Telemedicine and cardiac implants: what is the benefit?

Niraj Varma, Renato Pietro Ricci, Niraj Varma, Renato Pietro Ricci

Abstract

Cardiac implantable electronic devices are increasing in prevalence. The post-implant follow-up is important for monitoring both device function and patient condition. However, practice is inconsistent. For example, ICD follow-up schedules vary from 3 monthly to yearly according to facility and physician preference and availability of resources. Recommended follow-up schedules impose significant burden. Importantly, no surveillance occurs between follow-up visits. In contrast, implantable devices with automatic remote monitoring capability provide a means for performing constant surveillance, with the ability to identify salient problems rapidly. Remote home monitoring reduces the volume of device clinic visits and provides early detection of patient and/or system problems.

Keywords: Atrial fibrillation; Cardiac implantable electronic devices; Early detection; Follow-up; ICD; Recall management; Remote monitoring.

Figures

Figure 1
Figure 1
Home Monitoring technology. Top: transmission steps in this fully automatic system (modified with permission from ref.). Both landline and portable GSM/landline patient modules are illustrated. Bottom Left >90% of transmissions were received in

Figure 2

The TRUST trial. Top left:…

Figure 2

The TRUST trial. Top left: continuous remote monitoring compared with standard of care.…

Figure 2
The TRUST trial. Top left: continuous remote monitoring compared with standard of care. Right: Home Monitoring reduced cardiac-resource utilization [scheduled and unscheduled clinic and hospital visits (including responses to Home Monitoring event notifications)] by 45% in 1 year. Bottom: time to physician evaluation of arrhythmias (left), and for silent events (right). Compiled with permission from ref.

Figure 3

Home Monitoring generator coupled to…

Figure 3

Home Monitoring generator coupled to a Fidelis (MDT 6949) lead. Two separate event…

Figure 3
Home Monitoring generator coupled to a Fidelis (MDT 6949) lead. Two separate event notifications that were transmitted immediately on occurrence of lead fracture, occurring silently during sleep at 4:43 am, 6 weeks after last clinic follow-up in November 14. Left: lead impedance suddenly increased; right: ventricular fibrillation detection due to irregular sensed events. Electrogram definition in current generation has improved resolution (1/128 s) and includes post-detection sequences (e.g. Figure 2). The patient was reviewed within hours (FU) of the notifications. Compiled with permission from ref.

Figure 4

Interrogation of a CRT-D device…

Figure 4

Interrogation of a CRT-D device in patient presenting with acute decompensated heart failure.…

Figure 4
Interrogation of a CRT-D device in patient presenting with acute decompensated heart failure. Compiled with permission from ref.

Figure 5

Home guide registry.

Figure 5

Home guide registry.

Figure 5
Home guide registry.
Figure 2
Figure 2
The TRUST trial. Top left: continuous remote monitoring compared with standard of care. Right: Home Monitoring reduced cardiac-resource utilization [scheduled and unscheduled clinic and hospital visits (including responses to Home Monitoring event notifications)] by 45% in 1 year. Bottom: time to physician evaluation of arrhythmias (left), and for silent events (right). Compiled with permission from ref.
Figure 3
Figure 3
Home Monitoring generator coupled to a Fidelis (MDT 6949) lead. Two separate event notifications that were transmitted immediately on occurrence of lead fracture, occurring silently during sleep at 4:43 am, 6 weeks after last clinic follow-up in November 14. Left: lead impedance suddenly increased; right: ventricular fibrillation detection due to irregular sensed events. Electrogram definition in current generation has improved resolution (1/128 s) and includes post-detection sequences (e.g. Figure 2). The patient was reviewed within hours (FU) of the notifications. Compiled with permission from ref.
Figure 4
Figure 4
Interrogation of a CRT-D device in patient presenting with acute decompensated heart failure. Compiled with permission from ref.
Figure 5
Figure 5
Home guide registry.

