Performance and Acceptability of iDesign
2013年5月16日 更新者:Abbott Medical Optics
A Prospective Study to Evaluate the Performance and Acceptability of the iDesign Advanced WaveScan Study(tm) System
The results of this trial will demonstrate that the iDesign System performs as intended and is acceptable in a clinical setting.
研究概览
详细说明
LASIK treatment across the range of myopia with or without astigmatism, hyperopia with or without astigmatism, and mixed astigmatism.
研究类型
介入性
注册 (实际的)
77
阶段
- 不适用
联系人和位置
本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。
学习地点
-
-
British Columbia
-
Victoria、British Columbia、加拿大、V8X 1X2
- Clearly LASIK
-
-
Manitoba
-
Winnipeg、Manitoba、加拿大、R3C 3J5
- Image Plus Laser Eye Center
-
-
Ontario
-
Ottawa、Ontario、加拿大、K1H 8L6
- University of Ottawa Eye Institute, The Ottawa Hospital
-
Toronto、Ontario、加拿大、M4P 2E5
- Yonge-Eglington Laser
-
-
参与标准
研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。
资格标准
适合学习的年龄
18年 及以上 (成人、年长者)
接受健康志愿者
不
有资格学习的性别
全部
描述
Inclusion Criteria:
- Male or female, of any race, and at least 18 years old at the time the consent form is signed.
- The refractive error, based on manifest refraction (adjusted for optical infinity), must be MRSE up to -15.00 D with manifest cylinder between 0.00 and -6.00 D; MRSE up to +9.00 D with cylinder between 0.00 and +6.00 D; or manifest cylinder (from 1.00 to 6.00 D) greater than the magnitude of sphere, and the cylinder and sphere have opposite signs.
- BSCVA of 20/20 or better.
UCVA:
- Myopes must be 20/40 or worse.
- Subjects with Hyperopia or mixed astigmatism, treatment may be based on difficulty maintaining UCVA of 20/40 evidenced by spectacle/contact lens dependence for distance, as documented by the investigator.
- Wavefront diameter ≥ 4.0 mm.
- Sufficient agreement between manifest refraction (adjusted for optical infinity) and iDesign System refraction (sphere, cylinder, and axis) to allow treatment selection and generation of an ablation profile as determined by the investigational iDesign System software.
- Manifest refraction sphere within ± 0.75 D of cycloplegic refraction sphere.
- Anticipated post-operative stromal bed thickness of at least 250 microns, based on pre-operative central corneal pachymetry minus the maximum treatment depth to be ablated (as calculated by the iDesign System) and the intended flap thickness.
- Anticipated post-operative keratometry value (based on pre-operative iDesign System refraction and keratometry) that is appropriate.
- A stable refractive error (based on a previous exam, medical records, or prescription) as compared to the unadjusted pre-operative manifest refraction.
- Subjects who currently wear contact lenses must demonstrate refractive stability in the operative eye. Rigid or toric lenses must be removed for at least 2 weeks and SCL for at least 3 days prior to the first refraction used to establish stability. A second refraction with a change of no more than ± 0.50 D MRSE at least 7 days after baseline.
- Willing and capable of returning for follow-up examinations for the duration of the study.
Exclusion Criteria:
- Women who are pregnant, breast-feeding, or intend to become pregnant over the course of the study, as determined by verbal inquiry, or another condition associated with the fluctuation of hormones that could lead to refractive changes.
- Subjects with an ametropic or amblyopic fellow eye not meeting all inclusion criteria that does not fall within the indications for treatment using the VISX® STAR S4 IR™ Excimer Laser.
- Concurrent use of systemic (including inhaled) medications that may impair healing, including but not limited to: steroids, antimetabolites, isotretinoin (Accutane™) within 6 months of treatment, and amiodarone hydrochloride (Cordarone™) within 12 months of treatment.
- Acute or chronic disease, illness, or treatment that would increase the operative risk (e.g., immuno-compromised, autoimmune disease, connective tissue disease, clinically significant atopic disease, diabetes, etc.).
- Subjects with a cardiac pacemaker, implanted defibrillator, implanted neurostimulator, or any active electrical implants.
- Ocular condition (other than high myopia) that may predispose the subject to future complications, such as history or evidence of active or inactive corneal disease
- Previous intraocular or corneal surgery that might confound the outcome of the study or increase risk to the subject.
- Known sensitivity or inappropriate responsiveness to any of the medications used in the post-operative course.
- Concurrent participation in any other clinical study.
学习计划
本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。
研究是如何设计的?
设计细节
- 主要用途:治疗
- 分配:非随机化
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:Refractive Error
|
CustomVue LASIK targeted for emmetropia
其他名称:
|
研究衡量的是什么?
主要结果指标
结果测量 |
大体时间 |
|---|---|
|
Percentage of Eyes With Uncorrected Visual Acuity (UCVA) of 20/40 or Better.
大体时间:6 months
|
6 months
|
其他结果措施
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
失去超过 2 条线的眼睛百分比 最佳眼镜矫正视力 (BSCVA)
大体时间:6个月
|
6个月
|
|
|
Percentage of Eyes With Manifest Refraction Spherical Equivalent Within 1.0 D
大体时间:6 months
|
6 months
|
|
|
Percentage of Eyes With Best Spectacle Corrected Visual Acuity (BSCVA) Worse Than 20/40
大体时间:6 Months
|
6 Months
|
|
|
Percentage of Eyes With Induced Manifest Refractive Astigmatism Greater Than 2.00 D of Absolute Cylinder as Compared to the Preoperative Refraction
大体时间:6 Months
|
Induced Manifest Refractive Astigmatism is an increase of astigmatism (cylinder) postoperatively that could be caused by the refractive treatment.
An increase of greater than 2.0 D is considered a safety endpoint per ANSI Z80.11-2007.
|
6 Months
|
合作者和调查者
在这里您可以找到参与这项研究的人员和组织。
研究记录日期
这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。
研究主要日期
学习开始
2010年10月1日
初级完成 (实际的)
2011年6月1日
研究完成 (实际的)
2011年12月1日
研究注册日期
首次提交
2010年9月24日
首先提交符合 QC 标准的
2010年10月12日
首次发布 (估计)
2010年10月14日
研究记录更新
最后更新发布 (估计)
2013年5月20日
上次提交的符合 QC 标准的更新
2013年5月16日
最后验证
2013年5月1日
更多信息