Evaluation of EV1000™/volumeView™ for Cardiac Output Monitoring in Liver Transplantation
Evaluation of a New Calibrated Pulse Wave Analysis Method(EV1000™/volumeView™) for Cardiac Output Monitoring in Adult Liver Transplantation
Studieoversigt
Detaljeret beskrivelse
In cirrhotic patient undergoing liver transplantation, there are an altered patterm of circulation, which is characterized by increased cardiac output, decreased peripheral vascular resistance, and reduced ventricular response to physiological, pharmacological, and surgical stress. Large blood loss and clamping/declamping of the inferior vena cava and portal vein during liver transplantation surgery can affect intravascular volume status, which in turn can lead to hemodynamic instability. Therefore, reliable cardiac output monitoring is particularly useful in the cirrhotic patient undergoing liver transplant. Pulmonary artery thermodilution has been used as a standard method for cardiac output assessment for many years. Pulmonary artery catheter can cause rare but serious complications. And it has some limitations for continuously monitoring cardiac output during rapid hemodynamic changes.
Recently, several minimally invasive CO monitors have been developed. The pulse contour technique continuously estimates CO through mathematical analysis of the waveform of arterial pressure. However, previous studies evaluating the reliability of the pulse contour techniques indicated conflicting results in cirrhotic patients during liver transplantation. Most of studies were performed by analyzing from radial artery pressure waveform. Because of rapid changing of intravascular volume, using inotropics, the monitoring of femoral artery pressure has been recommended during liver transplantation.
A new pulse wave analysis system has developed and introduced into clinical practice that consists of a specific thermistor -tipped femoral arterial catheter ( the VolumeView™ catheter) and the EV1000™ monitoring platform ( Edward lifesciences, Irvine, CA, USA). To continuously assess CO based on the femoral arterial pressure curve signal and it uses transpulmonary thermodilution for calibration. We are trying to evaluate the agreement of cardiac output measurements taken using EV1000™ / VolumeView™ with measurements taken using continuous pulmonary artery thermodilution cardiac output monitoring during orthotopic liver transplantation.
Undersøgelsestype
Tilmelding (Faktiske)
Kontakter og lokationer
Studiesteder
-
-
-
Seoul, Korea, Republikken, 135-710
- Samsung Medical Center
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Prøveudtagningsmetode
Studiebefolkning
Beskrivelse
Inclusion Criteria:
subjects undergoing living donor liver transplantation during the study period subjects older than 20yrs who can give informed consent
Exclusion Criteria:
those who are confirmed moderate severe aortic regurgitation by echocardiography those with infection on cannulation site
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
Kohorter og interventioner
Gruppe / kohorte |
Intervention / Behandling |
|---|---|
|
EV1000™/volumeView™
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
cardiac output
Tidsramme: intraoperative
|
L/minute, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
|
intraoperative
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
systemic vascular resistance
Tidsramme: intraoperative
|
dyne s/cm5,28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
|
intraoperative
|
|
stroke volume variability
Tidsramme: during surgery every 10 minute, event time for example, anhepatic, liver in, reperfusion, using inotropics, changing ventilator setting
|
%, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
|
during surgery every 10 minute, event time for example, anhepatic, liver in, reperfusion, using inotropics, changing ventilator setting
|
|
stroke volume
Tidsramme: intraoperative
|
ml, 28 participants, EV1000 by femiral artery, Pulmonary artery thermodilution
|
intraoperative
|
|
ejection fraction
Tidsramme: intraoperative
|
%, 28 participants, EV1000 by femoral artery, Pulmonary artery thermodilution
|
intraoperative
|
Samarbejdspartnere og efterforskere
Sponsor
Efterforskere
- Ledende efterforsker: Mi Hye Park, MD, Samsung Medical Center
Publikationer og nyttige links
Generelle publikationer
- Feltracco P, Biancofiore G, Ori C, Saner FH, Della Rocca G. Limits and pitfalls of haemodynamic monitoring systems in liver transplantation surgery. Minerva Anestesiol. 2012 Dec;78(12):1372-84. Epub 2012 Aug 3.
- Uemura K, Kawada T, Inagaki M, Sugimachi M. A minimally invasive monitoring system of cardiac output using aortic flow velocity and peripheral arterial pressure profile. Anesth Analg. 2013 May;116(5):1006-1017. doi: 10.1213/ANE.0b013e31828a75bd. Epub 2013 Mar 14. Erratum In: Anesth Analg. 2014 Jan;118(1):242.
- Biancofiore G, Critchley LA, Lee A, Bindi L, Bisa M, Esposito M, Meacci L, Mozzo R, DeSimone P, Urbani L, Filipponi F. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth. 2009 Jan;102(1):47-54. doi: 10.1093/bja/aen343.
- Biancofiore G, Critchley LA, Lee A, Yang XX, Bindi LM, Esposito M, Bisa M, Meacci L, Mozzo R, Filipponi F. Evaluation of a new software version of the FloTrac/Vigileo (version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg. 2011 Sep;113(3):515-22. doi: 10.1213/ANE.0b013e31822401b2. Epub 2011 Jun 16.
- Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP. Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Proc. 2012 Mar;44(2):433-7. doi: 10.1016/j.transproceed.2011.12.045.
- Galluccio ST, Chapman MJ, Finnis ME. Femoral-radial arterial pressure gradients in critically ill patients. Crit Care Resusc. 2009 Mar;11(1):34-8.
- Arnal D, Garutti I, Perez-Pena J, Olmedilla L, Tzenkov IG. Radial to femoral arterial blood pressure differences during liver transplantation. Anaesthesia. 2005 Aug;60(8):766-71. doi: 10.1111/j.1365-2044.2005.04257.x.
- Kim YK, Shin WJ, Song JG, Jun IG, Kim HY, Seong SH, Hwang GS. Comparison of stroke volume variations derived from radial and femoral arterial pressure waveforms during liver transplantation. Transplant Proc. 2009 Dec;41(10):4220-8. doi: 10.1016/j.transproceed.2009.09.050.
- Sakka SG, Kozieras J, Thuemer O, van Hout N. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth. 2007 Sep;99(3):337-42. doi: 10.1093/bja/aem177. Epub 2007 Jul 4.
- Bendjelid K, Marx G, Kiefer N, Simon TP, Geisen M, Hoeft A, Siegenthaler N, Hofer CK. Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients. Br J Anaesth. 2013 Oct;111(4):573-9. doi: 10.1093/bja/aet116. Epub 2013 Apr 26.
- Kiefer N, Hofer CK, Marx G, Geisen M, Giraud R, Siegenthaler N, Hoeft A, Bendjelid K, Rex S. Clinical validation of a new thermodilution system for the assessment of cardiac output and volumetric parameters. Crit Care. 2012 May 30;16(3):R98. doi: 10.1186/cc11366.
- Bendjelid K, Giraud R, Siegenthaler N, Michard F. Validation of a new transpulmonary thermodilution system to assess global end-diastolic volume and extravascular lung water. Crit Care. 2010;14(6):R209. doi: 10.1186/cc9332. Epub 2010 Nov 23.
Datoer for undersøgelser
Studer store datoer
Studiestart
Primær færdiggørelse (Faktiske)
Studieafslutning (Faktiske)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Skøn)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Skøn)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- 2014-08-080-002
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Lever sygdom
-
National Health Research Institutes, TaiwanRekrutteringMindfulness | Senlivsdepression | Voksne for sent liv | Prodromal depression af sent liv | HjernestimuleringsinterventionTaiwan
-
Universidad Complutense de MadridRekruttering
-
L'OrealRekrutteringPigmentering fra det virkelige livIndien
-
Université de Reims Champagne-ArdenneAfsluttetAdolescentes følelsesmæssige og seksuelle livFrankrig
-
Sisli Hamidiye Etfal Training and Research HospitalIkke rekrutterer endnu
-
Bournemouth UniversityDorset County Hospital NHS Foundation TrustAfsluttetErfaring, livDet Forenede Kongerige
-
University of California, Los AngelesVA Greater Los Angeles Healthcare SystemTrukket tilbageVeteranfamilier | Familiekommunikation | Civilt liv reintegrationForenede Stater
-
Swiss Federal Institute of TechnologyAfsluttetÆldret | Uafhængigt livSchweiz
-
Diakonie Kliniken ZschadraßChemnitz University of TechnologyAfsluttetMantra meditation | Mantra Meditation + Kropsorienteret Yoga | Mantra Meditation + Etisk Liv | Mantra Meditation + Kropsorienteret Yoga + Etisk LivTyskland
-
CalydialOZ'IRIS SantéAfsluttetPatientengagement | Patientdeltagelse | Patienttilfredshed | Dialyse | Patientforhold, Sygeplejerske | Erfaring, livFrankrig
Kliniske forsøg med EV1000™/volumeView™
-
Iuliu Hatieganu University of Medicine and PharmacyUkendtSepsis | Septisk chok | Hæmodynamisk ustabilitetRumænien
-
CAMC Health SystemEdwards LifesciencesUkendtKritisk sygdom | TraumaForenede Stater
-
BaroNova, Inc.Afsluttet
-
London Health Sciences CentreUniversity of Western Ontario, Canada; Synaptive MedicalUkendtHjerneskade, kronisk | Cerebellar Kognitiv Affektiv Syndrom | Cerebellar mutisme
-
Medtronic CardiovascularMedtronicAfsluttetAortastenoseNew Zealand, Det Forenede Kongerige, Australien
-
Gülçin Özalp GerçekerAfsluttetKateterkomplikationer | KateterlækageKalkun
-
CereVasc IncAlvaMed, Inc.; Simplified Clinical Data Systems, LLC; Bioscience Consulting...Aktiv, ikke rekrutterendeHydrocephalus | Hydrocephalus, KommunikerendeArgentina
-
Asklepios Kliniken Hamburg GmbHUniversity of KielAfsluttetLuftvejsstyring | Larynx maske Airway | Fibreoptisk intubation
-
Evasc Medical Systems Corp.Trukket tilbageIntrakranielle aneurismerCanada
-
Andrew ParrentUniversity of Western Ontario, Canada; Synaptive MedicalUkendtTemporallappepilepsi