipo di studio

Osservativo

Iscrizione (Effettivo)

63

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

      • Madrid, Spagna, 28040
        • Hospital Clinico San Carlos

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

21 anni e precedenti (Adulto, Adulto più anziano)

Accetta volontari sani

No

Sessi ammissibili allo studio

Tutto

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

The study population consists of patients undergoing routine cataract surgery. The study participants receive a bilateral implantation of trifocal intraocular lenses. Depending on the preoperative corneal astigmatism, the patients receive either non-toric trifocal IOLs (POD F) or toric trifocal IOLs (POD FT) in both eyes.

Descrizione

Inclusion Criteria:

  • Age-related cataract
  • Preoperative Corneal astigmatism < 0.75 D
  • Age 21 and older
  • Visual Acuity > 0.05
  • Normal findings in the medical history and physical examination unless the investigator considers an abnormality to be clinically irrelevant.
  • Patient willing multifocal implantation and with realistic expectations.

Exclusion Criteria:

  • Relevant other ophthalmic diseases such as pseudoexfoliation syndrome, floppy iris syndrome, corneal pathologies, retinal pathology (diabetic maculopathy, myopic maculopathy, age related macular degeneration...)
  • Previous ocular surgery or trauma.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Modelli osservazionali: Caso di controllo
  • Prospettive temporali: Prospettiva

Coorti e interventi

Gruppo / Coorte
Intervento / Trattamento
POD 26P AY FT
26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT
Routine cataract surgery
POD 26P AY F
26 patients with corneal astigmatism lower than 1.0D in both eyes will be implanted with the POD 26P AY F
Routine cataract surgery

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Determine the equivalence of the glare
Lasso di tempo: 3 Months
The primary endpoint is to show statistically equivalent outcomes on glare at the 3 months follow up visit. A significance level of p = 0.05 will be considered statistically significant.
3 Months

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Measurement of far, intermediate and near visual acuity
Lasso di tempo: 6 Months
The secondary objective is to measure far, intermediate and near visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Refraction
Lasso di tempo: 6 Months
The secondary objective is to measure the refraction of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Centration
Lasso di tempo: 6 Months
The secondary objective is to measure the centration of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Rotational stability
Lasso di tempo: 6 Months
The secondary objective is to measure the rotational stability of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Lasso di tempo: 6 Months
The secondary objective is to determine the equivalence of the visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Lasso di tempo: 6 Months
The secondary objective is to determine the contrast sensitivity at far of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Investigatori

  • Investigatore principale: Francisco Poyales Galan, PhD, CEIC Hospital Clinical San Carlos

Pubblicazioni e link utili

La persona responsabile dell'inserimento delle informazioni sullo studio fornisce volontariamente queste pubblicazioni. Questi possono riguardare qualsiasi cosa relativa allo studio.

Pubblicazioni generali

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

16 luglio 2014

Completamento primario (Effettivo)

20 luglio 2017

Completamento dello studio (Effettivo)

20 luglio 2017

Date di iscrizione allo studio

Primo inviato

18 giugno 2021

Primo inviato che soddisfa i criteri di controllo qualità

21 giugno 2021

Primo Inserito (Effettivo)

29 giugno 2021

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

1 luglio 2021

Ultimo aggiornamento inviato che soddisfa i criteri QC

29 giugno 2021

Ultimo verificato

1 giugno 2021

Maggiori informazioni

Termini relativi a questo studio

Termini MeSH pertinenti aggiuntivi

Altri numeri di identificazione dello studio

  • Poyales/POD FT - POD F

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

No

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Prove cliniche su Cataratta

Prove cliniche su POD 26P AY FT

Sottoscrivi