Standardized Procedure for the Assessment of New-to-market Continuous Glucose Monitoring Systems (SPACE2)

2013年12月13日 更新者:J.H. DeVries、Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)
The purpose of this study is to assess the accuracy of two continuous glucose monitoring devices (the Dexcom G4 Platinum and Medtronic Enlite systems) in patients with type 1 diabetes mellitus.

研究概览

地位

完全的

条件

详细说明

The study will use two CE marked Continuous Glucose Monitoring (CGM) systems; Dexcom G4 Platinum CGM CE-marked 2012 (Dexcom, USA) and the Medtronic Paradigm Veo system with Enlite Sensor CE-marked 2011 (Medtronic, USA). Both the Dexcom G4 Platinum and the Medtronic Enlite sensors will be obtained through normal commercial channels rather than directly from the manufacturer. The two CGM sensors will be worn concomitantly by the participant during the visit to the Clinical Research Center (CRC). This visit will have a duration of 6 hours. Blood will be drawn for the determination of glucose levels. The patient will receive his usual breakfast and an increased insulin bolus (180% of the patient's calculated mealtime dose) will be administered to correct the breakfast glucose excursion with the aim of inducing a period of minor hypoglycaemia. Blood sampling will continue until the end of the admission. At the end of this CRC part, the patient will continue to wear the two sensors at home. In the case of sensor failure before the CRC session on the third day into the study, patients will be instructed to insert a new sensor per the manufacturer's instruction for use and to notify study coordinators of the event. Sensors will be worn until the end of the six day study duration. Patients will be asked to perform at least 6 fingersticks per day for blood glucose measurements with the study glucometer. The study will end on the 6th day after initial sensor insertion. Patients will return to the CRC to have the sensor removed and their CGM sensor and blood glucometer data downloaded from the devices. In case of failure of both sensors after the CRC session but before the scheduled six day study duration, patients will return to the CRC for sensor removal and for data download.

研究类型

介入性

注册 (实际的)

24

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

    • Noord-Holland
      • Amsterdam、Noord-Holland、荷兰、1105AZ
        • Academic Medical Center

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 及以上 (成人、年长者)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria:

  • aged 18 years or above
  • diagnosed with Type 1 diabetes mellitus at least 6 months according to the WHO definition
  • Body Mass Index (BMI) <35 kg/m²
  • willing and able to wear a CGM device for the duration of the study and undergo all study procedures
  • HbA1c <10%
  • signed informed consent form prior to study entry

Exclusion Criteria:

  • Patient is pregnant, or breast feeding during the period of the study.
  • Patient is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
  • Patient may not use acetaminophen (paracetamol) while participating in the study
  • Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the patient's safety or successful participation in the study.
  • Patient is actively enrolled in another clinical trial or took part in a study within 30 days
  • Known adrenal gland problem, pancreatic tumour, or insulinoma
  • Inability of the patient to comply with all study procedures
  • Inability of the patient to understand the patient information.
  • Patient donated blood in the last 3 months

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:治疗
  • 分配:不适用
  • 介入模型:单组作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
实验性的:CGM Monitoring
Fitting of Dexcom G4 Platinum CGM monitor and Medtronic Enlite CGM monitor
All patients will be fitted with the two study CGM systems. Paradigm Veo will be configured to use the CGM part only if another pump is used.
其他名称:
  • Dexcom G4 Platinum and Medtronic Enlite CGM

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Mean Absolute Relative Difference (MARD)
大体时间:up to day 6 of use
MARD will be assessed as an average of the first 6 days of wear
up to day 6 of use

次要结果测量

结果测量
措施说明
大体时间
Accuracy of sensors per glycemic range and trial phase
大体时间:up to day 6 of use
Accuracy of the sensor in the hypoglycaemic (defined as a blood glucose value ≤3.9 mmol/L) and hyperglycaemic area (defined as a blood glucose value ≥10.0 mmol/L) as an average of the first six days of use. Additionally, a separate analysis will be performed to assess sensor performance and accuracy per day of sensor life.
up to day 6 of use

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:J. Hans DeVries, MD, PhD、Academic Medical Center - Department of Internal Medicine, Amsterdam, The Netherlands
  • 首席研究员:Eric Renard, PhD、Medical University Montpellier, France
  • 首席研究员:Angelo Avogaro, PhD、Medical University Padova, Italy
  • 研究主任:Julia Mader, MD、Medical University Graz, Austria
  • 首席研究员:Thomas Pieber, MD、Medical University Graz, Austria

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始

2013年1月1日

初级完成 (实际的)

2013年9月1日

研究完成 (实际的)

2013年9月1日

研究注册日期

首次提交

2012年12月13日

首先提交符合 QC 标准的

2012年12月14日

首次发布 (估计)

2012年12月18日

研究记录更新

最后更新发布 (估计)

2013年12月16日

上次提交的符合 QC 标准的更新

2013年12月13日

最后验证

2013年12月1日

更多信息

与本研究相关的术语

其他研究编号

  • SPACE2

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