Diese Seite wurde automatisch übersetzt und die Genauigkeit der Übersetzung wird nicht garantiert. Bitte wende dich an die englische Version für einen Quelltext.

Comparison of Visual Outcomes After Implantation of the POD FT and the POD F (PoyPODF/FT)

29. Juni 2021 aktualisiert von: Beaver-Visitec International, Inc.

Comparison of Visual Outcomes After Implantation of the POD 26P AY FT (PhysIOL) and the POD 26P AY F (PhysIOL)

There is no difference of the photic phenomena in the two groups.

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

Studienübersicht

Status

Abgeschlossen

Bedingungen

Detaillierte Beschreibung

The study will be a prospective non-randomised controlled study. The target is then to determine the equivalence in the glare. Assumption of Type I error is p=0.05 as level of significance and Type II error is alpha=0.8 as power.

We used the 'TrialSize' package of the R software. Here are the parameters defined for the sample size calculation:

  1. Alpha: 0.05
  2. Power : 0.90
  3. Two sided test
  4. Equal group allocation
  5. Standard deviation for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.10
  6. Expected mean for the two groups (POD 26P AY F et POD 26P AY FT) equal to 0.89 (true difference between the means equal 0)
  7. An equivalence limit of 0.05, 0.10 and 0.15 Equivalence limit: 0.15 In order to show no difference between these two groups, 12 patients by groups are needed. With a drop-out of 15%, we obtain 30 patients (15 patients by group) to recruit in order to obtain 24 valuable patients.

Equivalence limit: 0.10 In order to show no difference between these two groups, 26 patients by groups are needed. With a drop-out of 15%, we obtain 62 patients (31 patients by group) to recruit in order to obtain 52 valuable patients.

This total number of subjects with such requirement is 26 patients in each group.

Group 1 will be made of 26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT and Group 2 will be made of 26 patients with corneal astigmatism lower than 1.0D in both eyes.

Interim analysis after the 12th patient of each group will have completed the last follow-up will be achieved to confirm the equivalence with the equivalence limit of 0.15.

The POD platform in 26 % (non toric version) is available since 2008 in the monofocal version and since 2012 in the toric one (POD 6P AY T, Ankoris®, PhysIOL).The so-called POD design is characterised by two symmetrical bifid haptics with opened loops that provide four points of support to ensure IOL stability in the capsular bag. The dual C-loop profile shortens the loops, thus improving the pliability of the implant during its injection through a micro-incision. These shortened loops should permit a clockwise or anticlockwise rotary motion and should facilitate the comfortable fitting of the implant orientation after unfolding without elastic-restoring torque. The double C-loop POD design offers 4 support points for the implant within the capsule and ensures its stability. The V-shape of the upper and lower haptics together with the flexibility of the implant allow the loop separation angle to be adjusted and ensure the most accurate adaptation to the capsular bag size. A finite element structural analysis of the implant showed that the elasticity at the distal part of the loop prevents the constraints from being transmitted onto the optical portion of the implant. The trifocal optic is associated to the POD platform and is made of 26 % raw material.

The rotational stability of this non toric version has been clinically investigated on 117 consecutive eyes 1 week, 3 weeks, 3 and 12 months after implantation. The absolute rotation of the lens averaged 2.5 ± 2.6°. The position of the center of the implant relative to the pupillary center expressed by a vector was 0.01 mm at 70°.

The investigational product (POD 26P AY FT) is CE labeled since January 2014. Main IOL features are:

  • Biconvex toric aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptic design
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Cylinders at the IOL plane: 1, 1.5, 2.25, 3, 3.75, 4.5, 5.25, 6 D
  • Addition +1.75D / +3.5 D

The control product (POD 26P AY F) is CE labeled since 2010. Main IOL features are:

  • Biconvex, aspheric (-0.11um SA) blue-blocking, hydrophilic acrylic IOL with 26% equlibrium water content, implantable through ≥ 2.2mm incision.
  • 4-point haptics design which minimizes decentration and optimizes the rotational stability
  • 6 mm of optical diameter / 11.4 mm of total diameter
  • Dioptric range (SE): 6 to 35 D by 0.5 D increments
  • Addition +1.75D / +3.5 D