References

    1. Dubner S, Auricchio A, Steinberg JS, Vardas P, Stone P, Brugada J, Piotrowicz R, Hayes DL, Kirchhof P, Breithardt G, Zareba W, Schuger C, Aktas MK, Chudzik M, Mittal S, Varma N. ISHNE/EHRA expert consensus on remote monitoring of cardiovascular implantable electronic devices (CIEDs) Europace. 2012;14:278–293. doi: 10.1093/europace/eur303.
    1. Wilkoff BL, Auricchio A, Brugada J, Cowie M, Ellenbogen KA, Gillis AM, Hayes DL, Howlett JG, Kautzner J, Love CJ, Morgan JM, Priori SG, Reynolds DW, Schoenfeld MH, Vardas PE. HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations: developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the European Heart Rhythm Association (a registered branch of the ESC), the American College of Cardiology, the American Heart Association. Europace. 2008;10:707–725. doi: 10.1093/europace/eun122.
    1. Carlson MD, Wilkoff BL, Maisel WH, Carlson MD, Ellenbogen KA, Saxon LA, Prystowsky EN, Alpert JS, Cain ME, Ching EA, Curtis AB, Davies DW, Hammill SC, Hauser RG, Lampert R, Zipes DP. Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) and the International Coalition of Pacing and Electrophysiology Organizations (COPE) Heart Rhythm. 2006;3:1250–1273. doi: 10.1016/j.hrthm.2006.08.029.
    1. Maisel WH, Hauser RG, Hammill SC, Hauser RG, Ellenbogen KA, Epstein AE, Hayes DL, Alpert JS, Berger RD, Curtis AB, Dubin AM, Estes NA, 3rd, Gura MT, Krahn AD, Lampert R, Lindsay BD, Wilkoff BL. Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and Guidelines: developed in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Heart Rhythm. 2009;6:869–885. doi: 10.1016/j.hrthm.2009.04.024.
    1. Schoenfeld MH, Compton SJ, Mead RH, Weiss DN, Sherfesee L, Englund J, Mongeon LR. Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. Pacing Clin Electrophysiol. 2004;27:757–763. doi: 10.1111/j.1540-8159.2004.00524.x.
    1. Joseph GK, Wilkoff BL, Dresing T, Burkhardt J, Khaykin Y. Remote interrogation and monitoring of implantable cardioverter defibrillators. J Interv Card Electrophysiol. 2004;11:161–166. doi: 10.1023/B:JICE.0000042356.52369.89.
    1. Crossley GH, Chen J, Choucair W, Cohen TJ, Gohn DC, Johnson WB, Kennedy EE, Mongeon LR, Serwer GA, Qiao H, Wilkoff BL. Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. J Am Coll Cardiol. 2009;54:2012–2019. doi: 10.1016/j.jacc.2009.10.001.
    1. Al-Khatib SM, Piccini JP, Knight D, Stewart M, Clapp-Channing N, Sanders GD. Remote monitoring of implantable cardioverter defibrillators versus quarterly device interrogations in clinic: results from a randomized pilot clinical trial. J Cardiovasc Electrophysiol. 2010;21:545–550. doi: 10.1111/j.1540-8167.2009.01659.x.
    1. Cronin EM, Ching EA, Varma N, Martin DO, Wilkoff BL, Lindsay BD. Remote monitoring of cardiovascular devices - a time and activity analysis. Heart Rhythm. Advance Access published August 3, 2012, doi: 10.1016/j.hrthm.2012.08.002.
    1. Varma N, Stambler B, Chun S. Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing Clin Electrophysiol. 2005;28(Suppl. 1):S133–S136. doi: 10.1111/j.1540-8159.2005.00083.x.
    1. Varma N. Rationale and design of a prospective study of the efficacy of a remote monitoring system used in implantable cardioverter defibrillator follow-up: the Lumos-T Reduces Routine Office Device Follow-Up Study (TRUST) study. Am Heart J. 2007;154:1029–1034. doi: 10.1016/j.ahj.2007.07.051.
