Reduced wear in vitamin E-infused highly cross-linked polyethylene cups: 5-year results of a randomized controlled trial

Goulven Rochcongar, Matthieu Remazeilles, Emeline Bourroux, Julien Dunet, Valentin Chapus, Matthieu Feron, César Praz, Geoffrey Buia, Christophe Hulet, Goulven Rochcongar, Matthieu Remazeilles, Emeline Bourroux, Julien Dunet, Valentin Chapus, Matthieu Feron, César Praz, Geoffrey Buia, Christophe Hulet

Abstract

Background and purpose - Vitamin E-infused polyethylene is a relatively new material in joint arthroplasty; there are no long-term reports, and only few mid-term results. Using radiostereometric analysis (RSA), we primarily determined whether vitamin E-infused highly cross-linked polyethylene (HXLPE/VitE) acetabular cups show less wear than ultra-high molecular weight polyethylene (UHMWPE) acetabular cups at 5 years after total hip arthroplasty (THA). We also assessed whether wear rates correlate with increasing cup inclination angles or cup sizes.Patients and methods - This is a 5-year follow-up of our previously reported randomized controlled trial of 62 patients with 3 years' follow-up, who received THA with either an HXLPE/VitE or a UHMWPE acetabular cup. At 5 years, 40 patients were analyzed (22 in the HXLPE/VitE and 18 in the UHMWPE group).Results - HXLPE/VitE cups continued to show less cumulative femoral head penetration than UHMWPE cups (HXLPE/VitE: 0.24 mm, UHMWPE: 0.45 mm; p < 0.001). Distribution of wear was also more even with HXLPE/VitE cups than with UHMWPE cups (p = 0.002). Moreover, the difference in PE wear between 1 and 5 years in both groups showed no statistically significant correlation with increasing cup inclination angles or cup sizes. Finally, no osteolysis and implant loosening occurred, and no revision surgeries were required.Interpretation - Wear rates continue to be lower in HXLPE/VitE cups than in UHMWPE cups at 5 years of follow-up without correlation with increasing cup inclination angles or cup sizes. Finally, HXLPE/VitE cups may have the potential to prevent osteolysis and implant loosening.

Trial registration: ClinicalTrials.gov NCT02524587.

Figures

Figure 1.
Figure 1.
Study enrollment. The reasons for the losses to follow-up—if known—are indicated in the corresponding boxes. HXLPE/VitE: Vitamin E-infused highly cross-linked polyethylene, RSA: radiostereometric analysis, UHMWPE: ultra-high molecular weight polyethylene.
Figure 2.
Figure 2.
Creep and wear behavior of UHMWPE (n = 18, red line) and HXLPE/VitE (n = 22, blue line) over the first 5 years after implantation. HXLPE/VitE: Vitamin E-infused highly cross-linked polyethylene, UHMWPE: ultra-high molecular weight polyethylene.

