Standardized Procedure for the Assessment of New-to-market Continuous Glucose Monitoring Systems (SPACE2)

2013年12月13日 更新者:J.H. DeVries、Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)
The purpose of this study is to assess the accuracy of two continuous glucose monitoring devices (the Dexcom G4 Platinum and Medtronic Enlite systems) in patients with type 1 diabetes mellitus.

調査の概要

状態

完了

条件

詳細な説明

The study will use two CE marked Continuous Glucose Monitoring (CGM) systems; Dexcom G4 Platinum CGM CE-marked 2012 (Dexcom, USA) and the Medtronic Paradigm Veo system with Enlite Sensor CE-marked 2011 (Medtronic, USA). Both the Dexcom G4 Platinum and the Medtronic Enlite sensors will be obtained through normal commercial channels rather than directly from the manufacturer. The two CGM sensors will be worn concomitantly by the participant during the visit to the Clinical Research Center (CRC). This visit will have a duration of 6 hours. Blood will be drawn for the determination of glucose levels. The patient will receive his usual breakfast and an increased insulin bolus (180% of the patient's calculated mealtime dose) will be administered to correct the breakfast glucose excursion with the aim of inducing a period of minor hypoglycaemia. Blood sampling will continue until the end of the admission. At the end of this CRC part, the patient will continue to wear the two sensors at home. In the case of sensor failure before the CRC session on the third day into the study, patients will be instructed to insert a new sensor per the manufacturer's instruction for use and to notify study coordinators of the event. Sensors will be worn until the end of the six day study duration. Patients will be asked to perform at least 6 fingersticks per day for blood glucose measurements with the study glucometer. The study will end on the 6th day after initial sensor insertion. Patients will return to the CRC to have the sensor removed and their CGM sensor and blood glucometer data downloaded from the devices. In case of failure of both sensors after the CRC session but before the scheduled six day study duration, patients will return to the CRC for sensor removal and for data download.

研究の種類

介入

入学 (実際)

24

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • Noord-Holland
      • Amsterdam、Noord-Holland、オランダ、1105AZ
        • Academic Medical Center

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • aged 18 years or above
  • diagnosed with Type 1 diabetes mellitus at least 6 months according to the WHO definition
  • Body Mass Index (BMI) <35 kg/m²
  • willing and able to wear a CGM device for the duration of the study and undergo all study procedures
  • HbA1c <10%
  • signed informed consent form prior to study entry

Exclusion Criteria:

  • Patient is pregnant, or breast feeding during the period of the study.
  • Patient is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
  • Patient may not use acetaminophen (paracetamol) while participating in the study
  • Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the patient's safety or successful participation in the study.
  • Patient is actively enrolled in another clinical trial or took part in a study within 30 days
  • Known adrenal gland problem, pancreatic tumour, or insulinoma
  • Inability of the patient to comply with all study procedures
  • Inability of the patient to understand the patient information.
  • Patient donated blood in the last 3 months

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:処理
  • 割り当て:なし
  • 介入モデル:単一グループの割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:CGM Monitoring
Fitting of Dexcom G4 Platinum CGM monitor and Medtronic Enlite CGM monitor
All patients will be fitted with the two study CGM systems. Paradigm Veo will be configured to use the CGM part only if another pump is used.
他の名前:
  • Dexcom G4 Platinum and Medtronic Enlite CGM

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Mean Absolute Relative Difference (MARD)
時間枠:up to day 6 of use
MARD will be assessed as an average of the first 6 days of wear
up to day 6 of use

二次結果の測定

結果測定
メジャーの説明
時間枠
Accuracy of sensors per glycemic range and trial phase
時間枠:up to day 6 of use
Accuracy of the sensor in the hypoglycaemic (defined as a blood glucose value ≤3.9 mmol/L) and hyperglycaemic area (defined as a blood glucose value ≥10.0 mmol/L) as an average of the first six days of use. Additionally, a separate analysis will be performed to assess sensor performance and accuracy per day of sensor life.
up to day 6 of use

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:J. Hans DeVries, MD, PhD、Academic Medical Center - Department of Internal Medicine, Amsterdam, The Netherlands
  • 主任研究者:Eric Renard, PhD、Medical University Montpellier, France
  • 主任研究者:Angelo Avogaro, PhD、Medical University Padova, Italy
  • スタディディレクター:Julia Mader, MD、Medical University Graz, Austria
  • 主任研究者:Thomas Pieber, MD、Medical University Graz, Austria

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始

2013年1月1日

一次修了 (実際)

2013年9月1日

研究の完了 (実際)

2013年9月1日

試験登録日

最初に提出

2012年12月13日

QC基準を満たした最初の提出物

2012年12月14日

最初の投稿 (見積もり)

2012年12月18日

学習記録の更新

投稿された最後の更新 (見積もり)

2013年12月16日

QC基準を満たした最後の更新が送信されました

2013年12月13日

最終確認日

2013年12月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • SPACE2

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

CGM Monitoringの臨床試験

3
購読する