    1. Varma N, Epstein A, Irimpen A, Schweikert R, Shah J, Love CJ, Investigators T. Efficacy and safety of automatic remote monitoring for ICD Follow-Up: the TRUST trial. Circulation. 2010;122:325–332. doi: 10.1161/CIRCULATIONAHA.110.937409.
    1. Crossley G, Boyle A, Vitense H, Chang Y, Mead RH. The clinical evaluation of remote notification to reduce time to clinical decision (CONNECT) trial: the value of wireless remote monitoring with automatic clinician alerts. J Am Coll Cardiol. 2011;57:1181–1189. doi: 10.1016/j.jacc.2010.12.012.
    1. Heidbuchel H, Lioen P, Foulon S, Huybrechts W, Ector J, Willems R, Ector H. Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace. 2008;10:351–357. doi: 10.1093/europace/eun010.
    1. Kacet S. Safety and Effectiveness of ICD Follow-up using Remote Monitoring: ECOST Study. Presented in Hot Line Session at ESC 2011 Congress; 2011. .
    1. Varma N, Michalski J TRUST Investigators. Effect of transmission reliability on remote follow-up in ICD patients: Automatic Home Monitoring in the TRUST Trial. Heart Rhythm. 2012 (Abstract )
    1. Perings C, Bauer WR, Bondke HJ, Mewis C, James M, Bocker D, Broadhurst P, Korte T, Toft E, Hintringer F, Clementy J, Schwab JO. Remote monitoring of implantable-cardioverter defibrillators: results from the Reliability of IEGM Online Interpretation (RIONI) study. Europace. 2011;13:221–229. doi: 10.1093/europace/euq447.
    1. Mabo P, Victor F, Bazin P, Ahres S, Babuty D, Da Costa A, Binet D, Daubert JC. A randomized trial of long-term remote monitoring of pacemaker recipients (The COMPAS trial) Eur Heart J. 2012;33:1105–1011. doi: 10.1093/eurheartj/ehr419.
    1. Wilkoff BL, Auricchio A, Brugada J, Cowie M, Ellenbogen KA, Gillis AM, Hayes DL, Howlett JG, Kautzner J, Love CJ, Morgan JM, Priori SG, Reynolds DW, Schoenfeld MH, Vardas PE. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm. 2008;5:907–925. doi: 10.1016/j.hrthm.2008.04.013.
    1. Brugada P. What evidence do we have to replace in-hospital implantable cardioverter defibrillator follow-up? Clin Res Cardiol. 2006;95(Suppl. 3):III3–III9. doi: 10.1007/s00392-006-1302-x.
    1. Senges-Becker JC, Klostermann M, Becker R, Bauer A, Siegler KE, Katus HA, Schoels W. What is the ‘optimal’ follow-up schedule for ICD patients? Europace. 2005;7:319–326. doi: 10.1016/j.eupc.2005.02.117.
    1. Lee DS, Krahn AD, Healey JS, Birnie D, Crystal E, Dorian P, Simpson CS, Khaykin Y, Cameron D, Janmohamed A, Yee R, Austin PC, Chen Z, Hardy J, Tu JV. Evaluation of early complications related to De Novo cardioverter defibrillator implantation insights from the Ontario ICD database. J Am Coll Cardiol. 2010;55:774–782. doi: 10.1016/j.jacc.2009.11.029.
    1. FDA Home Monitoring. 2009. . P050023/S020.
    1. Zartner P, Handke R, Photiadis J, Brecher AM, Schneider MB. Performance of an autonomous telemonitoring system in children and young adults with congenital heart diseases. Pacing Clin Electrophysiol. 2008;31:1291–1299. doi: 10.1111/j.1540-8159.2008.01180.x.
    1. Medtronic. Clinician Manual Supplement Protecta™ XT/Protecta.™ Projected service life information related to remote monitoring. 22 Oct 2010. Manual no.: M945739A001A.