References

    1. Berend K R, Adams J B, Morris M J, Lombardi A V, Jr. Early experience with a new porous hemispheric acetabular component. Surg Technol Int 2015; 27: 263–7.
    1. Bitar D, Parvizi J.. Biological response to prosthetic debris. World J Orthop 2015; 6(2): 172–89.
    1. Bono J V, Sanford L, Toussaint J T.. Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J Arthroplasty 1994; 9(2): 119–25.
    1. Busch A, Jager M, Wegner A, Haversath M.. Vitamin E-blended versus conventional polyethylene liners in prostheses: prospective, randomized trial with 3-year follow-up. Orthopade 2019; Nov 6.
    1. Callary S A, Field J R, Campbell D G.. Low wear of a second-generation highly crosslinked polyethylene liner: a 5-year radiostereometric analysis study. Clin Orthop Relat Res 2013; 471(11): 3596–600.
    1. Callary S A, Solomon L B, Holubowycz O T, Campbell D G, Munn Z, Howie DW.. Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 2015; 86(2): 159–68.
    1. Dowd J E, Sychterz C J, Young A M, Engh C A.. Characterization of long-term femoral-head-penetration rates: association with and prediction of osteolysis. J Bone Joint Surg Am 2000; 82-A(8): 1102–7.
    1. Dumbleton J H, Manley M T, Edidin A A.. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 2002; 17(5): 649–61.
    1. Elke R, Rieker C B.. Estimating the osteolysis-free life of a total hip prosthesis depending on the linear wear rate and head size. Proc Inst Mech Eng H 2018; 232(8): 753–8.
    1. Galea V P, Rojanasopondist P, Laursen M, Muratoglu O K, Malchau H, Bragdon C.. Evaluation of vitamin E-diffused highly crosslinked polyethylene wear and porous titanium-coated shell stability: a seven-year randomized control trial using radiostereometric analysis. Bone Joint J 2019; 101-b(7): 760–7.
    1. Garling E H, Kaptein B L, Geleijns K, Nelissen R G, Valstar E R.. Marker configuration model-based roentgen fluoroscopic analysis. J Biomech 2005; 38(4): 893–901.
    1. Halma J J, Senaris J, Delfosse D, Lerf R, Oberbach T, van Gaalen S M, de Gast A.. Edge loading does not increase wear rates of ceramic-on-ceramic and metal-on-polyethylene articulations. J. Biomed. Mater. Res. Part B Appl. Biomater 2014; 102(8): 1627–38.
    1. Lindalen E, Nordsletten L, Hovik O, Rohrl S M.. E-vitamin infused highly cross-linked polyethylene: RSA results from a randomised controlled trial using 32 mm and 36 mm ceramic heads. Hip Int 2015; 25(1): 50–5.
    1. Little N J, Busch C A, Gallagher J A, Rorabeck C H, Bourne R B.. Acetabular polyethylene wear and acetabular inclination and femoral offset. Clin Orthop Relat Res 2009; 467(11): 2895–900.
    1. Nebergall A K, Troelsen A, Rubash H E, Malchau H, Rolfson O, Greene M E.. Five-year experience of Vitamin E-diffused highly cross-linked polyethylene wear in total hip arthroplasty assessed by radiostereometric analysis. J. Arthroplasty 2016; 31(6): 1251–5.
    1. Nebergall A K, Greene M E, Laursen M B, Nielsen P T, Malchau H, Troelsen A.. Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years: a randomised controlled trial using radiostereometric analysis. Bone Joint J 2017; 99-b(5): 577–84.
    1. Önsten I, Carlsson A S, Besjakov J.. Wear in uncemented porous and cemented polyethylene sockets: a randomised, radiostereometric study. Br J Surg 1998; 80(2): 345–50.
    1. Oral E, Greenbaum E S, Malhi A S, Harris W H, Muratoglu O K.. Characterization of irradiated blends of alpha-tocopherol and UHMWPE. Biomaterials 2005; 26(33): 6657–63.
    1. Oral E, Wannomae K K, Rowell S L, Muratoglu O K.. Diffusion of vitamin E in ultra-high molecular weight polyethylene. Biomaterials 2007; 28(35): 5225–37.
    1. Pineau V, Lebel B, Gouzy S, Dutheil J J, Vielpeau C.. Dual mobility hip arthroplasty wear measurement: experimental accuracy assessment using radiostereometric analysis (RSA). Orthop Traumatol Surg Res 2010; 96(6): 609–15.
    1. Rochcongar G, Buia G, Bourroux E, Dunet J, Chapus V, Hulet C.. Creep and wear in Vitamin E-infused highly cross-linked polyethylene cups for total hip arthroplasty: a prospective randomized controlled trial. J Bone Joint Surg Am 2018; 100(2): 107–14.
    1. Salemyr M, Muren O, Ahl T, Bodén H, Chammout G, Stark A, Sköldenberg O.. Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty: a randomized, controlled trial. Int Orthop 2015; 39(8): 1499–505.
    1. Scemama C, Anract P, Dumaine V, Babinet A, Courpied J P, Hamadouche M.. Does vitamin E-blended polyethylene reduce wear in primary total hip arthroplasty: a blinded randomised clinical trial. Int Orthop 2017; 41(6): 1113–8.
    1. Shareghi B, Johanson P E, Karrholm J.. Femoral head penetration of Vitamin E-infused highly cross-linked polyethylene liners: a randomized radiostereometric study of seventy hips followed for two years. J Bone Joint Surg Am 2015; 97(16): 1366–71.
    1. Shareghi B, Johanson P E, Karrholm J.. Wear of Vitamin E-infused highly cross-linked polyethylene at five years. J Bone Joint Surg Am 2017; 99(17): 1447–52.
    1. Sillesen N H, Greene M E, Nebergall A K, Nielsen P T, Laursen M B, Troelsen A, Malchau H.. Three year RSA evaluation of Vitamin E diffused highly cross-linked polyethylene liners and cup stability. J Arthroplasty 2015; 30(7): 1260–4.
    1. Sillesen N H, Greene M E, Nebergall A K, Huddleston J I, Emerson R, Gebuhr P, Troelsen A, Malchau H.. 3-year follow-up of a long-term registry-based multicentre study on vitamin E diffused polyethylene in total hip replacement. Hip Int 2016; 26(1): 97–103.
    1. Teeter M G, Lanting B A, Naudie D D, McCalden R W, Howard J L, MacDonald S J.. Highly crosslinked polyethylene wear rates and acetabular component orientation: a minimum ten-year follow-up. Bone Joint J 2018; 100-b(7): 891–7.
    1. Tian J L, Sun L, Hu R Y, Han W, Tian X B.. Correlation of cup inclination angle with liner wear for metal-on-polyethylene in hip primary arthroplasty. Orthop Surg 2017; 9(2): 186–90.
    1. Tian J, Sun L, Hu R, Han W, Tian X.. Long-term results of primary hip arthroplasty with cup inclination angle bigger than fifty degrees. J Clin Orthop Trauma 2018; 9(2): 133–6.
    1. Wyatt M, Weidner J, Pfluger D, Beck M.. The RM Pressfit vitamys: 5-year Swiss experience of the first 100 cups. Hip Int 2017; 27(4): 368–72.

Source: PubMed

3
订阅