Evaluation of EV1000™/volumeView™ for Cardiac Output Monitoring in Liver Transplantation

May 7, 2016 updated by: Gwak Mi Sook, Samsung Medical Center

Evaluation of a New Calibrated Pulse Wave Analysis Method(EV1000™/volumeView™) for Cardiac Output Monitoring in Adult Liver Transplantation

The investigators are trying to evaluate the agreement of cardiac output measurements taken using the specific thermistor-tipped arterial catheter ( the VolumeView ™ catheter) and the EV1000 ™ monitoring platform ( Edward lifesciences, Irvine, CA, USA) with measurements taken using continuous pulmonary artery thermodilution cardiac output monitoring during orthotopic liver transplantation.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

In cirrhotic patient undergoing liver transplantation, there are an altered patterm of circulation, which is characterized by increased cardiac output, decreased peripheral vascular resistance, and reduced ventricular response to physiological, pharmacological, and surgical stress. Large blood loss and clamping/declamping of the inferior vena cava and portal vein during liver transplantation surgery can affect intravascular volume status, which in turn can lead to hemodynamic instability. Therefore, reliable cardiac output monitoring is particularly useful in the cirrhotic patient undergoing liver transplant. Pulmonary artery thermodilution has been used as a standard method for cardiac output assessment for many years. Pulmonary artery catheter can cause rare but serious complications. And it has some limitations for continuously monitoring cardiac output during rapid hemodynamic changes.

Recently, several minimally invasive CO monitors have been developed. The pulse contour technique continuously estimates CO through mathematical analysis of the waveform of arterial pressure. However, previous studies evaluating the reliability of the pulse contour techniques indicated conflicting results in cirrhotic patients during liver transplantation. Most of studies were performed by analyzing from radial artery pressure waveform. Because of rapid changing of intravascular volume, using inotropics, the monitoring of femoral artery pressure has been recommended during liver transplantation.

A new pulse wave analysis system has developed and introduced into clinical practice that consists of a specific thermistor -tipped femoral arterial catheter ( the VolumeView™ catheter) and the EV1000™ monitoring platform ( Edward lifesciences, Irvine, CA, USA). To continuously assess CO based on the femoral arterial pressure curve signal and it uses transpulmonary thermodilution for calibration. We are trying to evaluate the agreement of cardiac output measurements taken using EV1000™ / VolumeView™ with measurements taken using continuous pulmonary artery thermodilution cardiac output monitoring during orthotopic liver transplantation.

Study Type

Observational

Enrollment (Actual)

28

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

20 years to 70 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

patients who are performed liver transplantation in Samsung medical center

Description

Inclusion Criteria:

subjects undergoing living donor liver transplantation during the study period subjects older than 20yrs who can give informed consent

Exclusion Criteria:

those who are confirmed moderate severe aortic regurgitation by echocardiography those with infection on cannulation site

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
EV1000™/volumeView™

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
cardiac output
Time Frame: intraoperative
L/minute, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
intraoperative

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
systemic vascular resistance
Time Frame: intraoperative
dyne s/cm5,28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
intraoperative
stroke volume variability
Time Frame: during surgery every 10 minute, event time for example, anhepatic, liver in, reperfusion, using inotropics, changing ventilator setting
%, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
during surgery every 10 minute, event time for example, anhepatic, liver in, reperfusion, using inotropics, changing ventilator setting
stroke volume
Time Frame: intraoperative
ml, 28 participants, EV1000 by femiral artery, Pulmonary artery thermodilution
intraoperative
ejection fraction
Time Frame: intraoperative
%, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
intraoperative

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Mi Hye Park, MD, Samsung Medical Center

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

October 1, 2014

Primary Completion (Actual)

September 1, 2015

Study Completion (Actual)

September 1, 2015

Study Registration Dates

First Submitted

October 17, 2014

First Submitted That Met QC Criteria

November 29, 2014

First Posted (Estimate)

December 3, 2014

Study Record Updates

Last Update Posted (Estimate)

May 10, 2016

Last Update Submitted That Met QC Criteria

May 7, 2016

Last Verified

May 1, 2016

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • 2014-08-080-002

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Liver Disease

Clinical Trials on EV1000™/volumeView™

Subscribe