    1. 2007. Medtronic. Physician Advisory Letter: Urgent Medical Device Information Sprint Fidelis© Lead Patient Management Recommendations. .
    1. Kallinen LM, Hauser RG, Lee KW, Almquist AK, Katsiyiannis WT, Tang CY, Melby DP, Gornick CC. Failure of impedance monitoring to prevent adverse clinical events caused by fracture of a recalled high-voltage implantable cardioverter-defibrillator lead. Heart Rhythm. 2008;5:775–779. doi: 10.1016/j.hrthm.2008.02.039.
    1. Neuzil P, Taborsky M, Holy F, Wallbrueck K. Early automatic remote detection of combined lead insulation defect and ICD damage. Europace. 2008;10:556–557. doi: 10.1093/europace/eun009.
    1. Varma N. Remote monitoring for advisories: automatic early detection of silent lead failure. Pacing Clin Electrophysiol. 2009;32:525–527. doi: 10.1111/j.1540-8159.2009.02314.x.
    1. Spencker S, Coban N, Koch L, Schirdewan A, Muller D. Potential role of home monitoring to reduce inappropriate shocks in implantable cardioverter-defibrillator patients due to lead failure. Europace. 2009;11:483–488. doi: 10.1093/europace/eun350.
    1. Varma N, Michalski J, Epstein AE, Schweikert R. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ Arrhythm Electrophysiol. 2010;3:428–436. doi: 10.1161/CIRCEP.110.951962.
    1. Varma N, Johnson MA. Prevalence of cancelled shock therapy and relationship to shock delivery in recipients of implantable cardioverter-defibrillators assessed by remote monitoring. Pacing Clin Electrophysiol. 2009;32(Suppl. 1):S42–S46. doi: 10.1111/j.1540-8159.2008.02288.x.
    1. Orlov MV, Ghali JK, Araghi-Niknam M, Sherfesee L, Sahr D, Hettrick DA. Asymptomatic atrial fibrillation in pacemaker recipients: incidence, progression, and determinants based on the atrial high rate trial. Pacing Clin Electrophysiol. 2007;30:404–411. doi: 10.1111/j.1540-8159.2007.00682.x.
    1. Glotzer TV, Hellkamp AS, Zimmerman J, Sweeney MO, Yee R, Marinchak R, Cook J, Paraschos A, Love J, Radoslovich G, Lee KL, Lamas GA. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the Atrial Diagnostics Ancillary Study of the MOde Selection Trial (MOST) Circulation. 2003;107:1614–1619. doi: 10.1161/01.CIR.0000057981.70380.45.
    1. Capucci A, Santini M, Padeletti L, Gulizia M, Botto G, Boriani G, Ricci R, Favale S, Zolezzi F, Di Belardino N, Molon G, Drago F, Villani GQ, Mazzini E, Vimercati M, Grammatico A. Monitored atrial fibrillation duration predicts arterial embolic events in patients suffering from bradycardia and atrial fibrillation implanted with antitachycardia pacemakers. J Am Coll Cardiol. 2005;46:1913–1920. doi: 10.1016/j.jacc.2005.07.044.
    1. Glotzer TV, Daoud EG, Wyse DG, Singer DE, Ezekowitz MD, Hilker C, Miller C, Qi D, Ziegler PD. The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study. Circ Arrhythm Electrophysiol. 2009;2:474–480. doi: 10.1161/CIRCEP.109.849638.
    1. Ricci RP, Morichelli L, Santini M. Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace. 2009;11:54–61. doi: 10.1093/europace/eun303.
    1. Ricci RP, Morichelli L, Gargaro A, Laudadio MT, Santini M. Home monitoring in patients with implantable cardiac devices: is there a potential reduction of stroke risk? Results from a computer model tested through Monte Carlo simulations. J Cardiovasc Electrophysiol. 2009;20:1244–1251. doi: 10.1111/j.1540-8167.2009.01543.x.