Studientyp

Beobachtungs

Einschreibung (Tatsächlich)

63

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienorte

      • Madrid, Spanien, 28040
        • Hospital Clinico San Carlos

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

21 Jahre und älter (Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

Nein

Studienberechtigte Geschlechter

Alle

Probenahmeverfahren

Nicht-Wahrscheinlichkeitsprobe

Studienpopulation

The study population consists of patients undergoing routine cataract surgery. The study participants receive a bilateral implantation of trifocal intraocular lenses. Depending on the preoperative corneal astigmatism, the patients receive either non-toric trifocal IOLs (POD F) or toric trifocal IOLs (POD FT) in both eyes.

Beschreibung

Inclusion Criteria:

  • Age-related cataract
  • Preoperative Corneal astigmatism < 0.75 D
  • Age 21 and older
  • Visual Acuity > 0.05
  • Normal findings in the medical history and physical examination unless the investigator considers an abnormality to be clinically irrelevant.
  • Patient willing multifocal implantation and with realistic expectations.

Exclusion Criteria:

  • Relevant other ophthalmic diseases such as pseudoexfoliation syndrome, floppy iris syndrome, corneal pathologies, retinal pathology (diabetic maculopathy, myopic maculopathy, age related macular degeneration...)
  • Previous ocular surgery or trauma.

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Beobachtungsmodelle: Fallkontrolle
  • Zeitperspektiven: Interessent

Kohorten und Interventionen

Gruppe / Kohorte
Intervention / Behandlung
POD 26P AY FT
26 patients with corneal astigmatism higher than 1.0D in both eyes and lower than 2.5D and will be implanted with the POD 26P AY FT
Routine cataract surgery
POD 26P AY F
26 patients with corneal astigmatism lower than 1.0D in both eyes will be implanted with the POD 26P AY F
Routine cataract surgery

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Determine the equivalence of the glare
Zeitfenster: 3 Months
The primary endpoint is to show statistically equivalent outcomes on glare at the 3 months follow up visit. A significance level of p = 0.05 will be considered statistically significant.
3 Months

Sekundäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Measurement of far, intermediate and near visual acuity
Zeitfenster: 6 Months
The secondary objective is to measure far, intermediate and near visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Refraction
Zeitfenster: 6 Months
The secondary objective is to measure the refraction of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Centration
Zeitfenster: 6 Months
The secondary objective is to measure the centration of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Rotational stability
Zeitfenster: 6 Months
The secondary objective is to measure the rotational stability of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Zeitfenster: 6 Months
The secondary objective is to determine the equivalence of the visual acuity of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months
Equivalence of the visual acuity and the contrast sensitivity
Zeitfenster: 6 Months
The secondary objective is to determine the contrast sensitivity at far of the POD AY 26P FT with respect to the POD AY 26P F.
6 Months

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Ermittler

  • Hauptermittler: Francisco Poyales Galan, PhD, CEIC Hospital Clinical San Carlos

Publikationen und hilfreiche Links

Die Bereitstellung dieser Publikationen erfolgt freiwillig durch die für die Eingabe von Informationen über die Studie verantwortliche Person. Diese können sich auf alles beziehen, was mit dem Studium zu tun hat.

Allgemeine Veröffentlichungen

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

16. Juli 2014

Primärer Abschluss (Tatsächlich)

20. Juli 2017

Studienabschluss (Tatsächlich)

20. Juli 2017

Studienanmeldedaten

Zuerst eingereicht

18. Juni 2021

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

21. Juni 2021

Zuerst gepostet (Tatsächlich)

29. Juni 2021

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

1. Juli 2021

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

29. Juni 2021

Zuletzt verifiziert

1. Juni 2021

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Zusätzliche relevante MeSH-Bedingungen

Andere Studien-ID-Nummern

  • Poyales/POD FT - POD F

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

Nein

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .

Klinische Studien zur POD 26P AY FT

3
Abonnieren