    1. Santini M, Gasparini M, Landolina M, Lunati M, Proclemer A, Padeletti L, Catanzariti D, Molon G, Botto GL, La Rocca L, Grammatico A, Boriani G. Device-detected atrial tachyarrhythmias predict adverse outcome in real-world patients with implantable biventricular defibrillators. J Am Coll Cardiol. 2011;57:167–172. doi: 10.1016/j.jacc.2010.08.624.
    1. Shanmugam N, Boerdlein A, Proff J, Ong P, Valencia O, Maier SK, Bauer WR, Paul V, Sack S. Detection of atrial high-rate events by continuous home monitoring: clinical significance in the heart failure-cardiac resynchronization therapy population. Europace. 2012;14:230–237. doi: 10.1093/europace/eur293.
    1. Ip J, Waldo AL, Lip GY, Rothwell PM, Martin DT, Bersohn MM, Choucair WK, Akar JG, Wathen MS, Rohani P, Halperin JL. Multicenter randomized study of anticoagulation guided by remote rhythm monitoring in patients with implantable cardioverter-defibrillator and CRT-D devices: rationale, design, and clinical characteristics of the initially enrolled cohort The IMPACT study. Am Heart J. 2009;158:364–370. doi: 10.1016/j.ahj.2009.07.002. e1.
    1. Santini M, Ricci RP, Lunati M, Landolina M, Perego GB, Marzegalli M, Schirru M, Belvito C, Brambilla R, Guenzati G, Gilardi S, Valsecchi S. Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes. J Interv Card Electrophysiol. 2009;24:53–61. doi: 10.1007/s10840-008-9321-3.
    1. Mabo P, Defaye P, Sadoul N, Davy J, Deharo J, Kacet S. Remote follow-up of patients implanted with an ICD. The prospective randomized EVATEL study. 2011. .
    1. Varma N, Wilkoff B. Device features for managing patients with heart failure. Heart Fail Clin. 2011;7:215–225. doi: 10.1016/j.hfc.2010.12.005. viii.
    1. Koplan BA, Kaplan AJ, Weiner S, Jones PW, Seth M, Christman SA. Heart failure decompensation and all-cause mortality in relation to percent biventricular pacing in patients with heart failure: is a goal of 100% biventricular pacing necessary? J Am Coll Cardiol. 2009;53:355–360. doi: 10.1016/j.jacc.2008.09.043.
    1. Arya A, Block M, Kautzner J, Lewalter T, Mortel H, Sack S, Schumacher B, Sogaard P, Taborsky M, Husser D, Hindricks G. Influence of Home Monitoring on the clinical status of heart failure patients: Design and rationale of the IN-TIME study. Eur J Heart Fail. 2008;10:1143–1148. doi: 10.1016/j.ejheart.2008.08.004.
    1. Burri H, Quesada A, Ricci RP, Boriani G, Davinelli M, Favale S, Da Costa A, Kautzner J, Moser R, Navarro X, Santini M. The MOnitoring Resynchronization dEvices and CARdiac patiEnts (MORE-CARE) study: rationale and design. Am Heart J. 2010;160:42–48. doi: 10.1016/j.ahj.2010.04.005.
    1. Conraads VM, Tavazzi L, Santini M, Oliva F, Gerritse B, Yu CM, Cowie MR. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. Eur Heart J. 2011;32:2266–2273. doi: 10.1093/eurheartj/ehr050.
    1. van Veldhuisen DJ, Braunschweig F, Conraads V, Ford I, Cowie MR, Jondeau G, Kautzner J, Aguilera RM, Lunati M, Yu CM, Gerritse B, Borggrefe M. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011;124:1719–1726. doi: 10.1161/CIRCULATIONAHA.111.043042.
    1. Hoppe UC, Vanderheyden M, Sievert H, Brandt MC, Tobar R, Wijns W, Rozenman Y. Chronic monitoring of pulmonary artery pressure in patients with severe heart failure: multicentre experience of the monitoring Pulmonary Artery Pressure by Implantable device Responding to Ultrasonic Signal (PAPIRUS) II study. Heart. 2009;95:1091–1097. doi: 10.1136/hrt.2008.153486.
    1. Ritzema J, Melton IC, Richards AM, Crozier IG, Frampton C, Doughty RN, Whiting J, Kar S, Eigler N, Krum H, Abraham WT, Troughton RW. Direct left atrial pressure monitoring in ambulatory heart failure patients: initial experience with a new permanent implantable device. Circulation. 2007;116:2952–2959. doi: 10.1161/CIRCULATIONAHA.107.702191.
    1. Khoury DS, Naware M, Siou J, Blomqvist A, Mathuria NS, Wang J, Shih HT, Nagueh SF, Panescu D. Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors. J Am Coll Cardiol. 2009;53:1075–1081. doi: 10.1016/j.jacc.2008.12.018.
    1. Abraham WT, Adamson PB, Bourge RC, Aaron MF, Costanzo MR, Stevenson LW, Strickland W, Neelagaru S, Raval N, Krueger S, Weiner S, Shavelle D, Jeffries B, Yadav JS. Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet. 2011;377:658–666. doi: 10.1016/S0140-6736(11)60101-3.
    1. Whellan DJ, Ousdigian KT, Al-Khatib SM, Pu W, Sarkar S, Porter CB, Pavri BB, O'Connor CM. Combined heart failure device diagnostics identify patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF (Program to Access and Review Trending Information and Evaluate Correlation to Symptoms in Patients With Heart Failure) study. J Am Coll Cardiol. 2010;55:1803–1810. doi: 10.1016/j.jacc.2009.11.089.
    1. Mullens W, Grimm RA, Verga T, Dresing T, Starling RC, Wilkoff BL, Tang WH. Insights from a cardiac resynchronization optimization clinic as part of a heart failure disease management program. J Am Coll Cardiol. 2009;53:765–773. doi: 10.1016/j.jacc.2008.11.024.
    1. Ricci RP, Morichelli L, Santini M. Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: impact on medical management and health-care resource utilization. Europace. 2008;10:164–170. doi: 10.1093/europace/eum289.
    1. Ricci R, Calcagnini G, Castro A, Giada F, Igidbashan D, Landolina M, Melissano D, Perego G, Toselli T. Documento di consenso sul monitoraggio remoto dei dispositivi impiantabili: tecnologie disponibili, indicazioni, modelli organizzativi, accettabilità, responsabilità ed aspetti economici. G Ital Cardiol. 2011;12:450–467.
    1. Ricci R, Morichelli L, D'Onofrio A, Zanotto G, Vaccari D, Calò L, Buja G, Rovai N, Gargaro A. Home monitoring manpower, sensitivity and positive predictive value of adverse event detection. Preliminary results from the Home Guide registry. Eur Heart J. 2011;32(Suppl. 1):54.
    1. Marzegalli M, Lunati M, Landolina M, Perego GB, Ricci RP, Guenzati G, Schirru M, Belvito C, Brambilla R, Masella C, Di Stasi F, Valsecchi S, Santini M. Remote monitoring of CRT-ICD: the multicenter Italian CareLink evaluation–ease of use, acceptance, and organizational implications. Pacing Clin Electrophysiol. 2008;31:1259–1264. doi: 10.1111/j.1540-8159.2008.01175.x.
    1. Ricci RP, Morichelli L, Quarta L, Sassi A, Porfili A, Laudadio MT, Gargaro A, Santini M. Long-term patient acceptance of and satisfaction with implanted device remote monitoring. Europace. 2010;12:674–679. doi: 10.1093/europace/euq046.
    1. Petersen HH, Larsen MC, Nielsen OW, Kensing F, Svendsen JH. Patient satisfaction and suggestions for improvement of remote ICD monitoring. J Interv Card Electrophysiol. 2012;34:317–324.
    1. Morichelli L, Ricci R, Sassi A, Quarta L, Porfili A, Cadeddu N, Santini M. ICD remote monitoring is well accepted and easy to use even for elderly. Eur Heart J. 2011;32(Suppl. 1):32.
    1. Morichelli L, Ricci RP. Remote monitoring of implantable devices: the European experience. Heart Rhythm. 2009;6:1077–1080. doi: 10.1016/j.hrthm.2009.02.030.
    1. Varma N, Stambler B. Patient aspects of remote home monitoring of ICDs—The TRUST Trial. Circulation. 2010;123:e247. doi: 10.1161/CIRCULATIONAHA.110.983148.
    1. Varma N, Michalski J TRUST Investigators. Scheduled ICD patient follow up may be ‘in person or remote’- but are these techniques equivalent? Results from the TRUST Remote Monitoring Trial. Cardiostim Abstracts. 2012
    1. Elsner CH, Sommer P, Piorkowski C, Taborsky M, Neuser H, Bytesnik J, Geller JC, Kottkamp H, Wiesmeth H, Hindricks G. A Prospective Multicenter Comparison Trial of Home Monitoring against Regular Follow-up in MADIT II Patients: Additional Visits and Cost Impact. Comput Cardiol. 2006;33:241–244.
    1. Raatikainen MJ, Uusimaa P, van Ginneken MM, Janssen JP, Linnaluoto M. Remote monitoring of implantable cardioverter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up. Europace. 2008;10:1145–1151. doi: 10.1093/europace/eun203.
    1. Fauchier L, Sadoul N, Kouakam C, Briand F, Chauvin M, Babuty D, Clementy J. Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. Pacing Clin Electrophysiol. 2005;28(Suppl. 1):S255–S259. doi: 10.1111/j.1540-8159.2005.00071.x.
    1. Burri H, Heidbuchel H, Jung W, Brugada P. Remote monitoring: a cost or an investment? Europace. 2011;13(Suppl. 2):ii44–ii48. doi: 10.1093/europace/eur082.
    1. Ricci R, d'Onofrio A, Padeletti L, Sagone A, Vicentini A, Vincenti A, Morichelli L, Cavallaro C, Ricciardi G, Lombardi L, Fusco A, Rovaris G, Silvestri P, Guidotto T, Pollastrelli A, Santini M. Rationale and design of the health economics evaluation registry for remote follow-up: TARIFF. Europace. 2012;14:1661–1665.
    1. Powell BD, Cha YM, Asirvatham SJ, Cesario DA, Cao M, Jones PW, Seth M, Saxon LA, Gilliam FR., III Implantable cardioverter defibrillator electrogram adjudication for device registries: methodology and observations from ALTITUDE. Pacing Clin Electrophysiol. 2011;34:1003–1012. doi: 10.1111/j.1540-8159.2011.03093.x.
    1. Saxon LA, Hayes DL, Gilliam FR, Heidenreich PA, Day J, Seth M, Meyer TE, Jones PW, Boehmer JP. Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation. 2010;122:2359–2367. doi: 10.1161/CIRCULATIONAHA.110.960633.
    1. Hayes DL, Boehmer JP, Day JD, Gilliam FR, 3rd, Heidenreich PA, Seth M, Jones PW, Saxon LA. Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm. 2011;8:1469–1475. doi: 10.1016/j.hrthm.2011.04.015.
    1. Gilliam FR, Hayes DL, Boehmer JP, Day J, Heidenreich PA, Seth M, Jones PW, Stein KM, Saxon LA. Real world evaluation of dual-zone ICD and CRT-D programming compared to single-zone programming: the ALTITUDE REDUCES study. J Cardiovasc Electrophysiol. 2011;22:1023–1029. doi: 10.1111/j.1540-8167.2011.02086.x.
    1. Varma N, Wilkoff B. Advances in remote monitoring of implantable cardiac devices. Cardiol Electrophysiol Clin. 2011;3:463–472. doi: 10.1016/j.ccep.2011.05.011.
    1. Varma N. Automatic remote monitoring of implantable cardiac devices in heart failure: the TRUST Trial. J. Innov Card Rhythm Manage. 2010;1:22–29.

Source: PubMed

3
订阅